5 cm plastic water (in isocentre) 2 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 330 kHz on LGAD
14 min break with beam off (DAQ failed, had to power cycle TRB3s and all DIRICH boards-->out of synch DIRICH)
WET measurement
1 cm dense bone (in isocentre) 2.8 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 600 kHz on LGAD
2 cm dense bone (in isocentre) 2.1 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 315 kHz on LGAD
3 cm dense bone (in isocentre) 1.7 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 230 kHz on LGAD
4 cm dense bone (in isocentre) 1.4 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 175 kHz on LGAD
Mean rate per spill~ 2-8E5 p/s (online estimation without proper analysis)
Mean rate per spill~ 4Ee6 p/s (online estimation without proper analysis)
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000
5MHz with Ramon and only 500kHz TRB
400 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 65kHz on LGAD
TOF Calibration Curve
300 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 65kHz on LGAD
Tracker calibration run: mpw4-tj-monopix2_00-000999.xml Generated global xml file for MPW4 shift: mpw4-tj-monopix2_TLU_00-000999.xml
Previous local runs and xml files (mpw4-tj-monopix2_00-000212.xml) are unusable, because sensor L5.1.1.0 had a loose cotact and thus low gain. Fixed by disconnecting and reconnecting.
position of duts needs to be switched
new geo file with adjusted dut position
Low flux carbon did not work
1µs EFE, 30s Extraction, Deg100 we tried different energies (402.MeV/u)-> 120 MeV/u, but unfortunately we did not get any beam
5µs EFE, 10s Extraction, Deg100 we tried different energies (402.MeV/u)-> 238 MeV/u, but unfortunately we did not get any beam