Entry  Sat Oct 8 10:35:23 2022, Christian Irmler, ramon + tracker + DUT, 21, RD50-MPW3, 252.7, , , , Crap, lowflux low, , , , , , ,  
ASIC: TH 1350
BEAM OFF, for TH check

no noise hits
Entry  Sat Oct 8 10:32:44 2022, Christian Irmler, ramon + tracker + DUT, 20, RD50-MPW3, 252.7, , , , Crap, lowflux low, , , , , , ,  
ASIC: TH 1350
BEAM OFF, for TH check

MPW3 monitor not connected --> invalid
Entry  Sat Oct 8 10:30:38 2022, Christian Irmler, ramon + tracker + DUT, 19, RD50-MPW3, 252.7, , , , Good, lowflux low, , , , , , ,  
ASIC: TH 1300 again for verificationn - still too many noise hits
BEAM OFF, for TH check
Entry  Sat Oct 8 10:29:27 2022, Christian Irmler, ramon + tracker + DUT, 18, RD50-MPW3, 252.7, , , , Good, lowflux low, , , , , , ,  
ASIC: TH 1400
BEAM OFF, for TH check
Entry  Sat Oct 8 10:24:12 2022, Christian Irmler, ramon + tracker + DUT, 17, RD50-MPW3, 252.7, , , , Good, lowflux low, , , , , , ,  
ASIC: TH 1600 - too high
BEAM OFF, for TH check
Entry  Sat Oct 8 10:21:20 2022, Christian Irmler, ramon + tracker + DUT, 16, RD50-MPW3, 252.7, , , , Good, lowflux low, , , , , , ,  
ASIC: TH 1800 - too high, no noise hits
BEAM OFF, for TH check
Entry  Sat Oct 8 10:16:45 2022, Patrick Sieberer, ramon + tracker + DUT, 15, RD50-MPW3, 252.7, , , , Good, lowflux low, , , , , , ,  
ASIC: TH 1300
BEAM OFF, for TH check
Entry  Sat Oct 8 10:11:27 2022, Patrick Sieberer, ramon + tracker + DUT, 14, RD50-MPW3, 252.7, , , , Good, lowflux low, , , , , , ,  
ASIC: TH 1300
Entry  Sat Oct 8 09:46:15 2022, Patrick Sieberer, ramon + tracker + DUT, 13, RD50-MPW3, 252.7, , , , Good, lowflux low, , , , , , ,  timing_tracker_run13.pngtracker_hitmap_run9.png
Ratenwechsel

Tracker: latency 42 > timing much better

ASIC: TH 1200
Entry  Sat Oct 8 09:33:59 2022, Patrick Sieberer, ramon + tracker + DUT, 12, RD50-MPW3, 252.7, , , 9.8M, Good, lowflux high, , , , , , ,  
repeat run 9 again ;)

run 11 used for new local runs of tracker - hvcmos_mpw3_20221008_11_TLU_00-000011.xml

tracker seems fine, at least in the beginning
Entry  Sat Oct 8 09:19:01 2022, Patrick Sieberer, ramon + tracker + DUT, 10, RD50-MPW3, 252.7, , , , Good, lowflux high, , , , , , ,  run9_hitmap_tracker.png
repeat run 9

max_trgs: 100

tracker still not working.
Entry  Sat Oct 8 09:08:13 2022, Patrick Sieberer, ramon + tracker + DUT, 9, RD50-MPW3, 252.7, , , kurz, Good, lowflux high, , , , , , ,  
TLU: 6,8M Tracker nur 4 M -> Let's reset FADC
Entry  Sat Oct 8 09:07:48 2022, Patrick Sieberer, ramon + tracker + DUT, 8, RD50-MPW3, 252.7, , , kurz, Good, lowflux high, , , , , , ,  
short run for uploading to analysis
Entry  Sat Oct 8 09:02:20 2022, Patrick Sieberer, ramon + tracker + DUT, 7, RD50-MPW3, 252.7, , , 5 min data taking, Good, lowflux high, , , , , , ,  
First data run w/ ramon + tracker + MPW3 + monitor connected
Entry  Sat Oct 8 08:43:15 2022, Christian Irmler, ramon + tracker + DUT, 6, RD50-MPW3, 252.7, , , , Good, lowflux high, , , , , , ,  APV_sync_error.png
~4 MHz particle rate
MPW3 data file size after 5 minutes: 11 GB :-$

Tracker lost data again.
Entry  Sat Oct 8 07:40:51 2022, Christian Irmler, ramon + tracker + DUT, 5, RD50-MPW3, 800, , , 24260338, Good, lowflux high, , , , , , ,  
Tracker latency lat=42

First data run w/ ramon + tracker + MPW3 + monitor connected

Tracker dropped some udp packages, total tracker events 23166448
Entry  Sat Oct 8 07:30:37 2022, Christian Irmler, testrun ramon + tracker + DUT, 4, RD50-MPW3, 800, , , , Unknown, lowflux high, , , , , , ,  beam-profil-tracker.pngtiming_tracker.png
Tracker latency lat=42

RD50-MPW3 monitor connected first
Entry  Sat Oct 8 07:28:42 2022, Christian Irmler, testrun ramon + tracker + DUT, 3, RD50-MPW3, 800, , , , Unknown, lowflux high, , , , , , ,  
Tracker latency lat=42
Entry  Sat Oct 8 07:20:45 2022, Christian Irmler, testrun ramon + tracker + DUT, 2, RD50-MPW3, 800, , , , Crap, lowflux high, , , , , , ,  
Tracker timing wrong, latency too low, peak at 0 --> changed lat from 40 to 41 for next run.
Entry  Sat Oct 8 07:11:03 2022, Christian Irmler, Ramon only, 1, RD50-MPW3, 800, , Proton, , Crap, lowflux high, , , , , , ,  
 