  91   Sat Feb 27 18:44:23 2021 Felix Ulrich-PurWaveforms SiC+Si17SiC+Si+RaMon800VII  Good        

Start 11:55
Stopped at 12:05-->acc failure

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.100 V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x = 8
y = 68

  90   Sat Feb 27 18:44:06 2021 Felix Ulrich-PurWaveforms SiC+Si16SiC+Si+RaMon100.4I(EFE+SID0)  Good        

Start 11:36

schöne Signale

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.100 V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x = 8
y = 68

  89   Sat Feb 27 18:43:35 2021 Felix Ulrich-PurWaveforms SiC+Si15SiC+Si+RaMon100.4III(EFE+SID2)  Good        

Start 11:09

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.100 V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x = 8
y = 68

  88   Sat Feb 27 18:43:12 2021 Felix Ulrich-PurWaveforms SiC+Si14SiC+Si+RaMon145.4III(EFE+SID2)  Good        

Start 10:47

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.100 V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x = 8
y = 68

  87   Sat Feb 27 18:42:13 2021 Felix Ulrich-PurWaveforms SiC+Si13SiC+Si+RaMon252.7III(EFE+SID2)  Good        

Start 10:38
no signal in SiC

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.100 V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x = 8
y = 68

  86   Sat Feb 27 18:38:54 2021 Felix Ulrich-PurWaveforms SiC+Si12SiC+Si+RaMon252.7I(EFE+SID0)  Good        

Start 10:28
polarity now correct

now signal in SiC

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.100 V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x = 8
y = 68

  85   Sat Feb 27 18:38:09 2021 Felix Ulrich-PurWaveforms SiC+Si11SiC+Si+RaMon62.4I(EFE+SID0)  Good        

Start 10:24
polarity now correct
no signal in sic --> energy too low due to scintis

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.100 V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x = 8
y = 68

  84   Sat Feb 27 18:36:15 2021 Felix Ulrich-PurWaveforms SiC+Si10SiC+Si+RaMon145.4I(EFE+SID0)  Good        

Start 10:10
polarity now correct

Wiederholung von vorher, aber keine Aufnahme von neuen Waveforms

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.100 V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x = 8
y = 68

  83   Sat Feb 27 09:42:44 2021 Felix Ulrich-PurWaveforms SiC+Si9SiC+Si+RaMon145.4I(EFE+SID0)  Good        

Start 09:37

SiC setrup had to be rewired (wr0ng polarity somewhere (insert here))

 

Umstellen:
channel 2: LVDS-Signal vertauschen
channel 2 Referenz: 2.100 V 

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.475V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x = 8
y = 68

  82   Sat Feb 27 09:41:06 2021 Felix Ulrich-PurWaveforms SiC+Si8SiC+Si+RaMon252.7I(EFE+SID0)  Good        

08:34 room request, PC for SiC had to be restarted, trigger were connected as prior to shift

Start 08:53
rate looks fine

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.475V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

verändere Position, ab 7:49 bei Aufnahme von Waveforms Position:

x = 0
y = 0

7:51 verändere Position

7:58 neue Position:
x = 8
y = 68

 

FPGA shutdown --> Datenaufzeichnung bis etwa 8:27, ab dann Suchen des Fehlers

 

  81   Sat Feb 27 09:40:15 2021 Felix Ulrich-PurRate verification7SiC+Si+RaMon62.4I(EFE+SID0)  Good        

Start 07:40
rate looks fine, however operators mentioned, that CPD data might have been loaded incorrectly-->pause during run and reloaded (07:48), rate seems unchanged

 

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.475V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x: 0
y: 0 
 

  80   Sat Feb 27 09:31:57 2021 Felix Ulrich-PurRate verification6SiC+Si+RaMon252.7I(EFE+SID0)  Good        

Start 07:16
rate looks fine,

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.475V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x: 0
y: 0 

SiC: man sieht erstmals Signale 

  79   Sat Feb 27 09:30:32 2021 Felix Ulrich-PurRate verification5SiC+Si+RaMon62.4III(EFE+SID2)  Good        

Start 07:03
verified if run 1 was reliable-->Same results

 

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.475V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x: 0
y: 0 

  78   Sat Feb 27 09:29:41 2021 Felix Ulrich-PurRate verification4SiC+Si+RaMon252.7III(EFE+SID2)  Good        

Start 06:55
reallocated correctly, rate looks similar to old measurements now (kHz ish)

 

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.475V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x: 0
y: 0 

  77   Sat Feb 27 09:28:07 2021 Felix Ulrich-PurRate verification2SiC+Si+RaMon252.7III(EFE+SID2)  Crap        

Start 06:47
rate only 10Hz, stopped run, room request and rewired trigger.

NOTE: Problem was caused by operators: Optics were not set correctly after ramping up energy from 62.4 to 252.7, not trigger cable

 

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.475V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x: 0
y: 0 

  76   Sat Feb 27 09:26:27 2021 Felix Ulrich-PurRate verification1SiC+Si+RaMon62.4III(EFE+SID2)  Unknown        

Start 06:37
rate looks okish. few emtpy spills

 

2021-02-27, 06:07:

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.475V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.112V channel 1, 0-Level: 2.315V channel 1

position:

x: 0
y: 0 

2021-02-27, 06:30:

Referenz: 2.477V channel 1
channel 2 unverändert
 

2021-02-27, 06:37:

Setting: 3 

Run 1: 6:37 62.4 MeV, Rate geringer als erwartet 

 

  75   Wed Jan 20 17:57:16 2021 Felix Ulrich-PurLowflux p800 commissioning85Ramon+Tracker800  30 spillsUnknown100%60.5%20106dB   

MT=1, BSY=50,
setting V steering VII,beam looks centered, but rate and spot looks similar to setting VII, maybe wrong acc setting? --> don't use

  74   Wed Jan 20 17:56:06 2021 Felix Ulrich-PurLowflux p800 commissioning84Ramon+Tracker800  30 spills 12%60.5%20306dB   

MT=1, BSY=50,
setting VI steering VII,beam is off again

  73   Wed Jan 20 17:55:13 2021 Felix Ulrich-PurLowflux p800 commissioning83Ramon+Tracker800  30 spills 100%60.5%20306dB   

MT=1, BSY=50,
final setting VII
new steering VII +12.5mm x,+13mm y ( -beam direction)

  72   Wed Jan 20 17:52:49 2021 Felix Ulrich-PurLowflux p800 commissioning82Ramon+Tracker800  30 spills 100%60.5%20306dB   

MT=1, BSY=50,
potential setting VII
new steering +12.5mm x,+10mm y ( -beam direction), still off