Entry  Sun Jun 9 00:08:07 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 19, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Good, 23:20, 23:30, ,  

5 cm plastic water (in isocentre)
3.7 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 795 kHz on LGAD

WET measurement

Entry  Sun Jun 9 00:12:41 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Knochen, 21, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Good, 00:00, 00:14, ,  

1 cm dense bone (in isocentre)
4.5 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 1.2 MHz on LGAD

WET measurement

Entry  Sun Jun 9 01:25:26 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Knochen, 23, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Good, 00:30, 01:02, ,  

2 cm dense bone (in isocentre)
- Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 1 MHz on LGAD

interlock at 00:31 till 00:50 --> 19min break

WET measurement

Entry  Sun Jun 9 01:27:38 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Knochen, 25, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Good, 01:16, 01:30, ,  

3 cm dense bone (in isocentre)
2.8 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 600 kHz on LGAD

WET measurement

Entry  Sun Jun 9 01:48:51 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Knochen, 27, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Good, 01:48, 01:59, ,  

4 cm dense bone (in isocentre)
2 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 470 kHz on LGAD

WET measurement

Entry  Sun Jun 9 02:18:55 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Knochen, 29, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Good, 02:18, , ,  

5 cm dense bone (in isocentre)
1.8 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 350 kHz on LGAD

WET measurement

Entry  Sat Jun 8 20:05:27 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs Calibrationcurve, 4, TOF-Cal GSI, 175.3 MeV, Low Flux Medium, Proton, Good, 17:31, 17:54, ,  

500 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 170kHz on LGAD

TOF Calibration Curve

Entry  Sat Jun 8 20:07:44 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs Calibrationcurve, 5, TOF-Cal GSI, 145.4 MeV, Low Flux Medium, Proton, Good, 17:55, 18:24, ,  

400 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 120kHz on LGAD

TOF Calibration Curve

Entry  Sat Jun 8 21:43:26 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 11, TOF-Cal GSI, 145.4 MeV, Low Flux High, Proton, Good, 21:39, 21:48, ,  

1cm plastic water (in isocentre)
4 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 700 kHz on LGAD

WET measurement

Entry  Sat Jun 8 22:12:06 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 13, TOF-Cal GSI, 145.4 MeV, Low Flux High, Proton, Good, 22:05, 22:34, ,  

2cm plastic water (in isocentre)
3.2 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 600 kHz on LGAD

WET measurement

Entry  Sun Jun 9 00:07:17 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 18, TOF-Cal GSI, 145.4 MeV, Low Flux High, Proton, Good, 23:04, 23:20, ,  

4 cm plastic water (in isocentre)
2 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 500 kHz on LGAD

WET measurement

Entry  Sun Jun 9 00:10:38 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 20, TOF-Cal GSI, 145.4 MeV, Low Flux High, Proton, Good, 23:31, 00:00, ,  

5 cm plastic water (in isocentre)
2 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 330 kHz on LGAD

14 min break with beam off (DAQ failed, had to power cycle TRB3s and all DIRICH boards-->out of synch DIRICH)

WET measurement

Entry  Sun Jun 9 01:22:48 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Knochen, 22, TOF-Cal GSI, 145.4 MeV, Low Flux High, Proton, Good, 00:14, 00:23, ,  

1 cm dense bone (in isocentre)
2.8 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 600 kHz on LGAD

WET measurement

Entry  Sun Jun 9 01:26:56 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Knochen, 24, TOF-Cal GSI, 145.4 MeV, Low Flux High, Proton, Good, 01:02, 01:14, ,  

2 cm dense bone (in isocentre)
2.1 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 315 kHz on LGAD

WET measurement

Entry  Sun Jun 9 01:32:51 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Knochen, 26, TOF-Cal GSI, 145.4 MeV, Low Flux High, Proton, Good, 01:30, 01:45, ,  

3 cm dense bone (in isocentre)
1.7 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 230 kHz on LGAD

WET measurement

Entry  Sun Jun 9 02:02:16 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Knochen, 28, TOF-Cal GSI, 145.4 MeV, Low Flux High, Proton, Good, 02:00, 02:14, ,  

4 cm dense bone (in isocentre)
1.4 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 175 kHz on LGAD

WET measurement

Entry  Sat Jun 8 20:08:55 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs Calibrationcurve, 6, TOF-Cal GSI, 115.7MeV, Low Flux Medium, Proton, Good, 18:24, 19:00, ,  

400 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 65kHz on LGAD

TOF Calibration Curve

Entry  Sat Jun 8 20:09:25 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs Calibrationcurve, 7, TOF-Cal GSI, 100.4 MeV, Low Flux Medium, Proton, Good, 19:00, 19:42, ,  

300 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 65kHz on LGAD

TOF Calibration Curve

Entry  Sat Mar 30 06:34:39 2024, Bernhard Pilsl, And a new testbeam starts :D, , , , , , , , , ,  
 
Entry  Sat Mar 30 08:05:59 2024, Christian Irmler, RD50-MPW4, 999, , , , , , , , ,  

Tracker calibration run: mpw4-tj-monopix2_00-000999.xml
Generated global xml file for MPW4 shift: mpw4-tj-monopix2_TLU_00-000999.xml 

Previous local runs and xml files (mpw4-tj-monopix2_00-000212.xml) are unusable, because sensor L5.1.1.0 had a loose cotact and thus low gain.
Fixed by disconnecting and reconnecting.