Setup 1 (on 20.10.2019):
With Cividec
Position: beam center, ch3 (diode 1) in front, ch4 (diode 3) in the rear
RMS: ca. 23 mV, 24 mV
Trigger on Ch3: ca. 250 mV
Osci amplitude/div: 200 mV or 800 MeV, 400 mV for lower energies
Setup 2 (on 20.10.2019):
Same as setup 1 but 1 cm to the top from beam center
Setup 3 (on 25.10.2019):
Same as setup 1 but with Osci amplitude/div: 1V
Setup 4 (on 25.10.2019):
No Cividec
Postion: Same as setup 1
RMS: ca. 0.3 mV, 0.3 mV
Trigger: 3 mV
Osci amplitude/div: 10 mV
Sketch of the setup for experiment 3, proton CT
2019-11-23 21:30
TLU (Rack-mounted) failed: "netzwerk error"
2019-11-23 21:31
2019-11-23 21:32
2019-11-23 21:33
2019-11-23 21:46
TLU (Rack-mounted) failed: "netzwerk error" (even after a power cycle)
2019-11-24 13:43
no TLU error was observed
2019-11-24 14:05
promptly aborted due to "missing reason" (maybe Felix has some notes)
2019-11-24 14:06 - 14:20
Latency = 45 (found too late)
2019-11-24 14:24
Latency = 43 (found too late)
2019-11-24 14:26
Latency = 41 (this setting was kept)
2019-11-24 14:30
Random Triggers
2019-11-24 14:40
Coincidences Trigger
2019-11-24 15:08
Standalone TLU
2019-11-24 15:12
2019-11-24 15:13
2019-11-24 15:50
Standalone TLU + Tracker (Tera did not record data)
2019-11-24 16:42
"TLU not on network" (Alex does not know what was not on which network, maybe Felix' notes are more complete)
2019-11-24 16:48
"TLU not on network"
Data is good for Alignment