Entry  Sun Nov 21 13:55:05 2021, Simon Waid, Comparison SiC, Si, Si LGAD and Diamond, 2, SiC+Si, 62.4, I(EFE+SID0), Proton, 1000000, , , , , , , , ,  
Bias voltage SiC: 800

Bias voltage Si: -150

From a beam perspective the Si detector was the first target, on oscilloscope  channel 1, SiC was the second target connected to oscilloscope
Entry  Sun Nov 21 14:05:18 2021, Simon Waid, Comparison SiC, Si, Si LGAD and Diamond, 3, , 62.4, I(EFE+SID0), Proton, 10000, , , , , , , , ,  
Bias voltage SiC: 800

Bias voltage Si LGAD: -100

From a beam perspective the SiC detector was the first target, on oscilloscope channel 4, the Si LGAD was the second target connected to oscilloscope
Entry  Sun Nov 21 14:53:00 2021, Simon Waid, Comparison SiC, Si, Si LGAD and Diamond, 4, Diamond+LGAD, 62.4, I(EFE+SID0), Proton, 100000, , , , , , , , ,  
Bias voltage Diamond: 500

Bias voltage Si LGAD: -100

From a beam perspective the Diamond detector was the first target, on oscilloscope channel 4, the Si LGAD was the second target connected to
Entry  Sun Jan 10 06:27:26 2021, Maximilian Tomaschek, Lowflux recommissioning, 1, Ramon+Tracker, 252.7, , , 30 spills, , 10%, , 10, 1, , , ,  
Rate,15s spill pause, delay can be set by operator
Entry  Sun Jan 10 06:32:13 2021, Maximilian Tomaschek, Lowflux recommissioning, 2, Ramon+Tracker, 194.3, , , 30 spills, , 10%, , 10, 1, , , ,  
Rate
Entry  Sun Jan 10 06:33:22 2021, Maximilian Tomaschek, Lowflux recommissioning, 3, Ramon+Tracker, 175.3, , , 30 spills, , 10%, , 10, 1, , , ,  
Rate
Entry  Sun Jan 10 06:33:44 2021, Maximilian Tomaschek, Lowflux recommissioning, 5, Ramon+Tracker, 100.4, , , 30 spills, , 10%, , 10, 1, , , ,  
Rate
Entry  Sun Jan 10 06:38:16 2021, Maximilian Tomaschek, Lowflux recommissioning, 6, Ramon+Tracker, 83, , , 30 spills, , 10%, , 10, 1, , , ,  
Rate
Entry  Sun Jan 10 06:40:06 2021, Maximilian Tomaschek, Lowflux recommissioning, 7, Ramon+Tracker, 62.4, , , 30 spills, , 10%, , 10, 1, , , ,  
Rate, up till now looks like the same
Entry  Sun Jan 10 06:42:03 2021, Maximilian Tomaschek, Lowflux recommissioning, 8, Ramon+Tracker, 252.7, , , 30 spills, , , , 10, 1, , 2, ,  
Rate, factor 10 too high, role ~30kHz instead of 3kHz, Beta nominal
Entry  Sun Jan 10 06:44:59 2021, Maximilian Tomaschek, Lowflux recommissioning, 9, Ramon+Tracker, 62.4, , , 30 spills, , , , 10, 1, , 2, ,  
Rate, Beta nominal, factor 10 too high
Entry  Sun Jan 10 06:45:55 2021, Maximilian Tomaschek, Lowflux recommissioning, 10, Ramon+Tracker, 252.7, , , 30 spills, , 10%, , 10, 1, , 2, ,  
Rate, Beta 10% nominal, now 3kHz
Entry  Sun Jan 10 06:46:56 2021, Maximilian Tomaschek, Lowflux recommissioning, 11, Ramon+Tracker, 194.3, , , 30 spills, , 10%, , 10, 1, , 2, ,  
Rate, Beta 10% nominal
Entry  Sun Jan 10 06:50:45 2021, Maximilian Tomaschek, Lowflux recommissioning, 12, Ramon+Tracker, 175.3, , , 30 spills, , 10%, , 10, 1, , 2, ,  
Rate
Entry  Sun Jan 10 06:51:12 2021, Maximilian Tomaschek, Lowflux recommissioning, 14, Ramon+Tracker, 100.4, , , 30 spills, , 10%, , 10, 1, , 2, ,  
Rate, now Beta 10%
Entry  Sun Jan 10 06:51:57 2021, Maximilian Tomaschek, Lowflux recommissioning, 15, Ramon+Tracker, 83, , , 30 spills, , 10%, , 10, 1, , 2, ,  
Rate, now Beta 10%
Entry  Sun Jan 10 06:52:09 2021, Maximilian Tomaschek, Lowflux recommissioning, 16, Ramon+Tracker, 62.4, , , 30 spills, , 10%, , 10, 1, , 2, ,  
Rate, now Beta 10%
Entry  Sun Jan 10 06:52:24 2021, Maximilian Tomaschek, Lowflux recommissioning, 17, Ramon+Tracker, 252.7, , , 30 spills, , 10%, , 10, 1, , 0, ,  
Rate, now Beta 10%, looks like Beta EFE also ~400kHz
Entry  Sun Jan 10 07:01:04 2021, Maximilian Tomaschek, Lowflux recommissioning, 18, Ramon+Tracker, 252.7, , , 30 spills, , , , 10, 1, , 0, ,  
Rate, now Beta nominal again, looks like old setting 4-6 Mhz
Entry  Sun Jan 10 07:02:15 2021, Maximilian Tomaschek, Lowflux recommissioning, 19, Ramon+Tracker, 194.3, , , 30 spills, Crap, , , 10, 1, , 0, ,  
Rate, Beta nominal, ERROR tracker  -> reset TLU