Home CMS Production Clean room MedAustron HEPHY testbeams old
testbeam-MA2018 testbeam-MA2019 testbeam-MA2020 testbeam-MA2021 testbeam-MA2023 testbeam-MA2024
  Testbeams at MedAustron in 2024, Page 4 of 5  Not logged in ELOG logo
New entries since:Thu Jan 1 01:00:00 1970
Entry  Sat Jun 8 19:57:39 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs Calibrationcurve, 3, TOF-Cal GSI, 194.3 MeV, Low Flux Medium, Proton, Good, 16:51, 17:28, ,  

600 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 235kHz on LGAD

TOF Calibration Curve

Entry  Sat Jun 8 20:05:27 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs Calibrationcurve, 4, TOF-Cal GSI, 175.3 MeV, Low Flux Medium, Proton, Good, 17:31, 17:54, ,  

500 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 170kHz on LGAD

TOF Calibration Curve

Entry  Sat Jun 8 20:07:44 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs Calibrationcurve, 5, TOF-Cal GSI, 145.4 MeV, Low Flux Medium, Proton, Good, 17:55, 18:24, ,  

400 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 120kHz on LGAD

TOF Calibration Curve

Entry  Sat Jun 8 20:08:55 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs Calibrationcurve, 6, TOF-Cal GSI, 115.7MeV, Low Flux Medium, Proton, Good, 18:24, 19:00, ,  

400 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 65kHz on LGAD

TOF Calibration Curve

Entry  Sat Jun 8 20:09:25 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs Calibrationcurve, 7, TOF-Cal GSI, 100.4 MeV, Low Flux Medium, Proton, Good, 19:00, 19:42, ,  

300 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 65kHz on LGAD

TOF Calibration Curve

Entry  Sat Jun 8 20:11:00 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs Calibrationcurve, 8, TOF-Cal GSI, 83 MeV, Low Flux Medium, Proton, Good, 19:42, 20:51, ,  

300 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 35kHz on LGAD

TOF Calibration Curve

Entry  Sat Jun 8 21:41:04 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs Calibrationcurve, 9, TOF-Cal GSI, 62.4 MeV, Low Flux High, Proton, Good, 20:55, 21:21, ,  

changed back to high due to beamsize vs LGAD size factor
650 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 230kHz on LGAD

TOF Calibration Curve

Entry  Sat Jun 8 21:42:13 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 10, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Good, 21:27, 21:39, ,  

1cm plastic water (in isocentre)
6 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 1.5 MHz on LGAD

WET measurement

Entry  Sat Jun 8 21:43:26 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 11, TOF-Cal GSI, 145.4 MeV, Low Flux High, Proton, Good, 21:39, 21:48, ,  

1cm plastic water (in isocentre)
4 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 700 kHz on LGAD

WET measurement

Entry  Sat Jun 8 21:54:59 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 12, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Good, 21:53, 22:05, ,  

2cm plastic water (in isocentre)
5 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 1.2 MHz on LGAD

WET measurement

Entry  Sat Jun 8 22:12:06 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 13, TOF-Cal GSI, 145.4 MeV, Low Flux High, Proton, Good, 22:05, 22:34, ,  

2cm plastic water (in isocentre)
3.2 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 600 kHz on LGAD

WET measurement

Entry  Sat Jun 8 22:38:41 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 14, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Good, 22:19, 22.34, ,  

3 cm plastic water (in isocentre)
4.3 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 1MHz on LGAD

WET measurement

Entry  Sun Jun 9 00:04:23 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 15, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Good, 22:34, 22:50, ,  

3 cm plastic water (in isocentre)
2.6 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 400 kHz on LGAD

WET measurement

Entry  Sun Jun 9 00:06:01 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 17, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Good, 22:53, 23:04, ,  

4 cm plastic water (in isocentre)
3.8 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 900 kHz on LGAD

WET measurement

Entry  Sun Jun 9 00:07:17 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 18, TOF-Cal GSI, 145.4 MeV, Low Flux High, Proton, Good, 23:04, 23:20, ,  

4 cm plastic water (in isocentre)
2 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 500 kHz on LGAD

WET measurement

Entry  Sun Jun 9 00:08:07 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 19, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Good, 23:20, 23:30, ,  

5 cm plastic water (in isocentre)
3.7 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 795 kHz on LGAD

WET measurement

Entry  Sun Jun 9 00:10:38 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 20, TOF-Cal GSI, 145.4 MeV, Low Flux High, Proton, Good, 23:31, 00:00, ,  

5 cm plastic water (in isocentre)
2 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 330 kHz on LGAD

14 min break with beam off (DAQ failed, had to power cycle TRB3s and all DIRICH boards-->out of synch DIRICH)

WET measurement

Entry  Sun Jun 9 00:12:41 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Knochen, 21, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Good, 00:00, 00:14, ,  

1 cm dense bone (in isocentre)
4.5 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 1.2 MHz on LGAD

WET measurement

Entry  Sun Jun 9 01:22:48 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Knochen, 22, TOF-Cal GSI, 145.4 MeV, Low Flux High, Proton, Good, 00:14, 00:23, ,  

1 cm dense bone (in isocentre)
2.8 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 600 kHz on LGAD

WET measurement

Entry  Sun Jun 9 01:25:26 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Knochen, 23, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Good, 00:30, 01:02, ,  

2 cm dense bone (in isocentre)
- Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 1 MHz on LGAD

interlock at 00:31 till 00:50 --> 19min break

WET measurement

ELOG V3.1.5-fc6679b