61
|
Mon Nov 30 19:01:54 2015 |
Hao Yin | Belle II | system | Hardware_Cold | |
ADC Cold diff to hot: 1-3 adc delay config (max 1.5 ns), FIRRun_Cold_001
Temp = 0°C
target: CE
|
Attachment 1: Cold_ce_n.png
|
|
Attachment 2: Cold_ce_p.png
|
|
|
62
|
Mon Nov 30 19:48:05 2015 |
Hao Yin | Belle II | system | Hardware_Room_With_Cold_I2C | |
ADC Cold diff to hot: 1-3 adc delay config (max 1.5 ns), FIRRun_Cold_001
with room temp i2c config
Temp = 0°C
target: CE
|
Attachment 1: RoomTemp_with_cold_config_ce_p.png
|
|
Attachment 2: RoomTemp_with_cold_config_ce_n.png
|
|
|
63
|
Mon Nov 30 20:06:15 2015 |
Hao Yin | Belle II | system | Hardware_Room_With_Cold_I2C | |
ADC Cold diff to hot: 1-3 adc delay config (max 1.5 ns), FIRRun_Cold_001
with room temp i2c config
Temp = room
target: -Z
|
Attachment 1: RoomTemp_with_cold_config_-z_p.png
|
|
Attachment 2: RoomTemp_with_cold_config_-z_n.png
|
|
|
64
|
Mon Nov 30 20:16:37 2015 |
Hao Yin | Belle II | system | Hardware_Room_With_Cold_I2C | |
ADC Cold diff to hot: 1-3 adc delay config (max 1.5 ns), FIRRun_Cold_001
ADC delay shifted by 3 (1.5ns) all sensors
with room temp i2c config
Temp = room
target: -Z
|
Attachment 1: RoomTemp_with_cold_config_-z_p_3.png
|
|
Attachment 2: RoomTemp_with_cold_config_-z_n_3.png
|
|
|
65
|
Mon Nov 30 20:27:42 2015 |
Hao Yin | Belle II | system | Hardware_Room_With_Cold_I2C | |
ADC Cold diff to hot: 1-3 adc delay config (max 1.5 ns), FIRRun_Cold_001
ADC shifted by 5 (2.5ns) all sensors
with room temp i2c config
Temp = room
target: -Z
Note : n-side (strip nr 200 - 220 shadowhits) |
Attachment 1: RoomTemp_with_cold_config_-z_p_5.png
|
|
Attachment 2: RoomTemp_with_cold_config_-z_n_5.png
|
|
|
67
|
Mon Nov 30 22:35:41 2015 |
Hao Yin | Belle II | system | Hardware_Room_With_Cold_I2C | |
ADC Cold diff to hot: 1-3 adc delay config (max 1.5 ns), FIRRun_Cold_001
ADC shifted by 9 (4.5ns) all sensors
with room temp i2c config
Temp = room
target: -Z |
Attachment 1: RoomTemp_with_cold_config_-z_p_9.png
|
|
Attachment 2: RoomTemp_with_cold_config_-z_n_9.png
|
|
|
18
|
Wed May 7 14:36:29 2014 |
Benedikt Würkner | Belle II | source | L3 module | No 1 of 6 room temperature measurements using Sr90 Source (multi6) |
Sr90 Radioactive Source
No cooling
Module position: -5:-20(top left)
HV Turned to +40,-40V with currents. +1.2 and -1.2 µA
N+1.25V current: 0.22A
N+2.5 V current: 0.56A
P+1.25V current: 0.26A
P+2.50V current: 0.70A
Events: 200000 |
Attachment 1: run01_all_hits_apv0.png
|
|
Attachment 2: run01_all_hits_apv1.png
|
|
Attachment 3: run01_all_hits_apv2.png
|
|
Attachment 4: run01_all_hits_apv3.png
|
|
Attachment 5: run01_all_hits_apv4.png
|
|
Attachment 6: run01_all_hits_apv5.png
|
|
Attachment 7: run01_all_hits_apv6.png
|
|
Attachment 8: run01_all_hits_apv7.png
|
|
Attachment 9: run01_all_hits_apv8.png
|
|
Attachment 10: run01_all_hits_apv9.png
|
|
|
19
|
Wed May 7 15:04:32 2014 |
Benedikt Würkner | Belle II | source | L3 module | No 2 of 6 room temperature measurements using Sr90 Source (multi6) |
Sr90 Radioactive Source
No cooling
Module position: -32:-20(top center)
HV Turned to +40,-40V with currents. +1.23 and -1.25 µA
N+1.25V current: 0.22A
N+2.50V current: 0.56A
P+1.25V current: 0.26A
P+2.50V current: 0.71A
Events: 200000 |
Attachment 1: run02_all_hits_apv0.png
|
|
Attachment 2: run02_all_hits_apv1.png
|
|
Attachment 3: run02_all_hits_apv2.png
|
|
Attachment 4: run02_all_hits_apv3.png
|
|
Attachment 5: run02_all_hits_apv4.png
|
|
Attachment 6: run02_all_hits_apv5.png
|
|
Attachment 7: run02_all_hits_apv6.png
|
|
Attachment 8: run02_all_hits_apv7.png
|
|
Attachment 9: run02_all_hits_apv8.png
|
|
Attachment 10: run02_all_hits_apv9.png
|
|
|
20
|
Wed May 7 15:53:38 2014 |
Benedikt Würkner | Belle II | source | L3 module | No 3 of 6 room temperature measurements using Sr90 Source (multi6) |
Sr90 Radioactive Source
No cooling
Module position: -59:-20(top right)
HV Turned to +40,-40V with currents. +1.25 and -1.26 µA
N+1.25V current: 0.22A
N+2.50V current: 0.56A
P+1.25V current: 0.26A
P+2.50V current: 0.71A
Events: 200000 |
Attachment 1: run03_all_hits_apv0.png
|
|
Attachment 2: run03_all_hits_apv1.png
|
|
Attachment 3: run03_all_hits_apv2.png
|
|
Attachment 4: run03_all_hits_apv3.png
|
|
Attachment 5: run03_all_hits_apv4.png
|
|
Attachment 6: run03_all_hits_apv5.png
|
|
Attachment 7: run03_all_hits_apv6.png
|
|
Attachment 8: run03_all_hits_apv7.png
|
|
Attachment 9: run03_all_hits_apv8.png
|
|
Attachment 10: run03_all_hits_apv9.png
|
|
|
21
|
Wed May 7 16:43:37 2014 |
Benedikt Würkner | Belle II | source | L3 module | No 4 of 6 room temperature measurements using Sr90 Source (multi6) |
Sr90 Radioactive Source
No cooling
Module position: -5:-30(bottom left)
HV Turned to +40,-40V with currents. +1.25 and -1.26 µA
N+1.25V current: 0.22A
N+2.50V current: 0.57A
P+1.25V current: 0.26A
P+2.50V current: 0.72A
Events: 200000 |
Attachment 1: run04_all_hits_apv0.png
|
|
Attachment 2: run04_all_hits_apv1.png
|
|
Attachment 3: run04_all_hits_apv2.png
|
|
Attachment 4: run04_all_hits_apv3.png
|
|
Attachment 5: run04_all_hits_apv4.png
|
|
Attachment 6: run04_all_hits_apv5.png
|
|
Attachment 7: run04_all_hits_apv6.png
|
|
Attachment 8: run04_all_hits_apv7.png
|
|
Attachment 9: run04_all_hits_apv8.png
|
|
Attachment 10: run04_all_hits_apv9.png
|
|
|
22
|
Wed May 7 17:33:21 2014 |
Benedikt Würkner | Belle II | source | L3 module | No 5 of 6 room temperature measurements using Sr90 Source (multi6) |
Sr90 Radioactive Source
No cooling
Module position: -32:-30(bottom center)
HV Turned to +40,-40V with currents. +1.22 and -1.24 µA
N+1.25V current: 0.22A
N+2.50V current: 0.56A
P+1.25V current: 0.26A
P+2.50V current: 0.70A
Events: 200000 |
Attachment 1: run05_all_hits_apv0.png
|
|
Attachment 2: run05_all_hits_apv1.png
|
|
Attachment 3: run05_all_hits_apv2.png
|
|
Attachment 4: run05_all_hits_apv3.png
|
|
Attachment 5: run05_all_hits_apv4.png
|
|
Attachment 6: run05_all_hits_apv5.png
|
|
Attachment 7: run05_all_hits_apv6.png
|
|
Attachment 8: run05_all_hits_apv7.png
|
|
Attachment 9: run05_all_hits_apv8.png
|
|
Attachment 10: run05_all_hits_apv9.png
|
|
|
23
|
Wed May 7 18:20:36 2014 |
Benedikt Würkner | Belle II | source | L3 module | No 6 of 6 room temperature measurements using Sr90 Source (multi6) |
Sr90 Radioactive Source
No cooling
Module position: -57:-30(bottom right)
HV Turned to +40,-40V with currents. +1.2 and -1.22 µA
N+1.25V current: 0.22A
N+2.50V current: 0.56A
P+1.25V current: 0.26A
P+2.50V current: 0.70A
Events: 200000 |
Attachment 1: run06_all_hits_apv0.png
|
|
Attachment 2: run06_all_hits_apv1.png
|
|
Attachment 3: run06_all_hits_apv2.png
|
|
Attachment 4: run06_all_hits_apv3.png
|
|
Attachment 5: run06_all_hits_apv4.png
|
|
Attachment 6: run06_all_hits_apv5.png
|
|
Attachment 7: run06_all_hits_apv6.png
|
|
Attachment 8: run06_all_hits_apv7.png
|
|
Attachment 9: run06_all_hits_apv8.png
|
|
Attachment 10: run06_all_hits_apv9.png
|
|
|
26
|
Thu May 8 16:51:03 2014 |
Benedikt Würkner | Belle II | source | L3 module | No 1 of 6 room temperature measurements using Sr90 Source (single) |
Sr90 Radioactive Source
No cooling
Module position: 0:0(top left)
HV Turned to +40,-40V with currents. +1.0 and -1.0 µA
N+1.25V current: 0.22A
N+2.50V current: 0.57A
P+1.25V current: 0.26A
P+2.50V current: 0.72A
Events: 200000 |
Attachment 1: run07_all_hits_apv0.png
|
|
Attachment 2: run07_all_hits_apv1.png
|
|
Attachment 3: run07_all_hits_apv2.png
|
|
Attachment 4: run07_all_hits_apv3.png
|
|
Attachment 5: run07_all_hits_apv4.png
|
|
Attachment 6: run07_all_hits_apv5.png
|
|
Attachment 7: run07_all_hits_apv6.png
|
|
Attachment 8: run07_all_hits_apv7.png
|
|
Attachment 9: run07_all_hits_apv8.png
|
|
Attachment 10: run07_all_hits_apv9.png
|
|
|
27
|
Thu May 8 16:51:57 2014 |
Benedikt Würkner | Belle II | source | L3 module | No 2 of 6 room temperature measurements using Sr90 Source (single) |
Failed with Error: "***ERROR*** Init NECO failed, Return code: -32761"
Restarted and it works again
Sr90 Radioactive Source
No cooling
Module position: -27:0(top center)
HV Turned to +40,-40V with currents. +0.97 and -1.0 µA
N+1.25V current: 0.22A
N+2.50V current: 0.57A
P+1.25V current: 0.26A
P+2.50V current: 0.73A
Events: 200000 |
|
28
|
Fri May 9 15:12:44 2014 |
Benedikt Würkner | Belle II | source | L3 module | No 2 of 6 room temperature measurements using Sr90 Source (single) |
Sr90 Radioactive Source
No cooling
Module position: -27:0(top center)
HV Turned to +40,-40V with currents. +0.95 and -0.98 µA
N+1.25V current: 0.22A
N+2.50V current: 0.57A
P+1.25V current: 0.26A
P+2.50V current: 0.72A
Events: 200000 |
Attachment 1: run08_3rd_all_hits_apv0.png
|
|
Attachment 2: run08_3rd_all_hits_apv1.png
|
|
Attachment 3: run08_3rd_all_hits_apv2.png
|
|
Attachment 4: run08_3rd_all_hits_apv3.png
|
|
Attachment 5: run08_3rd_all_hits_apv4.png
|
|
Attachment 6: run08_3rd_all_hits_apv5.png
|
|
Attachment 7: run08_3rd_all_hits_apv6.png
|
|
Attachment 8: run08_3rd_all_hits_apv7.png
|
|
Attachment 9: run08_3rd_all_hits_apv8.png
|
|
Attachment 10: run08_3rd_all_hits_apv9.png
|
|
|
29
|
Fri May 9 16:44:41 2014 |
Benedikt Würkner | Belle II | source | L3 module | No 3 of 6 room temperature measurements using Sr90 Source (single) |
Sr90 Radioactive Source
No cooling
Module position: -54:0(top center)
HV Turned to +40,-40V with currents. +1.2 and -1.22 µA
N+1.25V current: 0.22A
N+2.50V current: 0.54A
P+1.25V current: 0.26A
P+2.50V current: 0.68A
Events: 200000 |
Attachment 1: run09_all_hits_apv0.png
|
|
Attachment 2: run09_all_hits_apv1.png
|
|
Attachment 3: run09_all_hits_apv2.png
|
|
Attachment 4: run09_all_hits_apv3.png
|
|
Attachment 5: run09_all_hits_apv4.png
|
|
Attachment 6: run09_all_hits_apv5.png
|
|
Attachment 7: run09_all_hits_apv6.png
|
|
Attachment 8: run09_all_hits_apv7.png
|
|
Attachment 9: run09_all_hits_apv8.png
|
|
Attachment 10: run09_all_hits_apv9.png
|
|
|
30
|
Fri May 16 11:37:05 2014 |
Benedikt Würkner | Belle II | source | L3 module | No 4 of 6 room temperature measurements using Sr90 Source (single) |
Sr90 Radioactive Source
No cooling
Module position: 0:-10(bottom left)
HV Turned to +40,-40V with currents. +0.85 and -0-9 µA
N+1.25V current: 0.22A
N+2.50V current: 0.57A
P+1.25V current: 0.26A
P+2.50V current: 0.71A
Events: 200000 |
Attachment 1: run10_all_hits_apv0.png
|
|
Attachment 2: run10_all_hits_apv1.png
|
|
Attachment 3: run10_all_hits_apv2.png
|
|
Attachment 4: run10_all_hits_apv3.png
|
|
Attachment 5: run10_all_hits_apv4.png
|
|
Attachment 6: run10_all_hits_apv5.png
|
|
Attachment 7: run10_all_hits_apv6.png
|
|
Attachment 8: run10_all_hits_apv7.png
|
|
Attachment 9: run10_all_hits_apv8.png
|
|
Attachment 10: run10_all_hits_apv9.png
|
|
|
31
|
Fri May 16 11:38:44 2014 |
Benedikt Würkner | Belle II | source | L3 module | No 5 of 6 room temperature measurements using Sr90 Source (single) |
Sr90 Radioactive Source
No cooling
Module position: -27:-10(bottom center)
HV Turned to +40,-40V with currents. +1.05 and -1.1 µA
N+1.25V current: 0.22A
N+2.50V current: 0.55A
P+1.25V current: 0.26A
P+2.50V current: 0.67A
Events: 200000 |
Attachment 1: run11_all_hits_apv0.png
|
|
Attachment 2: run11_all_hits_apv1.png
|
|
Attachment 3: run11_all_hits_apv2.png
|
|
Attachment 4: run11_all_hits_apv3.png
|
|
Attachment 5: run11_all_hits_apv4.png
|
|
Attachment 6: run11_all_hits_apv5.png
|
|
Attachment 7: run11_all_hits_apv6.png
|
|
Attachment 8: run11_all_hits_apv7.png
|
|
Attachment 9: run11_all_hits_apv8.png
|
|
Attachment 10: run11_all_hits_apv9.png
|
|
|
32
|
Fri May 16 11:40:29 2014 |
Benedikt Würkner | Belle II | source | L3 module | No 6 of 6 room temperature measurements using Sr90 Source (single) |
Sr90 Radioactive Source
No cooling
Module position: -54:-10(bottom right)
HV Turned to +40,-40V with currents. +1.1 and -1.1 µA
N+1.25V current: 0.22A
N+2.50V current: 0.55A
P+1.25V current: 0.26A
P+2.50V current: 0.67A
Events: 200000
|
Attachment 1: run12_all_hits_apv0.png
|
|
Attachment 2: run12_all_hits_apv1.png
|
|
Attachment 3: run12_all_hits_apv2.png
|
|
Attachment 4: run12_all_hits_apv3.png
|
|
Attachment 5: run12_all_hits_apv4.png
|
|
Attachment 6: run12_all_hits_apv5.png
|
|
Attachment 7: run12_all_hits_apv6.png
|
|
Attachment 8: run12_all_hits_apv7.png
|
|
Attachment 9: run12_all_hits_apv8.png
|
|
Attachment 10: run12_all_hits_apv9.png
|
|
|
47
|
Mon May 25 10:38:03 2015 |
Hao Yin | Belle II | system | L5 Ladder Emission Test | Emission |
(TestNr: 16)
base line noise measurement are saved in SYSTEM_CHECK_AFTER_WE.
(noise wings at fw n side decreased !!!)
Measuring noise emitted by our system.
1) noise emmision at the cable from dockbox to the hybrid (fw n and p side)
|