Entry  Sun Dec 6 15:44:51 2020, Christian Irmler, MPW2, 200, W13_04, 252.7, , , , 1, , SID0, , HV = during ramping, 10 spills, Crap 

HV = 

r0 c2

Entry  Sun Dec 6 15:45:45 2020, Christian Irmler, MPW2, 201, W13_04, 252.7, , , , 1, , SID0, , HV = 100V, 10 spills,  

r0 c2

Entry  Sun Dec 6 15:53:12 2020, Christian Irmler, MPW2, 202, W13_04, 252.7, , , , 1, , SID0, , HV off, 10 spills,  

r0 c2

HV set to 0V, but KeithRC SW turned SMU output off 

Entry  Sun Dec 6 15:55:13 2020, Christian Irmler, MPW2, 203, W13_04, 252.7, , , , 1, , SID0, , HV = 1V, 10 spills,  

r0 c2

lowest Voltage, which could be set with SMU outputs on.

Entry  Sun Dec 6 15:57:28 2020, Christian Irmler, MPW2, 204, W13_04, 252.7, , , , 1, , SID0, , HV = 20V, 10 spills,  

r0 c2

Entry  Sun Dec 6 16:01:02 2020, Christian Irmler, MPW2, 205, W13_04, 252.7, , , , 1, , SID0, , HV = 80V, 10 spills,  

pixel scan, starting with r0 c0

stopped after r1 c5 due to synch issues

Entry  Sun Dec 6 16:32:54 2020, Christian Irmler, MPW2, 206, W13_04, 252.7, , , , 1, , SID0, , HV = 80V, 10 spills,  

pixel scan, starting with r1 c6

stopped after r3 c2

stopped due to time difference of ~10s between TLU and Caribou

Entry  Sun Dec 6 16:53:35 2020, Christian Irmler, MPW2, 207, W13_04, 252.7, , , , 1, , SID0, , HV = 80V, 10 spills,  

pixel scan, starting with r3 c3

tried to syncronized clocks of Caribou and TLU syncronized before run started.
clock of caribou is still 3s ahead. --> need to use ntp server next time.

was interrupted between spill 64 an 65 for about 40 minutes due to MAPTA error (HL-tank error)

stopped after r7 c7

Entry  Sun Dec 6 18:59:48 2020, Christian Irmler, MPW2, 208, W13_04, 252.7, , , , 1, , SID0, , HV = 80V, 10 spills,  

single pixel r0 c2

shutter time set to 2.5s (half of previous value)

Entry  Sun Dec 6 19:07:20 2020, Christian Irmler, MPW2, 209, W13_04, 252.7, , , , 1, , SID0, , HV = 80V, 10 spills,  

single pixel r0 c2

shutter time set to 6s 

Entry  Fri May 8 15:17:39 2020, Felix Ulrich-Pur, , , DSSD-Tracker, Noise, , , , , , , , , ,  
 
Entry  Sun May 10 13:43:49 2020, Christian Irmler, setup, 1, Diamond + DSSD tracker, , , , , , , , none, , ,  setup.jpg
 
Entry  Sun May 10 14:51:12 2020, Christian Irmler, Calibration Runs, 1, Diamond + DSSD tracker, Noise, , , , , , , none, , ,  

ADC delay, FIR, ped, noise,cal with DSSD tracker + diamond
DSSD HV= +/- 50V
Dia HV = +/- 250V

diamond not yet pumped

calibration files: pct_dia_01_silimond_tb_01-000001.xml

Entry  Sun May 10 16:44:46 2020, Christian Irmler, Calibration Diamond 500V, 3, Diamond, 252.7, none, none, , , , , none, , ,  

diamond only
Dia HV = +/- 250V

ADC, FIR, ped, noise, cal

Entry  Sun May 10 16:47:10 2020, Christian Irmler, Calibration Diamond 600V, 4, Diamond, 252.7, none, none, , , , , none, , ,  

diamond only
Dia HV = +/- 300V

ADC, FIR, ped, noise, cal

Entry  Sun May 10 17:01:42 2020, Christian Irmler, Calibration Diamond + pCT-6 600V, 6, Diamond + DSSD-pCT6, 252.7, none, none, , , , , none, , ,  

pCT-6 + diamond

DSSD HV = +/-50V
Dia HV = +/- 300V

ADC, FIR, ped, noise, cal

Entry  Sun May 10 17:14:00 2020, Christian Irmler, Calibration Diamond + pCT-6 300V, 7, Diamond + DSSD-pCT6, 252.7, none, none, , , , , none, , ,  

pCT-6 + diamond

DSSD HV = +/-50V
Dia HV = +/- 150V

ADC, FIR, ped, noise, cal

Entry  Sun May 10 17:16:58 2020, Christian Irmler, Calibration Diamond 300V, 8, Diamond, 252.7, none, none, , , , , none, , ,  

Diamond only
Dia HV = +/- 150V

ADC, FIR, ped, noise, cal

Entry  Sun May 10 17:21:28 2020, Felix Ulrich-Pur, Calibration Diamond 400V, 9, Diamond, 252.7, none, none, , , , , none, , ,  

Diamond only
Dia HV = +/- 200V

ADC, FIR, ped, noise, cal

Entry  Sun May 10 17:38:55 2020, Felix Ulrich-Pur, Calibration Diamond+pCT6 400V, 10, Diamond + DSSD-pCT6, 252.7, none, none, , , , , none, , ,  

Diamond+pCT6
Dia HV = +/- 200V

ADC, FIR, ped, noise, cal

ELOG V3.1.5-fc6679b