Home CMS Production Clean room MedAustron HEPHY testbeams old
testbeam-MA2018 testbeam-MA2019 testbeam-MA2020 testbeam-MA2021 testbeam-MA2023 testbeam-MA2024
  MedAustron 2020, Page 1 of 15  Not logged in ELOG logo
Entry  Fri Jun 12 23:10:40 2020, Felix Ulrich-Pur, Calibration TERA MPV 200, 1, TERA-Cal, 800, 10, 60, 20, 10, 6dB, , none, , ,  

Setting 4

was stopped, cause timout setting was not adjusted to spill structure

Entry  Fri Jun 12 23:47:07 2020, Felix Ulrich-Pur, Calibration TERA MPV 200, 2, TERA-Cal, 800, 10, 60, 20, 10, 6dB, , none, , ,  

Setting 4

slice R1 did not converge therofor 9 steps

Entry  Sat Jun 13 00:01:43 2020, Felix Ulrich-Pur, Calibration TERA MPV 300, 3, TERA-Cal, 800, 10, 60, 20, 10, 6dB, , none, , ,  

Setting 4

 

Entry  Sat Jun 13 00:14:44 2020, Felix Ulrich-Pur, Calibration TERA MPV 400, 4, TERA-Cal, 800, 10, 60, 20, 10, 6dB, , none, , ,  

Setting 4

stopped because one slice could not be calibrated due to SW error

Entry  Sat Jun 13 00:29:22 2020, Felix Ulrich-Pur, Calibration TERA MPV 400, 5, TERA-Cal, 800, 10, 60, 20, 10, 6dB, , none, , ,  

Setting 4

S9 did not converge run stopped

 

Entry  Sat Jun 13 00:34:32 2020, Felix Ulrich-Pur, Calibration TERA MPV 400, 6, TERA-Cal, 800, 10, 60, 20, 10, 6dB, , none, , ,  

Setting 4

 

Entry  Sat Jun 13 00:54:47 2020, Felix Ulrich-Pur, Calibration TERA MPV 300, 7, TERA-Cal, 800, 10, 60, 20, 10, 6dB, , none, , ,  

Setting 4

repeated 300, because of SW error in previous run

 

Entry  Sat Jun 13 01:07:52 2020, Felix Ulrich-Pur, Calibration TERA MPV 200, 8, TERA-Cal, 800, 10, 60, 20, 10, 6dB, , none, , ,  

Setting 4

repeated 200, because of SW error in previous run

 

Entry  Sat Jun 13 01:29:35 2020, Felix Ulrich-Pur, Calibration TERA MPV 600, 9, TERA-Cal, 800, 10, 60, 20, 10, 6dB, , none, , ,  

Setting 4

few slices did not converge -> adapt range setting

 

Entry  Sat Jun 13 01:51:17 2020, Felix Ulrich-Pur, Calibration TERA MPV 600, 10, TERA-Cal, 800, 10, 60, 20, 10, 6dB, , none, , ,  

Setting 4

few slices did not converge -> adapted range setting

q9 and p3 did not converge-->choose mean voltage of group

 

Entry  Sat Jun 13 02:16:06 2020, Felix Ulrich-Pur, Calibration TERA Sr-90 Voltages, 11, TERA-Cal, 800, 10, 60, 20, 10, 6dB, , none, , 60000,  

1Setting 4

used same voltages as calibrated with Sr-90 (MPV 300)

 

Entry  Sun Jun 21 06:12:08 2020, Felix Ulrich-Pur, Silimond 800MeV 500V, 32, Diamond + DSSD tracker, 800, 10, 60, 20, 10, 6dB, , , , 24116,  

800 MeV global run (tracking effciency)
setting4

Entry  Sun Jun 21 07:53:56 2020, Felix Ulrich-Pur, Diamond only 800MeV 500V, 33, Diamond, 800, 10, 60, 20, 10, 6dB, , , , 109363,  

800 MeV global run (tracking effciency)
setting4

 

telescope shifted 6mm to the left (in beam direction)

Entry  Sun Jun 21 07:54:38 2020, Felix Ulrich-Pur, Diamond only 800MeV 600V, 34, Diamond, 800, 10, 60, 20, 10, 6dB, , , , 101831,  

800 MeV global run (tracking effciency)
setting4

 

telescope shifted 6mm to the left (in beam direction)

Entry  Sun Jun 21 08:15:13 2020, Felix Ulrich-Pur, Diamond only 800MeV 700V, 35, Diamond, 800, 10, 60, 20, 10, 6dB, , , , 101522,  pct_dia_01_diamond_700V_beam_LocalHardware_02-000035_hitmap_n.png

800 MeV global run (tracking effciency)
setting4

 

telescope shifted 6mm to the left (in beam direction)

 

4 strips on n- side started to getted really noisy (408,409,483,484)

Entry  Sun Jun 21 08:43:23 2020, Felix Ulrich-Pur, Diamond only 800MeV 400V, 36, Diamond, 800, 10, 60, 20, 10, 6dB, , , , 101680,  

800 MeV global run (tracking effciency)
setting4

 

telescope shifted 6mm to the left (in beam direction)

strips are not noisy

Entry  Sun Jun 21 09:11:38 2020, Felix Ulrich-Pur, Diamond only 800MeV 300V, 37, Diamond, 800, 10, 60, 20, 10, 6dB, , , , 116365,  

800 MeV global run (tracking effciency)
setting4

 

telescope shifted 6mm to the left (in beam direction)

 

Entry  Sun Jun 21 09:36:52 2020, Felix Ulrich-Pur, Silimond 800MeV 800V, 38, Diamond + DSSD tracker, 800, 10, 60, 20, 10, 6dB, , , , 33035,  pct_dia_01_silimond_800V_beam_LocalHardware_02-000038_hitmap.png

800 MeV global run (tracking effciency)
setting4

 

telescope shifted 6mm to the left (in beam direction)

 

Entry  Sun Jun 21 10:21:53 2020, Felix Ulrich-Pur, Silimond 800MeV 800V, 39, Diamond + DSSD tracker, 800, 10, 60, 20, 10, 6dB, , , , 2498,  

800 MeV global run (tracking effciency)
setting4

 

telescope shifted 6mm to the left (in beam direction)

 

Entry  Sun Jun 21 10:22:37 2020, Felix Ulrich-Pur, Diamond only 800MeV 800V, 40, Diamond, 800, 10, 60, 20, 10, 6dB, , , , 108625,  

800 MeV global run (tracking effciency)
setting4

 

telescope shifted 6mm to the left (in beam direction)

 

ELOG V3.1.5-fc6679b