77
|
Sat Jun 21 16:30:44 2025 |
Julia Müllner | 205.2 MeV systemtest | 3 | TOF-Cal GSI | 205.2 MeV/u | Low Flux High | Helium | Unknown | 16:38 | 16:57 | 19 | | | |
Threshold set to 5000
dead channels on last LGAD, changed trigger from LGAD 12 to LGAD 11 |
76
|
Sat Jun 21 15:43:47 2025 |
Julia Müllner | 800MeV systemtest | 2 | TOF-Cal GSI | 800 MeV | Low Flux High | Proton | Good | 16:02 | 16:29 | 27 | | | |
Threshold set to 1000
Factor 4 chosen for future runs, 80 GB in total recorded in this run |
75
|
Sat Jun 21 14:30:29 2025 |
Julia Müllner | 800MeV systemtest | 1 | TOF-Cal GSI | 800 MeV | Low Flux High | Proton | Good | 14:45 | 15:42 | 57 | | | |
first spills were done with low flux medium, changed to high at 14:49
again changed to medium at 14:55
again changed to high at 15:00
30 MB pro event builder
last few spills were with the wrong threshold
thresholds were set to 2000 instead of 1000 |
74
|
Sat Mar 15 21:29:28 2025 |
RD50 Generic User | Slab and rate testing (with Michael) | 13 | Sandwich-iCT setup(GSI)+RaMon | 205.2 MeV/u | Low Flux High | Helium | Good | 20:44 | 21:22 | 38 | | Plastic Water Slabs | 4cm |
totzeit optimieren und deadtime finden
hex wert mal 10ns deadtime
13 ab file 10 besat effort
ab 52 wieder ohne best effort
600 |
73
|
Sat Mar 15 21:25:18 2025 |
RD50 Generic User | Slab Test Water 5cm | 14 | Sandwich-iCT setup(GSI)+RaMon | 205.2 MeV/u | Low Flux High | Helium | Good | 21:26 | 21:50 | 24 | | Plastic Water Slabs | 5cm |
RaMon rate: 4.5MHz |
72
|
Sat Mar 15 19:37:42 2025 |
RD50 Generic User | Slab Test Water 4cm | 12 | Sandwich-iCT setup(GSI)+RaMon | 205.2 MeV/u | Low Flux Medium | Helium | | 19:39 | 20:41 | 62 | | Plastic Water Slabs | 4cm |
RaMon Rate 1.4MHz |
71
|
Sat Mar 15 18:37:36 2025 |
RD50 Generic User | Slab Test Water 3cm | 11 | Sandwich-iCT setup(GSI)+RaMon | 205.2 MeV/u | Low Flux Medium | Helium | | 18:41 | 19:28 | 47 | | Plastic Water Slabs | 3cm |
RaMon Rate 1.5MHz |
70
|
Sat Mar 15 17:30:18 2025 |
RD50 Generic User | Slab Test Water 2cm | 10 | Sandwich-iCT setup(GSI)+RaMon | 205.2 MeV/u | Low Flux Medium | Helium | | 17:30 | 18:24 | 54 | | Plastic Water Slabs | 2cm |
RaMon Rate 1.5 MHz |
69
|
Sat Mar 15 16:50:39 2025 |
RD50 Generic User | Slab Test Water 1cm | 9 | Sandwich-iCT setup(GSI)+RaMon | 205.2 MeV/u | Low Flux Medium | Helium | Good | 16:35 | 17:23 | 48 | | Plastic Water Slabs | 1cm |
RaMon Rate: 1.3MHz |
68
|
Sat Mar 15 15:26:08 2025 |
RD50 Generic User | Rate testing | 8 | Sandwich-iCT setup(GSI)+RaMon | 205.2 MeV/u | Low Flux Medium | Helium | | 15:31 | 16:30 | 59 | | | |
RaMon Rate: 1.3MHz
Comments:
Tested different evenbuilder settings which made things worse, probably switch is the bottleneck, changed rate settings inbetween, requires sergey |
67
|
Sat Mar 15 15:02:39 2025 |
RD50 Generic User | Rate testing | 7 | Sandwich-iCT setup(GSI)+RaMon | 205.2 MeV/u | Low Flux High | Helium | Good | 15:04 | 15:25 | 21 | | | |
RaMon Rate 4.5MHz
Comments:
From File 7 until 21 optional false, and from 22 true (everything deleted) |
66
|
Sat Mar 15 13:15:25 2025 |
RD50 Generic User | Slab Measurement 205MeV | 6 | Sandwich-iCT setup(GSI)+RaMon | 205.2 MeV/u | Low Flux Medium | Helium | Good | 13:27 | 14:49 | 82 | | | |
RaMon Rate 1.3 MHz
Comments:
We entered the room and connected the second LAN Port to the ReadOut-PC. The Eventbuilder-parameter got changed. |
65
|
Sat Mar 15 11:46:28 2025 |
RD50 Generic User | Slab Measurement 145MeV | 5 | Sandwich-iCT setup(GSI)+RaMon | 145.4 MeV | Low Flux Medium | Proton | Good | 11:48 | 13:13 | 85 | | | |
RaMon Rate 600kHz
front sensors 200mu: ~1kHz pro Streifen
Comments:
change of the trigger settings inbetween, we saw inefficency on all detectors |
64
|
Sat Mar 15 10:07:28 2025 |
RD50 Generic User | Alignment Run 252MeV | 4 | Sandwich-iCT setup(GSI)+RaMon | 252.7 MeV | Low Flux Medium | Proton | Good | 10:07 | 11:43 | 96 | | | |
RaMon Rate 1MHz
front sensors 200mu: ~10-25kHz pro Streifen |
63
|
Sat Mar 15 09:51:23 2025 |
RD50 Generic User | Alignment Run 800MeV | 3 | Sandwich-iCT setup(GSI)+RaMon | 800 MeV | Low Flux High | Proton | Unknown | 9:44 | 10:03 | 19 | | | |
RaMon Rate 2MHz
front sensors 200mu: ~50-100kHz pro Streifen
Comments:
Factor 10 lost more data than measured
|
62
|
Sat Mar 15 08:58:26 2025 |
RD50 Generic User | Alignment Run 800MeV | 1 | Sandwich-iCT setup(GSI)+RaMon | 800 MeV | Low Flux Low | Proton | Good | 07:40 | 08:20 | 40 | | | |
RaMon Rate not measured
front sensors 200mu: ~40-60Hz pro Streifen
Comments:
stopped run due to too low rate
|
61
|
Sat Mar 15 08:52:52 2025 |
RD50 Generic User | Alignment Run 800MeV | 2 | Sandwich-iCT setup(GSI)+RaMon | 800 MeV | Low Flux Medium | Proton | Good | 08:22 | 9:40 | 78 | | | |
RaMon Rate 400kHz
front sensors 200mu: ~10-20kHz pro Streifen
Comments:
In between run, sensors were shifted by 2mm to the left (left and right in online DAQ is mirrored)
|
60
|
Sat Mar 15 05:42:14 2025 |
elog-producer | test 8b10b | 246 | TJ-Monopix2 | 800 MeV | Low Flux Medium | Proton | Unknown | 15.03.2025 05:38:42 | 15.03.2025 05:42:41 | 3 | HV | None | |
automatic log for run 246
Comment:
BDAQ Board to Monopix connection swapped - and now the W02R01 chip works fine with full rate |
59
|
Sat Mar 15 05:27:08 2025 |
elog-producer | test 8b10b | 245 | TJ-Monopix2 | 800 MeV | Low Flux Medium | Proton | Unknown | 15.03.2025 05:26:52 | 15.03.2025 05:27:34 | 0 | HV | None | |
automatic log for run 245
Comment:
test with swapped ethernet ports and bsaq IP adresses |
58
|
Sat Mar 15 05:12:19 2025 |
elog-producer | test 8b10b | 244 | TJ-Monopix2 | 800 MeV | Low Flux Medium | Proton | Unknown | 15.03.2025 05:08:03 | 15.03.2025 05:12:46 | 4 | HV | None | |
automatic log for run 244
Comment: