Fri Oct 8 23:43:07 2010, Christian Irmler, BELLE, run, run055, 12316, 08.10.2010 23:42:51, 09.10.2010 00:14:56
|
Belle stack w/o Origami2 after irradiation
HV=80V / 255 µA
cooling: Origami only, chiller set to -25°C,
2 suspicious APVs on n side of Micron due to burnt strip
same position as in previous run
telescope run 20197 - wrong config file @ telescope -> new run with correct settings |
Sat Oct 9 00:17:02 2010, Markus Friedl, BELLE, run, run056, 225k, 09.10.2010 00:15:53, 09.10.2010 09:54:28
|
Belle stack w/o Origami2 after irradiation
HV=80V / I_start=255 µA, I_stop=243 µA
cooling: Origami only, chiller set to -25°C,
2 suspicious APVs on n side of Micron (possibly due to burnt strip)
same position as in previous run
telescope run 20198 - now with correct telescope threshold settings
a lot of "Error detected in APV 21, subevent 2. Event discarded, reset sent." errors started at 09.10.2010 09:00:25 (after whole night without any error) |
Sat Oct 9 12:25:36 2010, Thomas Bergauer, BELLE, run, run057, 276475, 09.10.2010 12:22:21, 09.10.2010 15:08:46
|
Origami1 (n-side only) + Micron1 (non-irrad)
angle scan: 73 degree
HV=100V / 36 uA (beginning), 35uA (end of run)
cooling: chiller set to 15C
PROBLEM: wrong config file (single hit, not multi6)
telescope run 20200
Origami p-side does not receive triggers -> disabled
origami flex straight part needed to be folded due to severe rotation angle -> nanonics connector problem...
|
|
Sat Oct 9 15:29:09 2010, Markus Friedl, BELLE, run, run058, 30000, 09.10.2010 15:26:50, 09.10.2010 16:21:37
|
Origami1 (n-side only) + Micron1 (non-irrad)
angle scan: 73 - now with multi6 (previous run was single)
HV=100V / 35uA (beginning of run) 34uA (end of run)
cooling: chiller set to 15C
telescope run 20201
Origami p-side does not receive triggers -> disabled
origami flex straight part needed to be folded due to severe rotation angle -> nanonics connector problem...
|
|
Fri Apr 8 13:25:58 2011, Manfred Valentan, Micron Baby, sensors and modules, , , ,
|
TRK1: B2Micron_2826_1_baby2
TRK2: B2Micron_2826_1_baby3
DUT3: B2Micron_2826_7_baby2 (common p-stop)
DUT4: B2Micron_2826_19_baby4 (combined p-stop)
DUT5: B2Micron_2826_19_baby3 (atoll p-stop)
TRK6: B2Micron_2826_1_baby4
TRK7: B2Micron_2826_19_baby2
GEW: B2Micron_2826_7_baby4
not used: B2Micron_2826_7_baby3 |
Tue May 17 19:08:29 2011, Manfred Valentan, Micron Baby, sensors for source test with intermediate strips, , , ,
|
Every second bond of these sensors will be removed to mimic the effect of an intermediate strip. For comparison, this is done both with irradiated and unirradiated sensors.
DUT3: B2Micron_2826_7_baby2, common p-stop, irradiated
DUT4: B2Micron_2826_19_baby4, combined p-stop, irradiated
DUT5: B2Micron_2826_19_baby3, atoll p-stop, irradiated (Wolfgang T. took this sensor to the SEM, it's destroyed by now)
TRK6: B2Micron_2826_1_baby4, combined p-stop, unirradiated
TRK7: B2Micron_2826_19_baby2, common p-stop, unirradiated
B2Micron_2826_7_baby3, atoll p-stop, unirradiated
|
Tue Sep 28 16:50:31 2010, Thomas Bergauer, common, serial connection chiller, , , ,
|
Der chiller braucht ein gekreuztes serielles Kabel!! |
|