Home CMS Production Clean room MedAustron HEPHY testbeams old
testbeam-MA2018 testbeam-MA2019 testbeam-MA2020 testbeam-MA2021 testbeam-MA2023 testbeam-MA2024 testbeam-MA2025
  Testbeams at MedAustron in 2025, Page 15 of 19  Not logged in ELOG logo
ID Date Author Subject Run Number DUT Beam Energy Rate Setting Particle Type Data t-Start t-Stop Duration_min Frontend Phantom Phantom thickness Text Attachments
  83   Sat Jun 21 19:22:45 2025 Julia Müllner73.2 MeV systemtest 9TOF-Cal GSI73.2 MeV/uLow Flux HighHeliumGood19:2319:3512   Threshold change from 50 to 500

10GB collected
  
  82   Sat Jun 21 18:22:28 2025 Julia Müllner73.2 MeV systemtest 8TOF-Cal GSI73.2 MeV/uLow Flux HighHeliumCrap19:1019:2111   new callibration was done, dimanond detector
connected to readout

Threshold = 4000
  
  81   Sat Jun 21 18:05:21 2025 Julia Müllner97.2 MeV systemtest 7TOF-Cal GSI97.2MeV/uLow Flux HighHeliumGood18:0618:2115   Threshold set to 2000

50GB data collected

Room Entry Request: 18:21
  
  80   Sat Jun 21 17:54:27 2025 Julia Müllner97.2 MeV systemtest 6TOF-Cal GSI97.2MeV/uLow Flux HighHeliumGood17:5618:048   Threshold set to 4000

40GB data collected
  
  79   Sat Jun 21 17:13:34 2025 Julia Müllner97.2 MeV systemtest 5TOF-Cal GSI97.2MeV/uLow Flux HighHeliumGood17:3117:5423   Threshold set to 3000

moved Ripple filters back in 

100GB data collected
  
  78   Sat Jun 21 17:01:05 2025 Julia Müllner97.2 MeV systemtest 4TOF-Cal GSI97.2MeV/uLow Flux HighHeliumGood16:5817:1113   Data was good, we have to shuft the setup
in x direction and slightly down. Enter room
request 17:12.
  
  77   Sat Jun 21 16:30:44 2025 Julia Müllner205.2 MeV systemtest 3TOF-Cal GSI205.2 MeV/uLow Flux HighHeliumUnknown16:3816:5719   Threshold set to 5000

dead channels on last LGAD, changed
trigger from LGAD 12 to LGAD 11
  
  76   Sat Jun 21 15:43:47 2025 Julia Müllner800MeV systemtest 2TOF-Cal GSI800 MeVLow Flux HighProtonGood16:0216:2927   Threshold set to 1000

Factor 4 chosen for future runs,
80 GB in total recorded in this run
  
  75   Sat Jun 21 14:30:29 2025 Julia Müllner800MeV systemtest 1TOF-Cal GSI800 MeVLow Flux HighProtonGood14:4515:4257   first spills were done with low flux medium,
changed to high at 14:49

again changed to medium at 14:55
  
  74   Sat Mar 15 21:29:28 2025 RD50 Generic UserSlab and rate testing (with Michael)13Sandwich-iCT setup(GSI)+RaMon205.2 MeV/uLow Flux HighHeliumGood20:4421:2238 Plastic Water Slabs4cmtotzeit optimieren und deadtime finden

hex wert mal 10ns deadtime

13 ab file 10 besat effort
  
  73   Sat Mar 15 21:25:18 2025 RD50 Generic UserSlab Test Water 5cm14Sandwich-iCT setup(GSI)+RaMon205.2 MeV/uLow Flux HighHeliumGood21:2621:5024 Plastic Water Slabs5cmRaMon rate: 4.5MHz   
  72   Sat Mar 15 19:37:42 2025 RD50 Generic UserSlab Test Water 4cm12Sandwich-iCT setup(GSI)+RaMon205.2 MeV/uLow Flux MediumHelium 19:3920:4162 Plastic Water Slabs4cmRaMon Rate 1.4MHz   
  71   Sat Mar 15 18:37:36 2025 RD50 Generic UserSlab Test Water 3cm11Sandwich-iCT setup(GSI)+RaMon205.2 MeV/uLow Flux MediumHelium 18:4119:2847 Plastic Water Slabs3cmRaMon Rate 1.5MHz   
  70   Sat Mar 15 17:30:18 2025 RD50 Generic UserSlab Test Water 2cm10Sandwich-iCT setup(GSI)+RaMon205.2 MeV/uLow Flux MediumHelium 17:3018:2454 Plastic Water Slabs2cmRaMon Rate 1.5 MHz   
  69   Sat Mar 15 16:50:39 2025 RD50 Generic UserSlab Test Water 1cm9Sandwich-iCT setup(GSI)+RaMon205.2 MeV/uLow Flux MediumHeliumGood16:3517:2348 Plastic Water Slabs1cmRaMon Rate: 1.3MHz   
  68   Sat Mar 15 15:26:08 2025 RD50 Generic UserRate testing8Sandwich-iCT setup(GSI)+RaMon205.2 MeV/uLow Flux MediumHelium 15:3116:3059   RaMon Rate: 1.3MHz

Comments:

Tested different evenbuilder settings
  
  67   Sat Mar 15 15:02:39 2025 RD50 Generic UserRate testing7Sandwich-iCT setup(GSI)+RaMon205.2 MeV/uLow Flux HighHeliumGood15:0415:2521   RaMon Rate 4.5MHz

Comments:

From File 7 until 21 optional false,
  
  66   Sat Mar 15 13:15:25 2025 RD50 Generic UserSlab Measurement 205MeV6Sandwich-iCT setup(GSI)+RaMon205.2 MeV/uLow Flux MediumHeliumGood13:2714:4982   RaMon Rate 1.3 MHz

Comments:

We entered the room and connected
  
  65   Sat Mar 15 11:46:28 2025 RD50 Generic UserSlab Measurement 145MeV5Sandwich-iCT setup(GSI)+RaMon145.4 MeVLow Flux MediumProtonGood11:4813:1385   RaMon Rate 600kHz

front sensors 200mu: ~1kHz pro
Streifen
  
  64   Sat Mar 15 10:07:28 2025 RD50 Generic UserAlignment Run 252MeV4Sandwich-iCT setup(GSI)+RaMon252.7 MeVLow Flux MediumProtonGood10:0711:4396   RaMon Rate 1MHz

front sensors 200mu: ~10-25kHz
pro Streifen
  
ELOG V3.1.5-fc6679b