ID Date Author Subject DUT Beam Energyup Beam Setting Run Number Events PhantomType PhantomAngle
  1   Sat Jul 13 12:14:55 2019 Thomas Bergauer Drehtisch      

Settings 

  2   Sat Jul 13 13:32:30 2019 Thomas Bergauer2019-07-13f proton CTDSSD-Tracker EFE, QuadScan SID21 none 

Trigger: Coincidence using Scintillators

timing setting 41 (*32ns)

using 16ns cables from our scintialltors to TLU

  12   Sat Jul 13 18:42:24 2019 Alexander Burker2019-07-13f proton CTSketch pCT Setup    

Big Line: DSSD
Small Line: Malta
Small Square: Phantom
Small Rectangle: Trigger Scintillator
Big Rectangle: TERA

Beam is coming in from Left to Right

  13   Sat Jul 13 18:52:33 2019 Alexander Burker2019-07-13f proton CTSketch TERA Calibration Setup    
  46   Tue Sep 24 11:12:45 2019 Alexander Burker Sketch      

Notes of the planned order of modules within beam (and reserve modules), before switching out the first one

  75   Fri Oct 25 18:45:20 2019 Alexander Burker2019-10-20 Low Flux   2922002none 

  76   Fri Oct 25 18:50:08 2019 Alexander Burker2019-10-20 Low FluxSketch      

Spot Sizes Table: SpotsForBeamSettings.csv

Empirics, such as P2, P25, Median, P75 and P98 were obtained directly from the data. These may not be accurate if the beam is larger than the sensor.

To better describe these cases, fit values were obtained with a curve fit to a normal distribution. This can also be inaccurate, as many beam settings aren't normal and the automatic fit might have failed. A goodness-of-fit will be added to each row in the distant future

  77   Fri Oct 25 18:56:11 2019 Alexander Burker2019-07-13f proton CTSketch    none 

Spot sizes table: SpotsForBeamSettings.csv

Empirics, such as P2, P25, Median, P75 and P98 were obtained directly from the data. These may not be accurate if the beam is larger than the sensor.

To better describe these cases, fit values were obtained with a curve fit to a normal distribution. This can also be inaccurate, as many beam settings aren't normal and the automatic fit might have failed. A goodness-of-fit will be added to each row in the distant future

  79   Sat Oct 26 00:36:35 2019 Alexander Burker2019-10-25 Low Flux   21132none 

  99   Tue Oct 29 14:13:20 2019 Alexander Burker 2019-10-25 Low Flux   36656none 

  100   Tue Nov 5 10:50:08 2019 Florian PittersLGAD Setups on 20. and 25.10. ShiftLGAD      

Setup 1 (on 20.10.2019):

  • With Cividec

  • Position: beam center, ch3 (diode 1) in front, ch4 (diode 3) in the rear

  • RMS: ca. 23 mV, 24 mV

  • Trigger on Ch3: ca. 250 mV

  • Osci amplitude/div: 200 mV or 800 MeV, 400 mV for lower energies

 

Setup 2 (on 20.10.2019):

  • Same as setup 1 but 1 cm to the top from beam center

 

Setup 3 (on 25.10.2019):

  • Same as setup 1 but with Osci amplitude/div: 1V

 

Setup 4 (on 25.10.2019):

  • No Cividec

  • Postion: Same as setup 1

  • RMS: ca. 0.3 mV, 0.3 mV

  • Trigger: 3 mV

  • Osci amplitude/div: 10 mV

  101   Fri Nov 22 10:52:45 2019 Alexander Burker2019-11-23f proton CTSketch pCT Setup    

Sketch of the setup for experiment 3, proton CT

  6   Sat Jul 13 14:13:33 2019 Alexander Burker2019-07-13f proton CTTERA-Cal100.4EFE, QuadScan SID24531644nonenone

Trigger: Coincidence using Scintillators

TERA Linearitätsmessung

timing setting 41

 

  69   Fri Oct 25 18:40:33 2019 Alexander Burker2019-10-20 Low Flux 100.4Degrader 1023343040none 

12:43 

  70   Fri Oct 25 18:41:09 2019 Alexander Burker2019-10-20 Low Flux 100.4Degrader 2024185344none 

13:00

  90   Sat Oct 26 03:22:11 2019 Alexander Burker2019-10-25 Low Flux 100.4EFE 1, QuadScan SID 0, Degrader 10, Betatron 101548319none 

  97   Sat Oct 26 05:35:09 2019 Alexander Burker2019-10-25 Low Flux 100.4EFE, QuadScan SID0, Degrader 10227889none 

 

  118   Tue Nov 26 13:36:33 2019 Alexander Burker2019-11-23f proton CTDSSD-Tracker100.4 4039546none 

2019-11-24 16:42

"TLU not on network" (Alex does not know what was not on which network, maybe Felix' notes are more complete)

  120   Tue Nov 26 13:39:51 2019 Alexander Burker2019-11-23f proton CTDrehtisch100.4 42 Alu0

2019-11-24 17:21

TERA DAQ failed

  121   Tue Nov 26 13:40:21 2019 Alexander Burker2019-11-23f proton CTDrehtisch100.4 43313433Alu0

2019-11-24 17:21

Run was interrupted to re-load config. A regular pattern in the hit maps suggested a problem with FIR, however the data looks good and a projection could be reconstructed