Sun Nov 7 19:00:36 2021, Felix Ulrich-Pur, Low flux test 7.11.21, 12, , 62.4, III(EFE+SID2), , 30 spills, Good, , , , , , , none,
|
Start: 07:44 O(kHz) |
Sun Nov 7 19:00:54 2021, Felix Ulrich-Pur, Peakfinder WET meas 7.11.21, 13, , 175.3, nominal, , , Good, , , 10, , , , none,
|
Start: 08:30
WET measurement of PMMA slabs also used for TOF-CT (GSI)
coarse run (to find Bragg range |
Sun Nov 7 19:02:25 2021, Felix Ulrich-Pur, Peakfinder WET meas 7.11.21, 14, , 175.3, nominal, , , Good, , , 10, , , , none,
|
Start: 08:55
WET measurement of PMMA slabs also used for TOF-CT (GSI)
reference run (with Arduino Spill Start/Stop logic)
fringes in Bragg peak (trigger not optimal (maybe too much latency with Arduino))--> next measurement with "Accumalte Charge" instead of "Spill Start/Stop" |
Sun Nov 7 19:04:10 2021, Felix Ulrich-Pur, Peakfinder WET meas 7.11.21, 15, , 175.3, nominal, , , Good, , , 10, , , , none,
|
Start: 09:14
WET measurement of PMMA slabs also used for TOF-CT (GSI)
reference run
"Accumalte Charge" |
Sun Nov 7 19:05:15 2021, Felix Ulrich-Pur, Peakfinder WET meas 7.11.21, 16, , 175.3, nominal, , , Good, , , 10, , , , none,
|
Start: 09:38
WET measurement of PMMA slabs also used for TOF-CT (GSI)
1cm PMMA
"Accumalte Charge"
PMMA was activated. Had to wait ~10-15min after measurement |
Sun Nov 7 19:06:06 2021, Felix Ulrich-Pur, Peakfinder WET meas 7.11.21, 17, , 175.3, nominal, , , Good, , , 10, , , , none,
|
Start: 10:06
WET measurement of PMMA slabs also used for TOF-CT (GSI)
2cm PMMA
"Accumalte Charge"
PMMA was activated. Had to wait ~10-15min after measurement |
Sun Nov 7 19:06:26 2021, Felix Ulrich-Pur, Peakfinder WET meas 7.11.21, 18, , 175.3, nominal, , , Good, , , 10, , , , none,
|
Start: 10:30
WET measurement of PMMA slabs also used for TOF-CT (GSI)
3cm PMMA
"Accumalte Charge"
PMMA was activated. Had to wait ~10-15min after measurement |
Sun Nov 7 19:06:40 2021, Felix Ulrich-Pur, Peakfinder WET meas 7.11.21, 19, , 175.3, nominal, , , Good, , , 10, , , , none,
|
Start: 11:00
WET measurement of PMMA slabs also used for TOF-CT (GSI)
4cm PMMA
"Accumalte Charge"
PMMA was activated. Had to wait ~10-15min after measurement |
Sun Nov 7 19:06:58 2021, Felix Ulrich-Pur, Peakfinder WET meas 7.11.21, 20, , 175.3, nominal, , , Good, , , 10, , , , none,
|
Start: 11:24
WET measurement of PMMA slabs also used for TOF-CT (GSI)
5cm PMMA
"Accumalte Charge"
PMMA was activated. Had to wait ~10-15min after measurement |
Sun Nov 7 19:07:31 2021, Felix Ulrich-Pur, Peakfinder WET meas 7.11.21, 21, , 194.3, nominal, , , Good, , , 10, , , , none,
|
Start: 11:44
WET measurement of PMMA slabs also used for TOF-CT (GSI)
5cm PMMA. higher energy for Felix Feichtinger
"Accumalte Charge"
coarse run |
Sun Nov 7 19:08:47 2021, Felix Ulrich-Pur, Peakfinder WET meas 7.11.21, 22, , 194.3, nominal, , , Good, , , 10, , , , none,
|
Start: 12:00
WET measurement of PMMA slabs also used for TOF-CT (GSI)
5cm PMMA. higher energy for Felix Feichtinger
"Accumalte Charge"
more detailed measurement |
Sat Oct 8 07:10:31 2022, Christian Irmler, RD50-MPW3 - First Testbeam, , , , , , , , , , , , , , ,
|
Tracker xml file: hvcmos_mpw3_20221008_TLU_00-000053.xml
|
Sat Oct 8 07:20:45 2022, Christian Irmler, testrun ramon + tracker + DUT, 2, RD50-MPW3, 800, , , , Crap, lowflux high, , , , , , ,
|
Tracker timing wrong, latency too low, peak at 0 --> changed lat from 40 to 41 for next run. |
Sat Oct 8 07:28:42 2022, Christian Irmler, testrun ramon + tracker + DUT, 3, RD50-MPW3, 800, , , , Unknown, lowflux high, , , , , , ,
|
Tracker latency lat=42 |
Sat Oct 8 07:30:37 2022, Christian Irmler, testrun ramon + tracker + DUT, 4, RD50-MPW3, 800, , , , Unknown, lowflux high, , , , , , ,
|
Tracker latency lat=42
RD50-MPW3 monitor connected first |
Sat Oct 8 07:40:51 2022, Christian Irmler, ramon + tracker + DUT, 5, RD50-MPW3, 800, , , 24260338, Good, lowflux high, , , , , , ,
|
Tracker latency lat=42
First data run w/ ramon + tracker + MPW3 + monitor connected
Tracker dropped some udp packages, total tracker events 23166448 |
Sat Oct 8 08:43:15 2022, Christian Irmler, ramon + tracker + DUT, 6, RD50-MPW3, 252.7, , , , Good, lowflux high, , , , , , ,
|
~4 MHz particle rate
MPW3 data file size after 5 minutes: 11 GB :-$
Tracker lost data again.
[2022-10-08 08:49:11 CEST] Error::FADCGbEMerger::Event missmatch
[2022-10-08 08:49:11 CEST] Error::FADCGbEMerger::Event missmatch
[2022-10-08 08:49:11 CEST] Error::FADCGbEMerger::Event missmatch
[2022-10-08 08:49:11 CEST] Error::FADCGbEMerger::Event missmatch
[2022-10-08 08:49:11 CEST] Warning::UDPUnpacker::Expected package number 1953610, got: 1953990!
Sync-Error L5.1.1 and L5.1.2 |
Sat Oct 8 09:02:20 2022, Patrick Sieberer, ramon + tracker + DUT, 7, RD50-MPW3, 252.7, , , 5 min data taking, Good, lowflux high, , , , , , ,
|
First data run w/ ramon + tracker + MPW3 + monitor connected |
Sat Oct 8 09:07:48 2022, Patrick Sieberer, ramon + tracker + DUT, 8, RD50-MPW3, 252.7, , , kurz, Good, lowflux high, , , , , , ,
|
short run for uploading to analysis |
Sat Oct 8 09:08:13 2022, Patrick Sieberer, ramon + tracker + DUT, 9, RD50-MPW3, 252.7, , , kurz, Good, lowflux high, , , , , , ,
|
TLU: 6,8M Tracker nur 4 M -> Let's reset FADC |
|