Home CMS Production Clean room MedAustron HEPHY testbeams old
electronics module assembly SiDDaTA scratch
  HEPHY logbook of the Electronics Group, Page 3 of 4  Not logged in ELOG logo
New entries since:Thu Jan 1 01:00:00 1970
  ID Date Author Project Measurement Type Object ID Subject Text Attachments
  26   Thu May 8 16:51:03 2014 Benedikt WürknerBelle IIsourceL3 moduleNo 1 of 6 room temperature measurements using Sr90 Source (single)Sr90 Radioactive Source
No cooling
Module position: 0:0(top left)
 10x 
  23   Wed May 7 18:20:36 2014 Benedikt WürknerBelle IIsourceL3 moduleNo 6 of 6 room temperature measurements using Sr90 Source (multi6)Sr90 Radioactive Source
No cooling
Module position: -57:-30(bottom right)
 10x 
  22   Wed May 7 17:33:21 2014 Benedikt WürknerBelle IIsourceL3 moduleNo 5 of 6 room temperature measurements using Sr90 Source (multi6)Sr90 Radioactive Source
No cooling
Module position: -32:-30(bottom center)
 10x 
  21   Wed May 7 16:43:37 2014 Benedikt WürknerBelle IIsourceL3 moduleNo 4 of 6 room temperature measurements using Sr90 Source (multi6)Sr90 Radioactive Source
No cooling
Module position: -5:-30(bottom left)
 10x 
  20   Wed May 7 15:53:38 2014 Benedikt WürknerBelle IIsourceL3 moduleNo 3 of 6 room temperature measurements using Sr90 Source (multi6)Sr90 Radioactive Source
No cooling
Module position: -59:-20(top right)
 10x 
  19   Wed May 7 15:04:32 2014 Benedikt WürknerBelle IIsourceL3 moduleNo 2 of 6 room temperature measurements using Sr90 Source (multi6)Sr90 Radioactive Source
No cooling
Module position: -32:-20(top center)
 10x 
  18   Wed May 7 14:36:29 2014 Benedikt WürknerBelle IIsourceL3 moduleNo 1 of 6 room temperature measurements using Sr90 Source (multi6)Sr90 Radioactive Source
No cooling
Module position: -5:-20(top left)
 10x 
  17   Sun Jul 4 05:48:36 2010 Christian IrmlerBELLE UpgradesourceOrigami 6 - module 1run002: first analysis resultsRun name: run002

Run type: 0 (Hardware (Normal Run))
 7x 
  16   Wed Oct 7 14:24:06 2009 Dieter UhlBELLE Upgradehybrid#5hybrid-pitchadaptershorts at upper coat
 pitchadapter5_lower_coat_opens.gifpitchadapter5_lower_coat.gifpitchadapter5_upper_coat_opens.gifpitchadapter5_upper_coat.gif 
  15   Wed Oct 7 14:23:35 2009 Dieter UhlBELLE Upgradehybrid#4hybrid-pitchadapteropens at upper coat
 pitchadapter4_lower_coat.gifpitchadapter4_lower_coat_opens.gifpitchadapter4_upper_coat.gifpitchadapter4_upper_coat_opens.gif 
  14   Tue May 20 14:27:50 2008 Markus FriedlBELLE Upgradesourcemicronanalysis results of source test
*** NOTE: AFTER THIS MEASUREMENT WE REALIZED
THAT BIASING WAS NOT DONE PROPERLY
 9x 
  13   Fri May 9 10:04:26 2008 Christian IrmlerSPS Testbeam June08modulemodule 05/05properties (noise, intcal), APVs bonded to the sensorModule tested with 1 and 2 rows bonded to
the sensor, respectively. 
HV = 100 V
Ibias
(100 V) = 18.0 nA
Ibias (200 V) = 23.6 nA
 12x 
  12   Fri May 9 10:00:34 2008 Christian IrmlerSPS Testbeam June08modulemodule 03/10properties (noise, intcal), APVs bonded to the sensorModule tested with 1 and 2 rows bonded to
the sensor, respectively. 
HV = 100 V
Ibias
(100 V) = 18.9 nA
Ibias (200 V) = 25.5 nA
 12x 
  11   Fri May 9 09:56:15 2008 Christian IrmlerSPS Testbeam June08modulemodule 06/03properties (noise, intcal), APVs bonded to the sensorModule tested with 1 and 2 rows bonded to
the sensor, respectively. 
HV = 100 V
Ibias
(100 V) = 26.5 nA
Ibias (200 V) = 37.8 nA
 12x 
  10   Wed May 7 19:05:08 2008 Christian IrmlerSPS Testbeam June08modulemodule 04/04properties (noise, intcal), APVs bonded to the sensorModule tested with 1 and 2 rows bonded to
the sensor, respectively. 
HV = 100 V
Ibias
(100 V) = 27.8 nA
Ibias (200 V) = 32.7 nA
 12x 
  7   Wed May 7 16:12:58 2008 Christian IrmlerSPS Testbeam June08modulemodule 12/08properties (noise, intcal), APVs bonded to the sensorModule tested with 1 and 2 rows bonded to
the sensor, respectively. 
HV = 100 V
Ibias
(100 V) = 22.2 nA
Ibias (200 V) = 26.5 nA
 12x 
  8   Wed May 7 15:40:07 2008 Christian IrmlerSPS Testbeam June08modulemodule 07/07properties (noise, intcal), APVs bonded to the sensorModule tested with 1 and 2 rows bonded to
the sensor, respectively. 
HV = 100 V
Ibias
(100 V) = 20.2 nA
Ibias (200 V) = 21.9 nA
 12x 
  9   Wed May 7 15:26:24 2008 Christian IrmlerSPS Testbeam June08modulemodule 20/09properties (noise, intcal), APVs bonded to the sensorModule tested with 1 and 2 rows bonded to
the sensor, respectively. 
HV = 100 V
Ibias
(100 V) = 25.1 nA
Ibias (200 V) = 31.4 nA
 12x 
  6   Wed May 7 15:24:49 2008 Christian IrmlerSPS Testbeam June08modulemodule 10/02properties (noise, intcal), APVs bonded to the sensorModule tested with 1 and 2 rows bonded to
the sensor, respectively.
HV = 100 V
Ibias
(100V) = 19.1 nA
Ibias (200V) = 23.7 nA
 12x 
  5   Wed Apr 30 16:52:17 2008 Markus FriedlBELLE Upgrademodulemicronmicron sensor glued to framesoeben haben wir den micron-DSSD (double metal
layer) in den 2-teiligen rahmen geklebt und
auf beiden seiten
  
ELOG V3.1.5-fc6679b