ID |
Date |
Author |
Project |
Subject |
Run Number |
Events |
StartTime |
EndTime |
61
|
Sat Oct 2 18:36:57 2010 |
Markus Friedl | ITE | run | run028 | 162263 | 02.10.2010 18:34:06 | 03.10.2010 08:11:13 |
6 x SILC + 2 x ITE
HV=100V
telescope position: 119
short break ~2.10.2010 20:00 (beam off for a couple of minutes for access by marcos)
telescope run 20126 |
48
|
Thu Sep 30 23:09:29 2010 |
Thomas Bergauer | Micron Baby | run | run019 | 162000 | 30.09.2010 23:08:03 | 01.10.2010 9:33:38 |
micron baby: 4 DUTs
HV=100V I=12uA I_stop=15.4uA
Beam on
45° rotation
telescope run 20115 (after JTAG programming of telescope chips due to config errors)
|
Attachment 1: run019_sensordisplay1.jpg
|
 |
Attachment 2: Eudet-X_Cluster_Corr-run020115.jpg
|
 |
Attachment 3: Eudet-Num_Clusters-run020115.jpg
|
 |
49
|
Fri Oct 1 09:33:45 2010 |
Immanuel Gfall | common | Eudet Hit Correlation, Num Clusters Screenshots | 20115 | 161135 | 9:30 | 9:30 |

[IMG]49/2[/IMG] |
Attachment 2: Eudet-X_Cluster_Corr-run020115.jpg
|
 |
55
|
Sat Oct 2 09:57:20 2010 |
Thomas Bergauer | CNM | run | run025 | 150k | 02.10.2010 09:54:28 | 02.10.2010 14:21:20 |
CMN-SILC stack
HV_SILC (Katze)=75V, 0.25uA (6 SiLC)
HV_CNM (Keithley)=30V, CNM_1=84nA, CNM_2=4.3uA
new position along the CNM strips (XY-table to 143/76/187.5)
telescope run 20122
|
Attachment 1: run20122.png
|
 |
Attachment 2: slowcontrol_cnm.png
|
 |
Attachment 3: cnm_2d_1.png
|
 |
Attachment 4: cnm_2d_2.png
|
 |
28
|
Tue Sep 28 23:24:35 2010 |
Markus Friedl | BELLE | run | run007 | 1460 | 28.09.2010 23:23:29 | 28.09.2010 23:27:54 |
Belle stack: Origami-Micron-Micron-Origami
HV=100V, Ibias=38A, beam on
cooling: Origami only, chiller: -25C, Module: -10C
modified I2C settings for cooled modules
centered again
telescope run 20082
sensor config bug (only relevant for sensor display) found -> stopped |
29
|
Tue Sep 28 23:29:50 2010 |
Markus Friedl | BELLE | run | run008 | 140000 | 28.09.2010 23:29:01 | 29.09.2010 09:36:58 |
Belle stack: Origami-Micron-Micron-Origami
HV=100V, Ibias=38A, beam on
cooling: Origami only, chiller: -25C, Module: -10C
modified I2C settings for cooled modules
centered again
telescope run 20083 |
Attachment 1: run008_sensordisplay.jpg
|
 |
Attachment 2: run008_sensordisplay2.jpg
|
 |
102
|
Fri Oct 8 23:43:07 2010 |
Christian Irmler | BELLE | run | run055 | 12316 | 08.10.2010 23:42:51 | 09.10.2010 00:14:56 |
Belle stack w/o Origami2 after irradiation
HV=80V / 255 µA
cooling: Origami only, chiller set to -25°C,
2 suspicious APVs on n side of Micron due to burnt strip
same position as in previous run
telescope run 20197 - wrong config file @ telescope -> new run with correct settings |
35
|
Wed Sep 29 16:28:17 2010 |
Thomas Bergauer | common | EUDET telescope pedestal run | 20089 | 122850 (events built) | 29.09.2010 16:27 | 29.09.2010 17:33 |
EUDAQ config eudet2-test-6boards-in-2crates
rate: ~745.5 Hz
|
18
|
Tue Sep 28 03:06:31 2010 |
Markus Friedl | BELLE | run | run001 | 119516 | 28.09.2010 03:06:04 | 28.09.2010 12:18:33 |
first run with complete Belle stack: 2 x Origami, 2 x Micron
HV=100V I=49uA
beam is on
chiller is set to 15°C (module temperature is ~17°C)
both Origamis and both Microns are serially cooled
position:centered in both axes 55 / 30 / 277.5
telescope: run20076 (in zero suppressed mode) |
36
|
Wed Sep 29 17:38:15 2010 |
Thomas Bergauer | common | EUDET telescope pedestal run | 20097 | 118125 | 29.09.2010 17:37 | 29.09.2010 18:09 |
new pedestal run of EUDET telescope with new JTAG settings of plane 3 (sensor 11) [according to Ingrid]:
VRef2 98
VRef1A 87
VRef1B 143
VRef1C 108
VRef1D 153
|
41
|
Thu Sep 30 00:56:32 2010 |
Markus Friedl | Micron Baby | run | run014 | 100k | 30.09.2010 00:55:30 | 30.09.2010 11:20:02 |
micron baby stack
HV=100V I=11.0uA
I_end = 16uA
Beam on
telescope position: 105/40.5/7.5
telescope run 20108 |
113
|
Thu Mar 31 11:05:39 2011 |
Manfred Valentan | Micron Baby | DUT5, source run, post irrad, pre zupf | run100 | 100k | 31.03.2011 11:02:17 | 31.03.2011 11:15:41 |
DUT5
cal-file: micronbaby_multi6_irrad_run100.cfg
HV=+-50V
90Sr 1mCi ~centered
I_start=104µA
I_end=111µA
|
115
|
Thu Mar 31 12:24:24 2011 |
Manfred Valentan | Micron Baby | DUT4, source run, post irrad, pre zupf | run101 | 100k | 31.03.2011 12:23:00 | 31.03.2011 12:36:48 |
DUT4
config file: micronbaby_multi6_irrrad_run101.cfg
HV=+-50V
90Sr 1mCi ~centered
I_start=128µA
I_end=155µA
|
118
|
Mon Apr 11 18:37:28 2011 |
Manfred Valentan | Micron Baby | DUT5, source run, post irrad, pre zupf | run104 | 100k | 11.04.2011 18:25:28 | 11.04.2011 18:32:11 |
DUT5 (atoll)
config file: D:/cern10/micronbaby/data_lab/DUT5/micronbaby_multi6_irrrad.cfg
HV=+-50V
90Sr 1mCi ~centered
I_start=50µA
I_end=95µA
|
121
|
Wed May 11 16:28:07 2011 |
Manfred Valentan | Micron Baby | DUT5, source run, post irrad, post zupf | run106 | 100k | 11.05.2011 16:00 | 11.05.2011 16:25 |
DUT5 (atoll)
config file: D:/cern10/micronbaby/data_lab/DUT5/micronbaby_multi6_irrrad.cfg
HV=+-50V
90Sr 1mCi ~centered
I_start=52µA
I_end=120µA
DAQ modified so that only strips marked as good are considered when calculating the common mode correction. |
122
|
Wed May 11 16:28:20 2011 |
Manfred Valentan | Micron Baby | DUT5, source run, post irrad, post zupf | cal106 | 100k | 11.05.2011 16:00 | 11.05.2011 16:25 |
DUT5 (atoll)
config file: D:/cern10/micronbaby/data_lab/DUT5/micronbaby_multi6_irrrad.cfg
HV=+-50V
90Sr 1mCi ~centered
I_start=52µA
I_end=120µA
DAQ modified so that only strips marked as good are considered when calculating the common mode correction. |
124
|
Tue May 17 13:08:08 2011 |
Manfred Valentan | Micron Baby | TRK1, source run, unirrad, prezupf | run107 | 100k | 17.05.2011 12:39:34 | 17.05.2011 13:03:05 |
DUT candidate common p-stop
TRK1: B2Micron_2826_1_baby2 (common p-stop)
config file: D:/cern10/micronbaby/data_lab/TRK1/prezupf/micronbaby_multi6.cfg
HV=+-50V
90Sr 1mCi ~centered
I_start=1.1µA
I_end=2µA
n-side: half of zones 1 and 4 obsolete due to noisy strips, rest is okay. DUT Candidate |
Attachment 1: run107_all_hits_apv4.png
|
 |
Attachment 2: run107_all_hits_apv5.png
|
 |
Attachment 3: cal107_pednoise_apv4.png
|
 |
Attachment 4: cal107_pednoise_apv5.png
|
 |
126
|
Tue May 17 14:01:45 2011 |
Manfred Valentan | Micron Baby | TRK2, source run, unirrad, prezupf | run108 | 100k | 17.05.2011 13:28:48 | 17.05.2011 13:51:16 |
No DUT candidate (atoll p-stop)
TRK2: B2Micron_2826_1_baby3 (atoll p-stop)
config file: D:/cern10/micronbaby/data_lab/TRK2/prezupf/micronbaby_multi6.cfg
HV=+-50V
90Sr 1mCi ~centered
I_start=1.3µA
I_end=2µA
n-side: APV4 most strips unusable, no DUT candidate |
Attachment 1: run108_all_hits_apv4.png
|
 |
Attachment 2: run108_all_hits_apv5.png
|
 |
Attachment 3: cal108_pednoise_apv4.png
|
 |
Attachment 4: cal108_pednoise_apv5.png
|
 |
128
|
Tue May 17 14:37:23 2011 |
Manfred Valentan | Micron Baby | TRK6, source run, unirrad, prezupf | run109 | 100k | 17.05.2011 14:34:32 | 17.05.2011 15:00:02 |
DUT candidate combined p-stop
TRK6: B2Micron_2826_1_baby4 (combined p-stop)
config file: D:/cern10/micronbaby/data_lab/TRK6/prezupf/micronbaby_multi6.cfg
HV=+-50V
90Sr 1mCi ~centered
I_start=1.9µA
I_end=2.4µA
n-side zone 2 several unusable strips, zone 3 & 4 32 strips each suspicious. Nevertheless DUT candidate! |
Attachment 1: run109_all_hits_apv4.png
|
 |
Attachment 2: run109_all_hits_apv5.png
|
 |
Attachment 3: cal109_pednoise_apv4.png
|
 |
Attachment 4: cal109_pednoise_apv5.png
|
 |
130
|
Tue May 17 17:43:24 2011 |
Manfred Valentan | Micron Baby | TRK7, source run, unirrad, prezupf | run110 | 100k | 17.05.2011 15:39:37 | 17.05.2011 16:02:54 |
DUT candidate common p-stop
TRK7: B2Micron_2826_19_baby2 (common p-stop)
config file: D:/cern10/micronbaby/data_lab/TRK7/prezupf/micronbaby_multi6.cfg
HV=+-50V
90Sr 1mCi ~centered
I_start=1.8µA
I_end=2.1µA
n-side: highest 32 strips of APV5 noisy = half of zone 4. Rest okay, DUT candidate |
Attachment 1: run110_all_hits_apv4.png
|
 |
Attachment 2: run110_all_hits_apv5.png
|
 |
Attachment 3: cal110_pednoise_apv4.png
|
 |
Attachment 4: cal110_pednoise_apv5.png
|
 |