  Testbeams at MedAustron in 2025, Page 1 of 4  Not logged in ELOG logo
ID Date Author Subject Run Number DUT Beam Energy Rate Setting Particle Type Data t-Start t-Stop Duration_min Frontend Phantom Phantom thickness
  76   Wed Mar 19 11:27:46 2025 bpjust testing            
This is message text
  75   Wed Mar 19 11:23:53 2025 bpjust testing            
This is message text
  74   Sat Mar 15 21:29:28 2025 RD50 Generic UserSlab and rate testing (with Michael)13Sandwich-iCT setup(GSI)+RaMon205.2 MeV/uLow Flux HighHeliumGood20:4421:2238 Plastic Water Slabs4cm

totzeit optimieren und deadtime finden

hex wert mal 10ns deadtime

13 ab file 10 besat effort
ab 52 wieder ohne best effort

600

  73   Sat Mar 15 21:25:18 2025 RD50 Generic UserSlab Test Water 5cm14Sandwich-iCT setup(GSI)+RaMon205.2 MeV/uLow Flux HighHeliumGood21:2621:5014 Plastic Water Slabs5cm

RaMon rate: 4.5MHz

  72   Sat Mar 15 19:37:42 2025 RD50 Generic UserSlab Test Water 4cm12Sandwich-iCT setup(GSI)+RaMon205.2 MeV/uLow Flux MediumHelium 19:3920:4162 Plastic Water Slabs4cm

RaMon Rate 1.4MHz

  71   Sat Mar 15 18:37:36 2025 RD50 Generic UserSlab Test Water 3cm11Sandwich-iCT setup(GSI)+RaMon205.2 MeV/uLow Flux MediumHelium 18:4119:2847 Plastic Water Slabs3cm

RaMon Rate 1.5MHz

  70   Sat Mar 15 17:30:18 2025 RD50 Generic UserSlab Test Water 2cm10Sandwich-iCT setup(GSI)+RaMon205.2 MeV/uLow Flux MediumHelium 17:3018:2454 Plastic Water Slabs2cm

RaMon Rate 1.5 MHz

  69   Sat Mar 15 16:50:39 2025 RD50 Generic UserSlab Test Water 1cm9Sandwich-iCT setup(GSI)+RaMon205.2 MeV/uLow Flux MediumHeliumGood16:3517:2348 Plastic Water Slabs1cm

RaMon Rate: 1.3MHz

  68   Sat Mar 15 15:26:08 2025 RD50 Generic UserRate testing8Sandwich-iCT setup(GSI)+RaMon205.2 MeV/uLow Flux MediumHelium 15:3116:3059   

RaMon Rate: 1.3MHz

Comments:

Tested different evenbuilder settings which made things worse, probably switch is the bottleneck, changed rate settings inbetween, requires sergey

  67   Sat Mar 15 15:02:39 2025 RD50 Generic UserRate testing7Sandwich-iCT setup(GSI)+RaMon205.2 MeV/uLow Flux HighHeliumGood15:0415:2521   

RaMon Rate 4.5MHz

Comments:

From File 7 until 21 optional false, and from 22 true (everything deleted)

  66   Sat Mar 15 13:15:25 2025 RD50 Generic UserSlab Measurement 205MeV6Sandwich-iCT setup(GSI)+RaMon205.2 MeV/uLow Flux MediumHeliumGood13:2714:4982   

RaMon Rate 1.3 MHz

Comments:

We entered the room and connected the second LAN Port to the ReadOut-PC. The Eventbuilder-parameter got changed.

  65   Sat Mar 15 11:46:28 2025 RD50 Generic UserSlab Measurement 145MeV5Sandwich-iCT setup(GSI)+RaMon145.4 MeVLow Flux MediumProtonGood11:4813:1385   

RaMon Rate 600kHz

front sensors 200mu: ~1kHz pro Streifen

Comments:

change of the trigger settings inbetween, we saw inefficency on all detectors

  64   Sat Mar 15 10:07:28 2025 RD50 Generic UserAlignment Run 252MeV4Sandwich-iCT setup(GSI)+RaMon252.7 MeVLow Flux MediumProtonGood10:0711:4396   

RaMon Rate 1MHz

front sensors 200mu: ~10-25kHz pro Streifen

  63   Sat Mar 15 09:51:23 2025 RD50 Generic UserAlignment Run 800MeV3Sandwich-iCT setup(GSI)+RaMon800 MeVLow Flux HighProtonUnknown9:4410:0319   

RaMon Rate 2MHz

front sensors 200mu: ~50-100kHz pro Streifen

Comments:

Factor 10 lost more data than measured

 

  62   Sat Mar 15 08:58:26 2025 RD50 Generic UserAlignment Run 800MeV1Sandwich-iCT setup(GSI)+RaMon800 MeVLow Flux LowProtonGood07:4008:2040   

RaMon Rate not measured

front sensors 200mu: ~40-60Hz pro Streifen

Comments:

stopped run due to too low rate

 

  61   Sat Mar 15 08:52:52 2025 RD50 Generic UserAlignment Run 800MeV2Sandwich-iCT setup(GSI)+RaMon800 MeVLow Flux MediumProtonGood08:229:4078   

RaMon Rate 400kHz

front sensors 200mu: ~10-20kHz pro Streifen

Comments:

In between run, sensors were shifted by 2mm to the left (left and right in online DAQ is mirrored)

 

  60   Sat Mar 15 05:42:14 2025 elog-producertest 8b10b246TJ-Monopix2800 MeVLow Flux MediumProtonUnknown15.03.2025 05:38:4215.03.2025 05:42:413HVNone 
automatic log for run 246
Comment:
BDAQ Board to Monopix connection swapped - and now the W02R01 chip works fine with full rate
  59   Sat Mar 15 05:27:08 2025 elog-producertest 8b10b245TJ-Monopix2800 MeVLow Flux MediumProtonUnknown15.03.2025 05:26:5215.03.2025 05:27:340HVNone 
automatic log for run 245
Comment:
test with swapped ethernet ports and bsaq IP adresses
  58   Sat Mar 15 05:12:19 2025 elog-producertest 8b10b244TJ-Monopix2800 MeVLow Flux MediumProtonUnknown15.03.2025 05:08:0315.03.2025 05:12:464HVNone 
automatic log for run 244
Comment:
both chips, more stat with small window on W02R01
  57   Sat Mar 15 05:06:44 2025 elog-producertest 8b10b243TJ-Monopix2800 MeVLow Flux MediumProtonUnknown15.03.2025 05:05:5915.03.2025 05:07:101HVNone 
automatic log for run 243
Comment: