  Alibava testbeam at DESY in Oct. 2017 (Dominic, Peter), Page 1 of 7  Not logged in ELOG logo
Entry  Wed Oct 18 12:22:03 2017, Peter Paulitsch, IFX_CenterBias, run000099, 99, 110000, 122.9, 23.5, -3.3°, No, Good 
run settings
Parameter Value
Alibava filename  run000099.dat
Telescope filename  run000099.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -160 nA
Sensor temperature 24.6 °C
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y 122.9, 23.5
Region of interest bias line
Trigger frequency  185 Hz
Meas. time 17 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 52.3%

 

Entry  Wed Oct 18 12:19:14 2017, Peter Paulitsch, IFX_CenterBias, run000093, 93, 1010000, 122.9, 26, -3.3°, No, Good 
run settings
Parameter Value
Alibava filename  run000093.dat
Telescope filename  run000093.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -165 nA
Sensor temperature 24.5 °C
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 122.9, 26
Region of interest all
Trigger frequency  185 Hz
Meas. time 90 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 44.1%

 

Entry  Wed Oct 18 12:12:55 2017, Peter Paulitsch, IFX_CenterBias, run000091, 91, 1010000, 122.9, 21, -3.3°, No, Good 
run settings
Parameter Value
Alibava filename  run000091.dat
Telescope filename  run000091.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -165 nA
Sensor temperature 24.6 °C
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 122.9, 21
Region of interest lower part
Trigger frequency  185 Hz
Meas. time 90 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 44.7%

 

Entry  Wed Oct 18 11:59:30 2017, Peter Paulitsch, IFX_CenterBias, run000088, 88, 1010000, 122.9, 31, -3.3°, No, Good 
run settings
Parameter Value
Alibava filename  run000088.dat
Telescope filename  run000088.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -165 nA
Sensor temperature 24.5 °C
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 122.9, 31
Region of interest upper part
Trigger frequency  185 Hz
Meas. time 90 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 45.4%

 

Entry  Wed Oct 18 11:48:51 2017, Peter Paulitsch, IFX_CenterBias, run000085, 85, 1010000, 122.9, 26, -3.3°, No, Good 
run settings
Parameter Value
Alibava filename  run000085.dat
Telescope filename  run000085.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -165 nA
Sensor temperature 24.5 °C
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 122.9, 26
Region of interest all
Trigger frequency  185 Hz
Meas. time 90 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 44.1%

 

Entry  Tue Oct 17 15:15:32 2017, Dominic Bloech, IFX_CenterBias, run000083 , 83, 260000, 122.9, 26, -3.3°, No, Unknown 
run settings
Parameter Value
Alibava filename  run000083.dat
Telescope filename  run000083.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -162 nA
Sensor temperature 22 °C
Latency 140 [a.u.]
Alibava triggers 260000
Telescope triggers  
Data synced yes
Table position x/y 122.9, 26
Region of interest all
Trigger frequency  220 Hz
Meas. time 90 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 44,2%

 

Entry  Tue Oct 17 15:13:42 2017, Dominic Bloech, IFX_CenterBias, run000081, 81, 1010000, 122.9, 26, -3.3°, No, Good 
run settings
Parameter Value
Alibava filename  run000081.dat
Telescope filename  run000081.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -162 nA
Sensor temperature 22 °C
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 122.9, 26
Region of interest all
Trigger frequency  220 Hz
Meas. time 90 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 44.1%

 

Entry  Tue Oct 17 14:46:26 2017, Dominic Bloech, IFX_CenterBias, run000080, 80, 500000, 122.9, 26, -3.3°, No, Unknown 

 

run settings
Parameter Value
Alibava filename  run000080.dat
Telescope filename  run000080.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -162 nA
Sensor temperature 22 °C
Latency 140 [a.u.]
Alibava triggers 500000
Telescope triggers  
Data synced yes
Table position x/y 122.9, 26
Region of interest all
Trigger frequency  220 Hz
Meas. time 90 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 43,6%

 

Entry  Mon Oct 30 12:04:14 2017, Peter Paulitsch, IFX_200_1e15, run000389, 389, 100000, 76.50, 66.00, 0°, Yes, Unknown 
Parameter Value
Alibava filename  run000389.dat
Telescope filename  run000389.raw
Pedestal filename KIT_pedestal_600V_26102017.ped
Bias voltage -600 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 100000
Telescope triggers 100000
Data synced yes
Table position x/y 76.50, 66.00
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency  
Entry  Mon Oct 30 12:03:27 2017, Peter Paulitsch, IFX_200_1e15, run000388, 388, 1010000, 76.50, 66.00, 0°, Yes, Good 
Parameter Value
Alibava filename  run000388.dat
Telescope filename  run000388.raw
Pedestal filename KIT_pedestal_600V_26102017.ped
Bias voltage -400 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 76.50, 66.00
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency  
Entry  Mon Oct 30 12:02:54 2017, Peter Paulitsch, IFX_200_1e15, run000387, 387, 1010000, 76.50, 66.00, 0°, Yes, Good 
Parameter Value
Alibava filename  run000387.dat
Telescope filename  run000387.raw
Pedestal filename KIT_pedestal_600V_26102017.ped
Bias voltage -800 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 76.50, 66.00
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency  
Entry  Mon Oct 30 12:02:05 2017, Peter Paulitsch, IFX_200_1e15, run000384, 384, 1010000, 76.50, 66.00, 0°, Yes, Good 
Parameter Value
Alibava filename  run000384.dat
Telescope filename  run000384.raw
Pedestal filename KIT_pedestal_600V_26102017.ped
Bias voltage -500 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 76.50, 66.00
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency  
Entry  Mon Oct 30 12:00:43 2017, Peter Paulitsch, IFX_200_1e15, run000383, 383, 2709999, 76.50, 66.00, 0°, Yes, Good 
Parameter Value
Alibava filename  run000383.dat
Telescope filename  run000383.raw
Pedestal filename KIT_pedestal_600V_26102017.ped
Bias voltage -600 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 2709999
Telescope triggers 2709950
Data synced yes
Table position x/y 76.50, 66.00
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency  
Entry  Mon Oct 30 11:55:04 2017, Peter Paulitsch, IFX_200_1e15, run000379, 379, 150000, 76.50, 66.00, 0°, Yes, Good 
Parameter Value
Alibava filename  run000379.dat
Telescope filename  run000379.raw
Pedestal filename KIT_pedestal_600V_26102017.ped
Bias voltage -400 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 150000
Telescope triggers 150000
Data synced yes
Table position x/y 76.50, 66.00
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency  
Entry  Mon Oct 30 11:54:14 2017, Peter Paulitsch, IFX_200_1e15, run000377, 377, 1010000, 76.50, 66.00, -10°, Yes, Good 
Parameter Value
Alibava filename  run000377.dat
Telescope filename  run000377.raw
Pedestal filename KIT_pedestal_600V_26102017.ped
Bias voltage -600 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 76.50, 66.00
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.7%
Entry  Mon Oct 30 11:53:33 2017, Peter Paulitsch, IFX_200_1e15, run000375, 375, 1010000, 75.00, 66.00, -20°, Yes, Good 
Parameter Value
Alibava filename  run000375.dat
Telescope filename  run000375.raw
Pedestal filename KIT_pedestal_600V_26102017.ped
Bias voltage -600 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 75.00, 66.00
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency  
Entry  Mon Oct 30 11:52:26 2017, Peter Paulitsch, IFX_200_1e15, run000373, 373, 1010000, 75.00, 66.00, -30°, Yes, Good 
Parameter Value
Alibava filename  run000373.dat
Telescope filename  run000373.raw
Pedestal filename KIT_pedestal_600V_26102017.ped
Bias voltage -600 V
Dark current 100 uA
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 75.00, 66.00
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency  
Entry  Mon Oct 30 11:51:46 2017, Peter Paulitsch, HPK_200_1e15, run000368, 368, 110000, 76.50, 66.00, -30°, Yes, Good 
Parameter Value
Alibava filename  run000368.dat
Telescope filename  run000368.raw
Pedestal filename KIT_pedestal_600V_26102017.ped
Bias voltage -600 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y 76.50, 66.00
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency  
Entry  Mon Oct 30 11:50:22 2017, Peter Paulitsch, HPK_200_1e15, run000365, 365, 510000, 76.50, 66.00, 4°, Yes, Good 
Parameter Value
Alibava filename  run000365.dat
Telescope filename  run000365.raw
Pedestal filename KIT_pedestal_600V_26102017.ped
Bias voltage -600 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 510000
Telescope triggers 500000
Data synced yes
Table position x/y 76.50, 66.00
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency  
Entry  Mon Oct 30 11:48:12 2017, Peter Paulitsch, HPK_200_1e15, run000364, 364, 250000, 76.50, 66.00, 5°, Yes, Good 
Parameter Value
Alibava filename  run000364.dat
Telescope filename  run000364.raw
Pedestal filename KIT_pedestal_600V_23102017.ped
Bias voltage -600 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 250000
Telescope triggers 250000
Data synced yes
Table position x/y 76.50, 66.00
Region of interest bias line
Trigger frequency  135 Hz
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency  