  Alibava testbeam at DESY in Oct. 2017 (Dominic, Peter), Page 6 of 7  Not logged in ELOG logo
New entries since:Thu Jan 1 01:00:00 1970
Entry  Fri Oct 20 14:58:56 2017, Peter Paulitsch, IFX_200_3e14, run000184, 184, 100000, 75.93, 65.30, -110.55°, Yes, Good 
run settings
Parameter Value
Alibava filename  run000184.dat
Telescope filename  run000184.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 40 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 100000
Telescope triggers 100000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.5%

 

Entry  Fri Oct 20 14:57:30 2017, Peter Paulitsch, IFX_200_3e14, run000182, 182, 100000, 75.93, 65.30, -115.55°, Yes, Good 
run settings
Parameter Value
Alibava filename  run000182.dat
Telescope filename  run000182.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 40 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 100000
Telescope triggers 100000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.6%

 

Entry  Fri Oct 20 14:54:11 2017, Peter Paulitsch, IFX_200_3e14, run000181, 181, 700000, 75.93, 65.30, -70.55°, Yes, Good 
run settings
Parameter Value
Alibava filename  run000181.dat
Telescope filename  run000181.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 40 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 700000
Telescope triggers 700000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.4%

 

Entry  Fri Oct 20 14:52:36 2017, Peter Paulitsch, IFX_200_3e14, run000180, 180, 400000, 75.93, 65.30, -70.55°, Yes, Good 
run settings
Parameter Value
Alibava filename  run000180.dat
Telescope filename  run000180.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 40 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 400000
Telescope triggers 400000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.1%

 

Entry  Fri Oct 20 14:50:03 2017, Peter Paulitsch, IFX_200_3e14, run000179, 179, 200000, 75.93, 65.30, -70.55°, Yes, Good 
run settings
Parameter Value
Alibava filename  run000179.dat
Telescope filename  run000179.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -700 V
Dark current 40 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 200000
Telescope triggers 200000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.2%

 

Entry  Fri Oct 20 14:35:48 2017, Peter Paulitsch, IFX_200_3e14, run000177, 177, 1010000, 75.93, 65.30, -70.55°, Yes, Good 
run settings
Parameter Value
Alibava filename  run000177.dat
Telescope filename  run000177.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -400 V
Dark current 40 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.8%

 

Entry  Fri Oct 20 14:29:57 2017, Peter Paulitsch, IFX_200_3e14, run000173, 173, 110000, 75.93, 65.30, -70.55°, Yes, Good 
run settings
Parameter Value
Alibava filename  run000173.dat
Telescope filename  run000173.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 90 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.2%

 

Entry  Fri Oct 20 14:26:59 2017, Peter Paulitsch, IFX_200_3e14, run000169, 169, 110000, 75.93, 65.30, -70.55°, Yes, Good 
run settings
Parameter Value
Alibava filename  run000169.dat
Telescope filename  run000169.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 90 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.5%

 

Entry  Fri Oct 20 14:23:59 2017, Peter Paulitsch, IFX_200_3e14, run000166, 166, 110000, 75.93, 65.30, -70.55°, Yes, Good 
run settings
Parameter Value
Alibava filename  run000166.dat
Telescope filename  run000166.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 90 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.2%

 

Entry  Fri Oct 20 14:16:03 2017, Peter Paulitsch, IFX_200_3e14, run000164, 164, 110000, 75.93, 65.30, -70.55°, Yes, Good 
run settings
Parameter Value
Alibava filename  run000164.dat
Telescope filename  run000164.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 90 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 110000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency  250 Hz
Meas. time 12 min
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.5%

 

Entry  Fri Oct 20 14:14:15 2017, Peter Paulitsch, IFX_200_3e14, run000162, 162, 110000, 75.93, 65.30, -70.55°, Yes, Good 
run settings
Parameter Value
Alibava filename  run000162.dat
Telescope filename  run000162.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 90 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency  250 Hz
Meas. time 12 min
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.2%

 

Entry  Fri Oct 20 14:12:20 2017, Peter Paulitsch, IFX_200_3e14, run000124, 124, 110000, 75.93, 65.30, -72.55°, Yes, Good 
run settings
Parameter Value
Alibava filename  run000124.dat
Telescope filename  run000124.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 39 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency  210 Hz
Meas. time 15 min
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.3%

 

Entry  Fri Oct 20 14:11:03 2017, Peter Paulitsch, IFX_200_3e14, run000122, 122, 110000, 75.93, 65.30, -72.55°, Yes, Good 
run settings
Parameter Value
Alibava filename  run000122.dat
Telescope filename  run000122.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 39 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency  210 Hz
Meas. time 15 min
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.7%

 

Entry  Fri Oct 20 14:09:11 2017, Peter Paulitsch, IFX_200_3e14, run000120, 120, 110000, 75.93, 64.80, -72.55°, Yes, Good 
run settings
Parameter Value
Alibava filename  run000120.dat
Telescope filename  run000120.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 39 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y 75.93, 64.80
Region of interest bias line
Trigger frequency  210 Hz
Meas. time 15 min
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.5%

 

Entry  Fri Oct 20 14:05:39 2017, Peter Paulitsch, IFX_200_3e14, run000118, 118, 110000, 75.93, 61.20, -72.55°, Yes, Good 
run settings
Parameter Value
Alibava filename  run000118.dat
Telescope filename  run000118.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 39 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y 75.93, 61.20
Region of interest bias line
Trigger frequency  210 Hz
Meas. time 15 min
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.2%

 

Entry  Fri Oct 20 12:15:47 2017, Peter Paulitsch, IFX_200_3e14, run000116, 116, 110000, 75.93, 63.00, -72.55°, Yes, Good 
run settings
Parameter Value
Alibava filename  run000116.dat
Telescope filename  run000116.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 39 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y 75.93, 63.00
Region of interest bias line
Trigger frequency  210 Hz
Meas. time 15 min
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.2%

 

Entry  Wed Oct 18 13:38:47 2017, Peter Paulitsch, IFX_CenterBias, run000112, 112, 3010000, 122.9, 24.25, -3.3°, No, Good 
run settings
Parameter Value
Alibava filename  run000112.dat
Telescope filename  run000112.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -160 nA
Sensor temperature 24.6 °C
Latency 140 [a.u.]
Alibava triggers 3010000
Telescope triggers 3000000
Data synced yes
Table position x/y 122.9, 24.25
Region of interest bias line
Trigger frequency  157 Hz
Meas. time 320 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 53.8%

 

Entry  Wed Oct 18 13:36:50 2017, Peter Paulitsch, IFX_CenterBias, run000109, 109, 110000, 122.9, 25.75, -3.3°, No, Good 
run settings
Parameter Value
Alibava filename  run000109.dat
Telescope filename  run000109.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -160 nA
Sensor temperature 24.6 °C
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y 122.9, 25.75
Region of interest bias line
Trigger frequency  200 Hz
Meas. time 15 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 53.4%

 

Entry  Wed Oct 18 13:34:54 2017, Peter Paulitsch, IFX_CenterBias, run000107, 107, 110000, 122.9, 23.5, -3.3°, No, Good 
run settings
Parameter Value
Alibava filename  run000107.dat
Telescope filename  run000107.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -160 nA
Sensor temperature 24.6 °C
Latency 140 [a.u.]
Alibava triggers 110000
Telescope triggers 100000
Data synced yes
Table position x/y 122.9, 23.5
Region of interest bias line
Trigger frequency  200 Hz
Meas. time 15 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 53.4%

 

Entry  Wed Oct 18 13:33:05 2017, Peter Paulitsch, IFX_CenterBias, run000106, 106, 60000, 122.9, 23.5, -3.3°, No, Good 
run settings
Parameter Value
Alibava filename  run000106.dat
Telescope filename  run000106.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -160 nA
Sensor temperature 24.6 °C
Latency 140 [a.u.]
Alibava triggers 60000
Telescope triggers 50000
Data synced yes
Table position x/y 122.9, 23.5
Region of interest bias line
Trigger frequency  200 Hz
Meas. time 8 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 53.4%

 