changed back to high due to beamsize vs LGAD size factor 650 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 230kHz on LGAD
TOF Calibration Curve
7.5 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 2.5 MHz on LGAD
1 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 350kHz on LGAD
600 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 235kHz on LGAD
1cm plastic water (in isocentre) 6 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 1.5 MHz on LGAD
WET measurement
2cm plastic water (in isocentre) 5 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 1.2 MHz on LGAD
3 cm plastic water (in isocentre) 4.3 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 1MHz on LGAD
3 cm plastic water (in isocentre) 2.6 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 400 kHz on LGAD
wrong energy don't use
4 cm plastic water (in isocentre) 3.8 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 900 kHz on LGAD
5 cm plastic water (in isocentre) 3.7 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 795 kHz on LGAD
1 cm dense bone (in isocentre) 4.5 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 1.2 MHz on LGAD
2 cm dense bone (in isocentre) - Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 1 MHz on LGAD
interlock at 00:31 till 00:50 --> 19min break
3 cm dense bone (in isocentre) 2.8 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 600 kHz on LGAD
4 cm dense bone (in isocentre) 2 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 470 kHz on LGAD
5 cm dense bone (in isocentre) 1.8 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 350 kHz on LGAD