Home CMS Production Clean room MedAustron HEPHY testbeams old
testbeam-MA2018 testbeam-MA2019 testbeam-MA2020 testbeam-MA2021 testbeam-MA2023 testbeam-MA2024
  MedAustron 2021, Page 14 of 18  Not logged in ELOG logo
Entry  Thu Oct 7 10:59:18 2021, Philipp Gaggl, Pad Sensoren, 18, SiC+Si, 62.4, II(BETA+EFE), Proton, , Unknown, , , , , , , ,  

beamtime on 18.09.2021
CH1: SiC
CH3: Si
Bias_Si: -100 V
Bias_SiC: 400 V

Entry  Sun Nov 7 19:06:26 2021, Felix Ulrich-Pur, Peakfinder WET meas 7.11.21, 18, , 175.3, nominal, , , Good, , , 10, , , , none,  

Start: 10:30
WET measurement of PMMA slabs also used for TOF-CT (GSI)

3cm PMMA
"Accumalte Charge"

PMMA was activated. Had to wait ~10-15min after measurement

Entry  Sat Oct 8 10:29:27 2022, Christian Irmler, ramon + tracker + DUT, 18, RD50-MPW3, 252.7, , , , Good, lowflux low, , , , , , ,  

ASIC: TH 1400
BEAM OFF, for TH check

Entry  Sun Jan 10 06:52:24 2021, Maximilian Tomaschek, Lowflux recommissioning, 17, Ramon+Tracker, 252.7, , , 30 spills, , 10%, , 10, 1, , 0, ,  

Rate, now Beta 10%, looks like Beta EFE also ~400kHz

Entry  Sat Feb 27 18:44:23 2021, Felix Ulrich-Pur, Waveforms SiC+Si, 17, SiC+Si+RaMon, 800, VII, , , Good, , , , , , , ,  

Start 11:55
Stopped at 12:05-->acc failure

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.100 V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x = 8
y = 68

Entry  Tue Mar 16 15:05:24 2021, Patrick Sieberer, Pad Sensoren, 17, , 194.3, II(BETA+EFE), Proton, , Good, , , , , , , none,  

Start: 11:46
Range Shifter Out, Filter 1 IN, Filter 2 IN

Entry  Tue Mar 16 15:16:38 2021, Patrick Sieberer, MPW2 Analog output, 17, MPW2+Picoscope, 800, VII, Proton, , Good, , , , , , , ,  

3er Koinzidenz
Pix 06, 33, 34

Entry  Sun Jun 20 20:05:38 2021, Felix Ulrich-Pur, HADES testrun with increased rate 800MeV, 17, HADES-LGADs, 800, nominal, Proton, , Good, , , 10, , , , none, 0 

TB 20.6.21

Start 20:10
Mean rate O(200MHz)
 

Entry  Thu Oct 7 10:58:47 2021, Philipp Gaggl, Pad Sensoren, 17, SiC+Si, 62.4, II(BETA+EFE), Proton, , Unknown, , , , , , , ,  

beamtime on 18.09.2021
CH1: SiC
CH3: Si
Bias_Si: -100 V
Bias_SiC: 1100 V
Trigger SiC

Entry  Sun Nov 7 19:06:06 2021, Felix Ulrich-Pur, Peakfinder WET meas 7.11.21, 17, , 175.3, nominal, , , Good, , , 10, , , , none,  

Start: 10:06
WET measurement of PMMA slabs also used for TOF-CT (GSI)

2cm PMMA
"Accumalte Charge"

PMMA was activated. Had to wait ~10-15min after measurement

Entry  Sat Oct 8 10:24:12 2022, Christian Irmler, ramon + tracker + DUT, 17, RD50-MPW3, 252.7, , , , Good, lowflux low, , , , , , ,  

ASIC: TH 1600 - too high
BEAM OFF, for TH check

Entry  Sun Jan 10 06:52:09 2021, Maximilian Tomaschek, Lowflux recommissioning, 16, Ramon+Tracker, 62.4, , , 30 spills, , 10%, , 10, 1, , 2, ,  

Rate, now Beta 10%

Entry  Sat Feb 27 18:44:06 2021, Felix Ulrich-Pur, Waveforms SiC+Si, 16, SiC+Si+RaMon, 100.4, I(EFE+SID0), , , Good, , , , , , , ,  

Start 11:36

schöne Signale

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.100 V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x = 8
y = 68

Entry  Tue Mar 16 15:05:08 2021, Patrick Sieberer, Pad Sensoren, 16, , 100.4, II(BETA+EFE), Proton, , Good, , , , , , , none,  

Start: 11:38
Range Shifter Out, Filter 1 IN, Filter 2 IN

Entry  Tue Mar 16 15:16:24 2021, Patrick Sieberer, MPW2 Analog output, 16, MPW2+Picoscope, 175.3, I(EFE+SID0), Proton, , Good, , , , , , , ,  

3er Koinzidenz
Pix 06, 33, 34

Entry  Sun Jun 20 20:00:05 2021, Felix Ulrich-Pur, HADES testrun 800MeV, 16, HADES-LGADs, 800, VII, Proton, , Good, , , , , , , none, 0 

TB 20.6.21

Start 19:19
Mean rate O(MHz)
 

Entry  Thu Oct 7 10:58:19 2021, Philipp Gaggl, Pad Sensoren, 16, SiC+Si, 62.4, II(BETA+EFE), Proton, , Unknown, , , , , , , ,  

beamtime on 18.09.2021
CH1: SiC
CH3: Si
Bias_Si: -100 V
Bias_SiC: 1000 V
Trigger SiC

Entry  Sun Nov 7 19:05:15 2021, Felix Ulrich-Pur, Peakfinder WET meas 7.11.21, 16, , 175.3, nominal, , , Good, , , 10, , , , none,  

Start: 09:38
WET measurement of PMMA slabs also used for TOF-CT (GSI)

1cm PMMA
"Accumalte Charge"

PMMA was activated. Had to wait ~10-15min after measurement

Entry  Sat Oct 8 10:21:20 2022, Christian Irmler, ramon + tracker + DUT, 16, RD50-MPW3, 252.7, , , , Good, lowflux low, , , , , , ,  

ASIC: TH 1800 - too high, no noise hits
BEAM OFF, for TH check

Entry  Sun Jan 10 06:51:57 2021, Maximilian Tomaschek, Lowflux recommissioning, 15, Ramon+Tracker, 83, , , 30 spills, , 10%, , 10, 1, , 2, ,  

Rate, now Beta 10%

ELOG V3.1.5-fc6679b