Home CMS Production Clean room MedAustron HEPHY testbeams old
testbeam-rd50-desy-Oct2024 testbeam-RD50-DESY-Apr2024 testbeam-RD50-DESY-Jul2023 testbeam-RD50-CERN-Oct2022 testbeam-DESY-Oct2017 testbeam-DESY-April2017 testbeam-SPS2015 testbeam-SPS2014 testbeam-DESY14 testbeam-SPS12 testbeam-SPS11 testbeam-SPS10 testbeam-SPS09 testbeam-SPS08 LP-TPC
  HEPHY logbook for testbeam at SPS September/October 2010, Page 1 of 8  Not logged in ELOG logo
Entry  Tue Oct 5 10:23:09 2010, Manfred Valentan, BELLE, run, run47, 30000, 05.10.2010 10:21:20, 05.10.2010  

angle scan Origami 1 and  Micron wedge 1
HV = 100V
chiller set to 14°C

angle: 0°
axis 1: 92.4 (horizontal)
axis 2: 35 (vertical)
axis 3: +7.5 (rotary table)

telescope run: 20153

Entry  Tue Oct 5 10:20:30 2010, Manfred Valentan, BELLE, run, run46, 350000, 05.10.2010 01:00:00, 05.10.2010 10:16:07 

angle scan Origami 1 and  Micron wedge 1
HV = 100V
chiller set to 14°C

angle: 15°
axis 1: 92.4 (horizontal)
axis 2: 35 (vertical)
axis 3: -7.5 (rotary table)

telescope run: 20152

Entry  Mon Oct 4 23:59:05 2010, Manfred Valentan, BELLE, run, run45, , 04.10.2010 ??:??:??, 05.10.2010 00:56:48 

angle scan Origami 1 and  Micron wedge 1
HV = 100V
chiller set to 14°C

angle: 45°
axis 1: 92.4 (horizontal)
axis 2: 35 (vertical)
axis 3: -37.5 (rotary table)

telescope run: 20151

Entry  Mon Oct 4 23:12:16 2010, Manfred Valentan, BELLE, run, run44, 30000, 04.10.2010 23:?:?, 04.10.2010 23.57.23 

angle scan Origami 1 and  Micron wedge 1
HV = 100V
chiller set to 14°C

angle: 30°
axis 1: 92.4 (horizontal)
axis 2: 35 (vertical)
axis 3: -22.5 (rotary table)

telescope run: 20150

Entry  Mon Oct 4 21:01:25 2010, Christian Irmler, BELLE, run, run43, 30000, 04.10.2010 20:59:48, 04.10.2010 23:04:24 

angle scan Origami 1 and  Micron wedge 1
HV = 100V
chiller set to 14°C

angle: 60°
axis 1: 92.4 (horizontal)
axis 2: 35 (vertical)
axis 3: -52.5 (rotary table)

telescope run: 20149

Entry  Thu May 19 11:04:04 2011, Manfred Valentan, Micron Baby, DUT4, source run, irrad, postzupf, run119, 100k, 19.05.2011 10:08:14, 19.05.2011 10:29:02 6x

DUT4: B2Micron_2826_19_baby4 (combined p-stop)
config file: D:/cern10/micronbaby/data_lab/DUT4/postzupf/micronbaby_multi6_irrad_postzupf.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=141µA
I_end=167µA

Every second bond zupfed...

Entry  Thu May 19 09:44:39 2011, Manfred Valentan, Micron Baby, DUT3, source run, irrad, postzupf, run118, 100k, 19.05.2011 09:07:33, 19.05.2011 09:27:21 6x

DUT3: B2Micron_2826_7_baby2 (common p-stop)
config file: D:/cern10/micronbaby/data_lab/DUT3/postzupf/micronbaby_multi6_irrad_postzupf.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=18.0µA
I_end=22.6µA

Every second bond zupfed

Entry  Wed May 18 16:33:29 2011, Manfred Valentan, Micron Baby, B2Micron_2826_7_baby3, source run, unirrad, prezupf, run117, 100k, 18.05.2011 15:59:39, 18.05.2011 16:21:24 6x

B2Micron_2826_7_baby3 (atoll p-stop)
config file: D:/cern10/micronbaby/data_lab/7baby3/postzupf/micronbaby_multi6_postzupf.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=1.3µA
I_end=1.7µA

Every second bond zupfed...

Entry  Wed May 18 15:38:29 2011, Manfred Valentan, Micron Baby, TRK7, source run, unirrad, postzupf, run116, 100k, 18.05.2011 15:02:05, 18.05.2011 15:24:07 6x

TRK7: B2Micron_2826_19_baby2 (common p-stop)
config file: D:/cern10/micronbaby/data_lab/TRK7/postzupf/micronbaby_multi6_postzupf.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=1.7µA
I_end=2.3µA

every second bond zupfed...

Entry  Wed May 18 14:42:49 2011, Manfred Valentan, Micron Baby, TRK6, source run, unirrad, postzupf, run115, 100k, 18.05.2011 13:53:54, 18.05.2011 13:53:54 6x

TRK6: B2Micron_2826_1_baby4 (combined p-stop)
config file: D:/cern10/micronbaby/data_lab/TRK6/postzupfzupf/micronbaby_multi6_postzupf.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=1.7µA
I_end=1.8µA

Every second bond zupfed. However, there was a mistake and the first ~60 channels are shifted by 1, see attachment!

Entry  Thu May 19 11:06:41 2011, Manfred Valentan, Micron Baby, DUT4, source run, irrad, prezupf, run114, 100k, 18.05.2011 11:57:13, 18.05.2011 12:13:12 cal114_pednoise_apv4.pngcal114_pednoise_apv5.png

DUT4: B2Micron_2826_19_baby4 (combined p-stop)
config file: D:/cern10/micronbaby/data_lab/DUT4/prezupf/micronbaby_multi6_irrad.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=118µA
I_end=168µA

keine hitmaps gespeichert...

Entry  Wed May 18 10:38:07 2011, Manfred Valentan, Micron Baby, DUT3, source run, irrad, prezupf, run113, 100k, 18.05.2011 10:03:15, 18.05.2011 10:25:14 run113_all_hits_apv4.pngrun113_all_hits_apv5.pngcal113_pednoise_apv4.pngcal113_pednoise_apv5.png

DUT3: B2Micron_2826_7_baby2 (common p-stop)
config file: D:/cern10/micronbaby/data_lab/DUT3/prezupf/micronbaby_single_irrad.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=20µA
I_end=26µA

Entry  Tue May 17 18:53:33 2011, Manfred Valentan, Micron Baby, B2Micron_2826_7_baby3, source run, unirrad, prezupf, run112, 100k, 17.05.2011 19:00:21, 17.05.2011 19:22:45 run112_all_hits_apv4.pngrun112_all_hits_apv5.pngcal112_pednoise_apv4.pngcal112_pednoise_apv5.png

DUT candidate atoll p-stop

B2Micron_2826_7_baby3 (atoll p-stop)
config file: D:/cern10/micronbaby/data_lab/7baby3/prezupf/micronbaby_multi6.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=1.4µA
I_end=1.8µA

n-side: looks very good except for a few strips in the middle of APV5 -> DUT candidate

Entry  Tue May 17 18:33:07 2011, Manfred Valentan, Micron Baby, GEW8, source run, unirrad, prezupf, run111, 100k, 17.05.2011 18:01:41, 17.05.2011 18:23:24 run111_all_hits_apv4.pngrun111_all_hits_apv5.pngcal111_pednoise_apv4.pngcal111_pednoise_apv5.png

NOT usable as DUT

GEW8: B2Micron_2826_7_baby4 (combined p-stop)
config file: D:/cern10/micronbaby/data_lab/GEW8/prezupf/micronbaby_multi6.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=1.6µA
I_end=2.2µA

n-side: most of zones 1 and 3 unusable

Entry  Tue May 17 17:43:24 2011, Manfred Valentan, Micron Baby, TRK7, source run, unirrad, prezupf, run110, 100k, 17.05.2011 15:39:37, 17.05.2011 16:02:54 run110_all_hits_apv4.pngrun110_all_hits_apv5.pngcal110_pednoise_apv4.pngcal110_pednoise_apv5.png

DUT candidate common p-stop

TRK7: B2Micron_2826_19_baby2 (common p-stop)
config file: D:/cern10/micronbaby/data_lab/TRK7/prezupf/micronbaby_multi6.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=1.8µA
I_end=2.1µA

n-side: highest 32 strips of APV5 noisy = half of zone 4. Rest okay, DUT candidate

Entry  Tue May 17 14:37:23 2011, Manfred Valentan, Micron Baby, TRK6, source run, unirrad, prezupf, run109, 100k, 17.05.2011 14:34:32, 17.05.2011 15:00:02 run109_all_hits_apv4.pngrun109_all_hits_apv5.pngcal109_pednoise_apv4.pngcal109_pednoise_apv5.png

DUT candidate combined p-stop

TRK6: B2Micron_2826_1_baby4 (combined p-stop)
config file: D:/cern10/micronbaby/data_lab/TRK6/prezupf/micronbaby_multi6.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=1.9µA
I_end=2.4µA

n-side zone 2 several unusable strips, zone 3 & 4 32 strips each suspicious. Nevertheless DUT candidate!

Entry  Tue May 17 14:01:45 2011, Manfred Valentan, Micron Baby, TRK2, source run, unirrad, prezupf, run108, 100k, 17.05.2011 13:28:48, 17.05.2011 13:51:16 run108_all_hits_apv4.pngrun108_all_hits_apv5.pngcal108_pednoise_apv4.pngcal108_pednoise_apv5.png

No DUT candidate (atoll p-stop)

TRK2: B2Micron_2826_1_baby3 (atoll p-stop)
config file: D:/cern10/micronbaby/data_lab/TRK2/prezupf/micronbaby_multi6.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=1.3µA
I_end=2µA

n-side: APV4 most strips unusable, no DUT candidate

Entry  Tue May 17 13:08:08 2011, Manfred Valentan, Micron Baby, TRK1, source run, unirrad, prezupf, run107, 100k, 17.05.2011 12:39:34, 17.05.2011 13:03:05 run107_all_hits_apv4.pngrun107_all_hits_apv5.pngcal107_pednoise_apv4.pngcal107_pednoise_apv5.png

DUT candidate common p-stop

TRK1: B2Micron_2826_1_baby2 (common p-stop)
config file: D:/cern10/micronbaby/data_lab/TRK1/prezupf/micronbaby_multi6.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=1.1µA
I_end=2µA

n-side: half of zones 1 and 4 obsolete due to noisy strips, rest is okay. DUT Candidate

Entry  Wed May 11 16:28:07 2011, Manfred Valentan, Micron Baby, DUT5, source run, post irrad, post zupf, run106, 100k, 11.05.2011 16:00, 11.05.2011 16:25 

DUT5 (atoll)
config file: D:/cern10/micronbaby/data_lab/DUT5/micronbaby_multi6_irrrad.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=52µA
I_end=120µA

 DAQ modified so that only strips marked as good are considered when calculating the common mode correction.

Entry  Mon Apr 11 18:37:28 2011, Manfred Valentan, Micron Baby, DUT5, source run, post irrad, pre zupf, run104, 100k, 11.04.2011 18:25:28, 11.04.2011 18:32:11 

DUT5 (atoll)
config file: D:/cern10/micronbaby/data_lab/DUT5/micronbaby_multi6_irrrad.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=50µA
I_end=95µA

 

ELOG V3.1.5-fc6679b