Entry  Tue May 19 11:01:29 2015, Hao Yin, Belle II, system, Pedestal, PedestalRun 16x

FIRRun001

Baseline measurement without injections, 50000 event,

room temperature.

noise results see attachments.

Entry is currently edited by Hao Yin on 255.255.255.255  Entry  Fri May 29 11:47:56 2015, Hao Yin, Belle II, system, PS Filter, Testing PS LV filter and quantifying the required min noise lvl 

Data of this entry is recorded in the folder: LV315kHz_Injections
Injecting noise to LV with a freq. of 314 kHz to emulate Caen PS with KenWood PS.

Run Name 315kHzKenWood_: (injecting cmc noise into p-side lv) (attachment 1)
000 ... baseline noise
001 ... 4mA
002 ... 1mA
003 ... 1mA wo Amplifier
004 ... 4mA wo Amplifier
005 ... 0.5mA wo Amplifier
006 ... 2mA wo Amplifier
007 ... 0.2mA wo Amplifier

Run Name 315kHzKenWood_Filter_BW_: (injecting cmc noise into p-side lv with filters (inductance with 470 mH x 6 ) conn. at bw) (attachment 1) (wo Amplifier)
000 ... baseline noise
001 ... 0.2mA
002 ... 0.4mA
003 ... 0.5mA
004 ... 1mA
005 ... 2mA
006 ... 4mA

Run Name 315kHzKenWood_Filer_BW_HVRET-GND:
000 ... baseline noise
001 ... 0.2mA
002 ... 0.4mA
 

Entry  Sun Jul 4 05:48:36 2010, Christian Irmler, BELLE Upgrade, source, Origami 6 - module 1, run002: first analysis results 7x

Run name: run002

Run type: 0 (Hardware (Normal Run))
Comments:
After ~40min warmup
HV=80
90Sr 1mCi source - moved compared to run001
multi6

Max. Events=100000      Trg delay=25

Origami 6 module #1
w/o cooling
sensor: B2HPK_10938-9239_8

n-side: all 4 APVs read out
p-side: APV #0 to #3  read out, analog output of #4 and #5 were not connected to FADC board

 

 

Entry  Tue May 19 14:48:06 2015, Hao Yin, Belle II, module, L6 Noise inJection Runs, L6 Noise inJection Runs 

Folder "L6_RLC_Test"

Configuration file: /mnt/data/ITA_NOISE_TESTS/L6_RLC_Test/SinglePeak_3Fadc.cfg

3 Tests: Injkection to 2.5V 1.25V. ground and their combination

Frequency scan with const ampl.:

points (index (run number) corresponds to frequency):

index  -  frequency[MHz]

1 - 0,5
2 - 0,7
3 - 0,8
4 - 1
5 - 2
6 - 3
7 - 5
8 - 7
9 - 8
10 - 10
11 - 15
12 - 20
13 - 30
14 - 40
15 - 50
16 - 80

0 - base line for comparison

Base line run with out injection.

Injection Test L6:

2.5V, 1.25V, ground (data prefix all_)

Entry  Mon May 25 10:38:03 2015, Hao Yin, Belle II, system, L5 Ladder Emission Test, Emission 

(TestNr: 16)

base line noise measurement are saved in SYSTEM_CHECK_AFTER_WE.
(noise wings at fw n side decreased !!!)

Measuring noise emitted by our system.
1) noise emmision at the cable from dockbox to the hybrid (fw n and p side)
 

Entry  Wed May 7 14:36:29 2014, Benedikt Würkner, Belle II, source, L3 module, No 1 of 6 room temperature measurements using Sr90 Source (multi6) 10x

Sr90 Radioactive Source
No cooling
Module position: -5:-20(top left)
HV Turned to +40,-40V with currents. +1.2 and -1.2 µA

N+1.25V current: 0.22A
N+2.5 V current: 0.56A
P+1.25V current: 0.26A
P+2.50V current: 0.70A

Events: 200000

Entry  Wed May 7 15:04:32 2014, Benedikt Würkner, Belle II, source, L3 module, No 2 of 6 room temperature measurements using Sr90 Source (multi6) 10x

Sr90 Radioactive Source
No cooling
Module position: -32:-20(top center)
HV Turned to +40,-40V with currents. +1.23 and -1.25 µA

N+1.25V current: 0.22A
N+2.50V current: 0.56A
P+1.25V current: 0.26A
P+2.50V current: 0.71A

Events: 200000

Entry  Wed May 7 15:53:38 2014, Benedikt Würkner, Belle II, source, L3 module, No 3 of 6 room temperature measurements using Sr90 Source (multi6) 10x

Sr90 Radioactive Source
No cooling
Module position: -59:-20(top right)
HV Turned to +40,-40V with currents. +1.25 and -1.26 µA

N+1.25V current: 0.22A
N+2.50V current: 0.56A
P+1.25V current: 0.26A
P+2.50V current: 0.71A

Events: 200000

Entry  Wed May 7 16:43:37 2014, Benedikt Würkner, Belle II, source, L3 module, No 4 of 6 room temperature measurements using Sr90 Source (multi6) 10x

Sr90 Radioactive Source
No cooling
Module position: -5:-30(bottom left)
HV Turned to +40,-40V with currents. +1.25 and -1.26 µA

N+1.25V current: 0.22A
N+2.50V current: 0.57A
P+1.25V current: 0.26A
P+2.50V current: 0.72A

Events: 200000

Entry  Wed May 7 17:33:21 2014, Benedikt Würkner, Belle II, source, L3 module, No 5 of 6 room temperature measurements using Sr90 Source (multi6) 10x

Sr90 Radioactive Source
No cooling
Module position: -32:-30(bottom center)
HV Turned to +40,-40V with currents. +1.22 and -1.24 µA

N+1.25V current: 0.22A
N+2.50V current: 0.56A
P+1.25V current: 0.26A
P+2.50V current: 0.70A

Events: 200000

Entry  Wed May 7 18:20:36 2014, Benedikt Würkner, Belle II, source, L3 module, No 6 of 6 room temperature measurements using Sr90 Source (multi6) 10x

Sr90 Radioactive Source
No cooling
Module position: -57:-30(bottom right)
HV Turned to +40,-40V with currents. +1.2 and -1.22 µA

N+1.25V current: 0.22A
N+2.50V current: 0.56A
P+1.25V current: 0.26A
P+2.50V current: 0.70A

Events: 200000

Entry  Thu May 8 16:51:03 2014, Benedikt Würkner, Belle II, source, L3 module, No 1 of 6 room temperature measurements using Sr90 Source (single) 10x

Sr90 Radioactive Source
No cooling
Module position: 0:0(top left)
HV Turned to +40,-40V with currents. +1.0 and -1.0 µA

N+1.25V current: 0.22A
N+2.50V current: 0.57A
P+1.25V current: 0.26A
P+2.50V current: 0.72A

Events: 200000

Entry  Thu May 8 16:51:57 2014, Benedikt Würkner, Belle II, source, L3 module, No 2 of 6 room temperature measurements using Sr90 Source (single) 

Failed with Error: "***ERROR*** Init NECO failed,  Return code: -32761"

Restarted and it works again

Sr90 Radioactive Source
No cooling
Module position: -27:0(top center)
HV Turned to +40,-40V with currents. +0.97 and -1.0 µA

N+1.25V current: 0.22A
N+2.50V current: 0.57A
P+1.25V current: 0.26A
P+2.50V current: 0.73A

Events: 200000

Entry  Fri May 9 15:12:44 2014, Benedikt Würkner, Belle II, source, L3 module, No 2 of 6 room temperature measurements using Sr90 Source (single) 10x

Sr90 Radioactive Source
No cooling
Module position: -27:0(top center)
HV Turned to +40,-40V with currents. +0.95 and -0.98 µA

N+1.25V current: 0.22A
N+2.50V current: 0.57A
P+1.25V current: 0.26A
P+2.50V current: 0.72A

Events: 200000

Entry  Fri May 9 16:44:41 2014, Benedikt Würkner, Belle II, source, L3 module, No 3 of 6 room temperature measurements using Sr90 Source (single) 10x

Sr90 Radioactive Source
No cooling
Module position: -54:0(top center)
HV Turned to +40,-40V with currents. +1.2 and -1.22 µA

N+1.25V current: 0.22A
N+2.50V current: 0.54A
P+1.25V current: 0.26A
P+2.50V current: 0.68A

Events: 200000

Entry  Fri May 16 11:37:05 2014, Benedikt Würkner, Belle II, source, L3 module, No 4 of 6 room temperature measurements using Sr90 Source (single) 10x

Sr90 Radioactive Source
No cooling
Module position: 0:-10(bottom left)
HV Turned to +40,-40V with currents. +0.85 and -0-9 µA

N+1.25V current: 0.22A
N+2.50V current: 0.57A
P+1.25V current: 0.26A
P+2.50V current: 0.71A

Events: 200000

Entry  Fri May 16 11:38:44 2014, Benedikt Würkner, Belle II, source, L3 module, No 5 of 6 room temperature measurements using Sr90 Source (single) 10x

Sr90 Radioactive Source
No cooling
Module position: -27:-10(bottom center)
HV Turned to +40,-40V with currents. +1.05 and -1.1 µA

N+1.25V current: 0.22A
N+2.50V current: 0.55A
P+1.25V current: 0.26A
P+2.50V current: 0.67A

Events: 200000

Entry  Fri May 16 11:40:29 2014, Benedikt Würkner, Belle II, source, L3 module, No 6 of 6 room temperature measurements using Sr90 Source (single) 10x

Sr90 Radioactive Source
No cooling
Module position: -54:-10(bottom right)
HV Turned to +40,-40V with currents. +1.1 and -1.1 µA

N+1.25V current: 0.22A
N+2.50V current: 0.55A
P+1.25V current: 0.26A
P+2.50V current: 0.67A

Events: 200000
 

Entry  Mon Nov 30 19:48:05 2015, Hao Yin, Belle II, system, Hardware_Room_With_Cold_I2C,  RoomTemp_with_cold_config_ce_p.pngRoomTemp_with_cold_config_ce_n.png

ADC Cold diff to hot: 1-3 adc delay config (max 1.5 ns), FIRRun_Cold_001

with room temp i2c config

Temp = 0°C

target: CE

 

Entry  Mon Nov 30 20:06:15 2015, Hao Yin, Belle II, system, Hardware_Room_With_Cold_I2C,  RoomTemp_with_cold_config_-z_p.pngRoomTemp_with_cold_config_-z_n.png

ADC Cold diff to hot: 1-3 adc delay config (max 1.5 ns), FIRRun_Cold_001

with room temp i2c config

Temp = room

target: -Z

 