  73   Sat Jun 8 19:56:37 2024 Felix Ulrich-PurTOF CAL HADES LGADs Calibrationcurve2TOF-Cal GSI252.7 MeVLow Flux MediumProtonGood16:2916:47    1 Mhz on Ramon (after 5 Layers of Silicon
and close to second iso centre) and 350kHz
on LGAD
  
  74   Sat Jun 8 19:57:39 2024 Felix Ulrich-PurTOF CAL HADES LGADs Calibrationcurve3TOF-Cal GSI194.3 MeVLow Flux MediumProtonGood16:5117:28    600 khz on Ramon (after 5 Layers of Silicon
and close to second iso centre) and 235kHz
on LGAD
  
  75   Sat Jun 8 20:05:27 2024 Felix Ulrich-PurTOF CAL HADES LGADs Calibrationcurve4TOF-Cal GSI175.3 MeVLow Flux MediumProtonGood17:3117:54    500 khz on Ramon (after 5 Layers of Silicon
and close to second iso centre) and 170kHz
on LGAD
  
  76   Sat Jun 8 20:07:44 2024 Felix Ulrich-PurTOF CAL HADES LGADs Calibrationcurve5TOF-Cal GSI145.4 MeVLow Flux MediumProtonGood17:5518:24    400 khz on Ramon (after 5 Layers of Silicon
and close to second iso centre) and 120kHz
on LGAD
  
  77   Sat Jun 8 20:08:55 2024 Felix Ulrich-PurTOF CAL HADES LGADs Calibrationcurve6TOF-Cal GSI115.7MeVLow Flux MediumProtonGood18:2419:00    400 khz on Ramon (after 5 Layers of Silicon
and close to second iso centre) and 65kHz
on LGAD
  
  78   Sat Jun 8 20:09:25 2024 Felix Ulrich-PurTOF CAL HADES LGADs Calibrationcurve7TOF-Cal GSI100.4 MeVLow Flux MediumProtonGood19:0019:42    300 khz on Ramon (after 5 Layers of Silicon
and close to second iso centre) and 65kHz
on LGAD
  
  79   Sat Jun 8 20:11:00 2024 Felix Ulrich-PurTOF CAL HADES LGADs Calibrationcurve8TOF-Cal GSI83 MeVLow Flux MediumProtonGood19:4220:51    300 khz on Ramon (after 5 Layers of Silicon
and close to second iso centre) and 35kHz
on LGAD
  
  80   Sat Jun 8 21:41:04 2024 Felix Ulrich-PurTOF CAL HADES LGADs Calibrationcurve9TOF-Cal GSI62.4 MeVLow Flux HighProtonGood20:5521:21    changed back to high due to beamsize vs
LGAD size factor
650 khz on Ramon (after 5 Layers of
  
  81   Sat Jun 8 21:42:13 2024 Felix Ulrich-PurTOF CAL HADES LGADs WET Messung Wasser10TOF-Cal GSI194.3 MeVLow Flux HighProtonGood21:2721:39    1cm plastic water (in isocentre)
6 Mhz on Ramon (after 5 Layers of
Silicon and close to second iso centre) and
  
  82   Sat Jun 8 21:43:26 2024 Felix Ulrich-PurTOF CAL HADES LGADs WET Messung Wasser11TOF-Cal GSI145.4 MeVLow Flux HighProtonGood21:3921:48    1cm plastic water (in isocentre)
4 Mhz on Ramon (after 5 Layers of
Silicon and close to second iso centre) and
  
  83   Sat Jun 8 21:54:59 2024 Felix Ulrich-PurTOF CAL HADES LGADs WET Messung Wasser12TOF-Cal GSI194.3 MeVLow Flux HighProtonGood21:5322:05    2cm plastic water (in isocentre)
5 Mhz on Ramon (after 5 Layers of
Silicon and close to second iso centre) and
  
  84   Sat Jun 8 22:12:06 2024 Felix Ulrich-PurTOF CAL HADES LGADs WET Messung Wasser13TOF-Cal GSI145.4 MeVLow Flux HighProtonGood22:0522:34    2cm plastic water (in isocentre)
3.2 Mhz on Ramon (after 5 Layers of
Silicon and close to second iso centre) and
  
  85   Sat Jun 8 22:38:41 2024 Felix Ulrich-PurTOF CAL HADES LGADs WET Messung Wasser14TOF-Cal GSI194.3 MeVLow Flux HighProtonGood22:1922.34    3 cm plastic water (in isocentre)
4.3 Mhz on Ramon (after 5 Layers of
Silicon and close to second iso centre) and
  
  86   Sun Jun 9 00:04:23 2024 Felix Ulrich-PurTOF CAL HADES LGADs WET Messung Wasser15TOF-Cal GSI194.3 MeVLow Flux HighProtonGood22:3422:50    3 cm plastic water (in isocentre)
2.6 Mhz on Ramon (after 5 Layers of
Silicon and close to second iso centre) and
  
  87   Sun Jun 9 00:05:11 2024 Felix Ulrich-PurTOF CAL HADES LGADs WET Messung Wasser (wrong energy)16TOF-Cal GSI194.3 MeVLow Flux HighProtonCrap22:5022:52    wrong energy don't use   
  88   Sun Jun 9 00:06:01 2024 Felix Ulrich-PurTOF CAL HADES LGADs WET Messung Wasser17TOF-Cal GSI194.3 MeVLow Flux HighProtonGood22:5323:04    4 cm plastic water (in isocentre)
3.8 Mhz on Ramon (after 5 Layers of
Silicon and close to second iso centre) and
  
  89   Sun Jun 9 00:07:17 2024 Felix Ulrich-PurTOF CAL HADES LGADs WET Messung Wasser18TOF-Cal GSI145.4 MeVLow Flux HighProtonGood23:0423:20    4 cm plastic water (in isocentre)
2 Mhz on Ramon (after 5 Layers of
Silicon and close to second iso centre) and
  
  90   Sun Jun 9 00:08:07 2024 Felix Ulrich-PurTOF CAL HADES LGADs WET Messung Wasser19TOF-Cal GSI194.3 MeVLow Flux HighProtonGood23:2023:30    5 cm plastic water (in isocentre)
3.7 Mhz on Ramon (after 5 Layers of
Silicon and close to second iso centre) and
  
  91   Sun Jun 9 00:10:38 2024 Felix Ulrich-PurTOF CAL HADES LGADs WET Messung Wasser20TOF-Cal GSI145.4 MeVLow Flux HighProtonGood23:3100:00    5 cm plastic water (in isocentre)
2 Mhz on Ramon (after 5 Layers of
Silicon and close to second iso centre) and
  
  92   Sun Jun 9 00:12:41 2024 Felix Ulrich-PurTOF CAL HADES LGADs WET Messung Knochen21TOF-Cal GSI194.3 MeVLow Flux HighProtonGood00:0000:14    1 cm dense bone (in isocentre)
4.5 Mhz on Ramon (after 5 Layers of
Silicon and close to second iso centre) and
  