ID Date Author Subjectdown Run Number DUT Beam Energy Rate Setting Particle Type Events Data Beam Betatron Beam Chopper Beam Degrader Beam EFE Beam Orb Beam QuadScan PhantomType PhantomAngle
  206   Tue Mar 16 15:32:48 2021 Patrick SiebererSiC strip sensor69SiC+DSSD tracker62.4II(BETA+EFE)Proton Crap        

Testrun for Beam
Ripple Filter jumping around

  207   Tue Mar 16 15:33:23 2021 Patrick SiebererSiC strip sensor70SiC+DSSD tracker62.4II(BETA+EFE)Proton1.1MGood        

Energy Scan
00:21
Filter vermutlich draussen

  208   Tue Mar 16 15:34:25 2021 Patrick SiebererSiC strip sensor71SiC+DSSD tracker100.4II(BETA+EFE)Proton1.2MGood        

Energy Scan
00:46
Filter vermutlich draussen

  209   Tue Mar 16 15:34:50 2021 Patrick SiebererSiC strip sensor72SiC+DSSD tracker83II(BETA+EFE)Proton1.2MGood        

Energy Scan
01:10
Filter vermutlich draussen

  210   Tue Mar 16 15:35:10 2021 Patrick SiebererSiC strip sensor73SiC+DSSD tracker145.4II(BETA+EFE)Proton1MGood        

Energy Scan
01:35
Filter vermutlich draussen

  211   Tue Mar 16 15:35:35 2021 Patrick SiebererSiC strip sensor74SiC+DSSD tracker252.7II(BETA+EFE)Proton3MGood        

Energy Scan
01:54
Filter vermutlich draussen

  212   Tue Mar 16 15:35:59 2021 Patrick SiebererSiC strip sensor75SiC+DSSD tracker800VProton400kGood        

Energy Scan
02:48
1-6M Hz!

  213   Tue Mar 16 15:37:05 2021 Patrick SiebererSiC strip sensor76SiC+DSSD tracker800VIProton Crap        

Energy Scan
03:02
4Hz, VII -> MHz ?

  214   Tue Mar 16 15:37:49 2021 Patrick SiebererSiC strip sensor77SiC+DSSD tracker800VProton6.6MGood        

Energy Scan
03:06-5:18
MHz Rate
Danach: Abbauen

  76   Sat Feb 27 09:26:27 2021 Felix Ulrich-PurRate verification1SiC+Si+RaMon62.4III(EFE+SID2)  Unknown        

Start 06:37
rate looks okish. few emtpy spills

 

2021-02-27, 06:07:

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.475V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.112V channel 1, 0-Level: 2.315V channel 1

position:

x: 0
y: 0 

2021-02-27, 06:30:

Referenz: 2.477V channel 1
channel 2 unverändert
 

2021-02-27, 06:37:

Setting: 3 

Run 1: 6:37 62.4 MeV, Rate geringer als erwartet 

 

  77   Sat Feb 27 09:28:07 2021 Felix Ulrich-PurRate verification2SiC+Si+RaMon252.7III(EFE+SID2)  Crap        

Start 06:47
rate only 10Hz, stopped run, room request and rewired trigger.

NOTE: Problem was caused by operators: Optics were not set correctly after ramping up energy from 62.4 to 252.7, not trigger cable

 

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.475V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x: 0
y: 0 

  78   Sat Feb 27 09:29:41 2021 Felix Ulrich-PurRate verification4SiC+Si+RaMon252.7III(EFE+SID2)  Good        

Start 06:55
reallocated correctly, rate looks similar to old measurements now (kHz ish)

 

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.475V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x: 0
y: 0 

  79   Sat Feb 27 09:30:32 2021 Felix Ulrich-PurRate verification5SiC+Si+RaMon62.4III(EFE+SID2)  Good        

Start 07:03
verified if run 1 was reliable-->Same results

 

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.475V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x: 0
y: 0 

  80   Sat Feb 27 09:31:57 2021 Felix Ulrich-PurRate verification6SiC+Si+RaMon252.7I(EFE+SID0)  Good        

Start 07:16
rate looks fine,

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.475V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x: 0
y: 0 

SiC: man sieht erstmals Signale 

  81   Sat Feb 27 09:40:15 2021 Felix Ulrich-PurRate verification7SiC+Si+RaMon62.4I(EFE+SID0)  Good        

Start 07:40
rate looks fine, however operators mentioned, that CPD data might have been loaded incorrectly-->pause during run and reloaded (07:48), rate seems unchanged

 

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.475V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x: 0
y: 0 
 

  312   Sat Oct 8 07:11:03 2022 Christian IrmlerRamon only1RD50-MPW3800 Proton Craplowflux high       
  311   Sat Oct 8 07:10:31 2022 Christian IrmlerRD50-MPW3 - First Testbeam               

Tracker xml file: hvcmos_mpw3_20221008_TLU_00-000053.xml

 

  285   Sun Nov 7 19:00:54 2021 Felix Ulrich-PurPeakfinder WET meas 7.11.2113 175.3nominal  Good  10   none 

Start: 08:30
WET measurement of PMMA slabs also used for TOF-CT (GSI)

coarse run (to find Bragg range

  286   Sun Nov 7 19:02:25 2021 Felix Ulrich-PurPeakfinder WET meas 7.11.2114 175.3nominal  Good  10   none 

Start: 08:55
WET measurement of PMMA slabs also used for TOF-CT (GSI)

reference run (with Arduino Spill Start/Stop logic)
fringes in Bragg peak (trigger not optimal (maybe too much latency with Arduino))--> next measurement with "Accumalte Charge" instead of "Spill Start/Stop"

  287   Sun Nov 7 19:04:10 2021 Felix Ulrich-PurPeakfinder WET meas 7.11.2115 175.3nominal  Good  10   none 

Start: 09:14
WET measurement of PMMA slabs also used for TOF-CT (GSI)

reference run
"Accumalte Charge"