Entry  Wed Jan 29 12:10:06 2014, Wolfgang Treberspurg, Infineon, Single mode Run Stack 2 Bias Scan, HW_Run_Infineon_Stack2_BiasScan_005, 200k, 28.01.2014 20:00, 28.01.2014 23:00, Good 

HV at 210V; I at app. 8.5µA 3GeV, Single mode

 

Entry  Wed Jan 29 12:11:00 2014, Wolfgang Treberspurg, Infineon, Single mode Run Stack 2 Bias Scan, HW_Run_Infineon_Stack2_BiasScan_006, 200k, 28.01.2014 20:30, 28.01.2014 21:00, Good 

HV at 220V; I at app. 8.5µA 3GeV, Single mode

 

Entry  Wed Jan 29 21:55:09 2014, Wolfgang Treberspurg, Infineon, Single mode Run Stack 2 Bias Scan, HW_Run_Infineon_Stack2_BiasScan_007, 200k, 29.01.2014 11:50, 29.01.2014 12:15, Good 

HV at 230V; I at app. 8µA 3GeV, Single mode

 

Entry  Wed Jan 29 21:56:15 2014, Wolfgang Treberspurg, Infineon, Single mode Run Stack 2 Bias Scan, HW_Run_Infineon_Stack2_BiasScan_008, 200k, 29.01.2014 12:15, 29.01.2014 12:45, Good 

HV at 240V; I at app. 7.5µA 3GeV, Single mode

 

Entry  Wed Jan 29 21:59:51 2014, Wolfgang Treberspurg, Infineon, Single mode Run Stack 2 Bias Scan, HW_Run_Infineon_Stack2_BiasScan_011, 200k, 29.01.2014 13:45, 29.01.2014 14:20, Good 

HV at 270V; I at app. 7.6µA 3GeV, Single mode

 

Entry  Wed Jan 29 22:01:08 2014, Wolfgang Treberspurg, Infineon, Single mode Run Stack 2 Bias Scan, HW_Run_Infineon_Stack2_BiasScan_012, 200k, 29.01.2014 14:25, 29.01.2014 15:26, Good 

HV at 280V; I at app. 8µA 3GeV, Single mode

 

Entry  Wed Jan 29 22:02:04 2014, Wolfgang Treberspurg, Infineon, Single mode Run Stack 2 Bias Scan, HW_Run_Infineon_Stack2_BiasScan_013, 200k, 29.01.2014 15:26, 29.01.2014 17:00, Good 

HV at 290V; I at app. 8µA 3GeV, Single mode

 

Entry  Wed Jan 29 22:03:15 2014, Wolfgang Treberspurg, Infineon, Single mode Run Stack 2 Bias Scan, HW_Run_Infineon_Stack2_BiasScan_014, 200k, 29.01.2014 18:15, 29.01.2014 18:45, Good 

HV at 260V; I at app. 7.5µA 3GeV, Single mode

 

Entry  Wed Jan 29 22:06:12 2014, Wolfgang Treberspurg, Infineon, Single mode Run Stack 2 Bias Scan, HW_Run_Infineon_Stack2_BiasScan_015, 200k, 29.01.2014 19:15, 29.01.2014 19:45, Good 

HV at 300V; I at app. 7.5µA 3GeV, Single mode

 

Entry  Wed Jan 29 22:36:35 2014, Wolfgang Treberspurg, Infineon, Single mode Run Stack 2 Bias Scan, HW_Run_Infineon_Stack2_pos1_001, 200k, 29.01.2014 22:15, 29.01.2014 22:35, Good 

HV at 300V; I at app. 7.5µA 3GeV, Single mode

Position changed to -3.910970 -17.800610

 

Entry  Wed Jan 29 22:40:09 2014, Wolfgang Treberspurg, Infineon, Single mode Run Stack 2 Bias Scan, HW_Run_Infineon_Stack2_pos2_002, 200k, 29.01.2014 22:40, 29.01.2014 23:15, Good 

HV at 300V; I at app. 7.5µA 3GeV, Single mode

Position changed to --8.91097 -22.80061

 

Entry  Tue Jan 7 08:36:29 2014, Thomas Bergauer, Infineon, run, run014, 20k, 07.01.2014 08:34:06, 07.01.2014 09:05:47, Good 
HV=290V  I=9.55uA
position centered horizontally (+5mm)

DESY table z=-346 x=-255 

6Gev/c,

bad strip region hardly hit

Entry  Tue Jan 7 09:12:24 2014, Thomas Bergauer, Belle SVD, run, run015, 20k, 07.01.2014 09:10:52, 07.01.2014 09:15:39, Good 

 HV=290V  I=9.55uA

position centered horizontally (+5mm)
DESY table z=-346 x=-255
 
energy switched back to 3Gev/c
Entry  Wed Jan 8 14:10:07 2014, Thomas Bergauer, AIDA, run, run035, 211681, 08.01.2014 14:01:25, 08.01.2014 17:03:55, Unknown 

 HV=100V  I=4.2uA

position adapted to center beam
 

accidentally hit reset button on PC which ended this run abruptly...

 

5 GeV beam

Entry  Tue Jan 7 02:18:26 2014, Markus Friedl, Infineon, run, run013, 215k, 07.01.2014 02:14:50, 07.01.2014 08:31:25, Crap 

HV=150V  I=7.90uA

position centered horizontally (+5mm)

DESY table z=-346 x=-255

 

3 GeV beam momentum

 

close to end: switched to 6GeV

bias voltage accidentally left at 150V instead of initially planned 290V

Entry  Sat Jan 25 21:14:46 2014, Wolfgang Treberspurg, Infineon, Mulit 6 mode Run Stack 2, HW_Run_Infineon_Stack2_Multi_002, 230k, 25.01.2014 19:00, 25.01.2014 21:15, Good 

HV 300V, I 10µA

Beam Energy  4GeV

Latency changed to 93

Entry  Tue Jan 7 13:20:48 2014, Markus Friedl, Infineon, run, run020, 300k, 07.01.2014 13:18:52, 07.01.2014 14:35:49, Good 

HV=300V  I=8.5uA

position adapted to center beam better (+10mm horizontal)

narrow trigger width (~5ns)

multi6 with same settings as run019 (htr+3)

Beam energy = 3 GeV

Entry  Sun Jan 12 10:36:37 2014, Markus Friedl, AIDA, run, run055, 310913, 12.01.2014 10:34:44, 12.01.2014 15:42:47, Good 

HV=100V  I=3.6uA
position = 23 / -35 mm relative to before
5GeV beam, 6 samples
narrow timing (trigger width)

Entry  Wed Jan 8 17:20:40 2014, Markus Friedl, AIDA, run, run036, 320k, 08.01.2014 17:18:51, 08.01.2014 22:14:58, Unknown 

 HV=100V  I=4.3uA

centered position

5GeV beam

narrow timing (trigger width), but 6 samples

 

Entry  Sat Jan 25 18:46:32 2014, Wolfgang Treberspurg, Infineon, Mulit 6 mode Run Stack 2, HW_Run_Infineon_Stack2_Multi_001, 340k, 25.01.2014 16:30, 25.01.2014 18:40, Unknown 

HV 300V, I 10µA

Beam Energy changed during the run between 2GeV - 5GeV - 4GeV

Latency perhaps not optimal set (has been 95) but exchanged trigger cable seems to require 93

ELOG V3.1.5-fc6679b