ID |
Date |
Author |
Project |
Measurement Type |
Object ID |
Subject |
18
|
Wed May 7 14:36:29 2014 |
Benedikt Würkner | Belle II | source | L3 module | No 1 of 6 room temperature measurements using Sr90 Source (multi6) |
Sr90 Radioactive Source
No cooling
Module position: -5:-20(top left)
HV Turned to +40,-40V with currents. +1.2 and -1.2 µA
N+1.25V current: 0.22A
N+2.5 V current: 0.56A
P+1.25V current: 0.26A
P+2.50V current: 0.70A
Events: 200000 |
19
|
Wed May 7 15:04:32 2014 |
Benedikt Würkner | Belle II | source | L3 module | No 2 of 6 room temperature measurements using Sr90 Source (multi6) |
Sr90 Radioactive Source
No cooling
Module position: -32:-20(top center)
HV Turned to +40,-40V with currents. +1.23 and -1.25 µA
N+1.25V current: 0.22A
N+2.50V current: 0.56A
P+1.25V current: 0.26A
P+2.50V current: 0.71A
Events: 200000 |
20
|
Wed May 7 15:53:38 2014 |
Benedikt Würkner | Belle II | source | L3 module | No 3 of 6 room temperature measurements using Sr90 Source (multi6) |
Sr90 Radioactive Source
No cooling
Module position: -59:-20(top right)
HV Turned to +40,-40V with currents. +1.25 and -1.26 µA
N+1.25V current: 0.22A
N+2.50V current: 0.56A
P+1.25V current: 0.26A
P+2.50V current: 0.71A
Events: 200000 |
21
|
Wed May 7 16:43:37 2014 |
Benedikt Würkner | Belle II | source | L3 module | No 4 of 6 room temperature measurements using Sr90 Source (multi6) |
Sr90 Radioactive Source
No cooling
Module position: -5:-30(bottom left)
HV Turned to +40,-40V with currents. +1.25 and -1.26 µA
N+1.25V current: 0.22A
N+2.50V current: 0.57A
P+1.25V current: 0.26A
P+2.50V current: 0.72A
Events: 200000 |
22
|
Wed May 7 17:33:21 2014 |
Benedikt Würkner | Belle II | source | L3 module | No 5 of 6 room temperature measurements using Sr90 Source (multi6) |
Sr90 Radioactive Source
No cooling
Module position: -32:-30(bottom center)
HV Turned to +40,-40V with currents. +1.22 and -1.24 µA
N+1.25V current: 0.22A
N+2.50V current: 0.56A
P+1.25V current: 0.26A
P+2.50V current: 0.70A
Events: 200000 |
23
|
Wed May 7 18:20:36 2014 |
Benedikt Würkner | Belle II | source | L3 module | No 6 of 6 room temperature measurements using Sr90 Source (multi6) |
Sr90 Radioactive Source
No cooling
Module position: -57:-30(bottom right)
HV Turned to +40,-40V with currents. +1.2 and -1.22 µA
N+1.25V current: 0.22A
N+2.50V current: 0.56A
P+1.25V current: 0.26A
P+2.50V current: 0.70A
Events: 200000 |
26
|
Thu May 8 16:51:03 2014 |
Benedikt Würkner | Belle II | source | L3 module | No 1 of 6 room temperature measurements using Sr90 Source (single) |
Sr90 Radioactive Source
No cooling
Module position: 0:0(top left)
HV Turned to +40,-40V with currents. +1.0 and -1.0 µA
N+1.25V current: 0.22A
N+2.50V current: 0.57A
P+1.25V current: 0.26A
P+2.50V current: 0.72A
Events: 200000 |
27
|
Thu May 8 16:51:57 2014 |
Benedikt Würkner | Belle II | source | L3 module | No 2 of 6 room temperature measurements using Sr90 Source (single) |
Failed with Error: "***ERROR*** Init NECO failed, Return code: -32761"
Restarted and it works again
Sr90 Radioactive Source
No cooling
Module position: -27:0(top center)
HV Turned to +40,-40V with currents. +0.97 and -1.0 µA
N+1.25V current: 0.22A
N+2.50V current: 0.57A
P+1.25V current: 0.26A
P+2.50V current: 0.73A
Events: 200000 |
28
|
Fri May 9 15:12:44 2014 |
Benedikt Würkner | Belle II | source | L3 module | No 2 of 6 room temperature measurements using Sr90 Source (single) |
Sr90 Radioactive Source
No cooling
Module position: -27:0(top center)
HV Turned to +40,-40V with currents. +0.95 and -0.98 µA
N+1.25V current: 0.22A
N+2.50V current: 0.57A
P+1.25V current: 0.26A
P+2.50V current: 0.72A
Events: 200000 |
29
|
Fri May 9 16:44:41 2014 |
Benedikt Würkner | Belle II | source | L3 module | No 3 of 6 room temperature measurements using Sr90 Source (single) |
Sr90 Radioactive Source
No cooling
Module position: -54:0(top center)
HV Turned to +40,-40V with currents. +1.2 and -1.22 µA
N+1.25V current: 0.22A
N+2.50V current: 0.54A
P+1.25V current: 0.26A
P+2.50V current: 0.68A
Events: 200000 |
30
|
Fri May 16 11:37:05 2014 |
Benedikt Würkner | Belle II | source | L3 module | No 4 of 6 room temperature measurements using Sr90 Source (single) |
Sr90 Radioactive Source
No cooling
Module position: 0:-10(bottom left)
HV Turned to +40,-40V with currents. +0.85 and -0-9 µA
N+1.25V current: 0.22A
N+2.50V current: 0.57A
P+1.25V current: 0.26A
P+2.50V current: 0.71A
Events: 200000 |
31
|
Fri May 16 11:38:44 2014 |
Benedikt Würkner | Belle II | source | L3 module | No 5 of 6 room temperature measurements using Sr90 Source (single) |
Sr90 Radioactive Source
No cooling
Module position: -27:-10(bottom center)
HV Turned to +40,-40V with currents. +1.05 and -1.1 µA
N+1.25V current: 0.22A
N+2.50V current: 0.55A
P+1.25V current: 0.26A
P+2.50V current: 0.67A
Events: 200000 |
32
|
Fri May 16 11:40:29 2014 |
Benedikt Würkner | Belle II | source | L3 module | No 6 of 6 room temperature measurements using Sr90 Source (single) |
Sr90 Radioactive Source
No cooling
Module position: -54:-10(bottom right)
HV Turned to +40,-40V with currents. +1.1 and -1.1 µA
N+1.25V current: 0.22A
N+2.50V current: 0.55A
P+1.25V current: 0.26A
P+2.50V current: 0.67A
Events: 200000
|
33
|
Thu Jun 5 10:33:46 2014 |
Benedikt Würkner | Belle II | source | Silc Module | Silc Angle Measurement 1° |
Measured the Silc 03/10 Module using the Sr90 Source to have a comparison for the Eta-Distribution at different angles.
Data can be found on heros in: /home/medialib/LAB_Silc_Angle.
Plots made with TuxOA for all different regions can be found in /home/users/bwuerkner/plots/.
|
34
|
Thu Jun 5 10:34:06 2014 |
Benedikt Würkner | Belle II | source | Silc Module | Silc Angle Measurement 4° |
Measured the Silc 03/10 Module using the Sr90 Source to have a comparison for the Eta-Distribution at different angles.
Data can be found on heros in: /home/medialib/LAB_Silc_Angle.
Plots made with TuxOA for all different regions can be found in /home/users/bwuerkner/plots/.
|
35
|
Thu Jun 5 10:34:29 2014 |
Benedikt Würkner | Belle II | source | Silc Module | Silc Angle Measurement 7° |
Measured the Silc 03/10 Module using the Sr90 Source to have a comparison for the Eta-Distribution at different angles.
Data can be found on heros in: /home/medialib/LAB_Silc_Angle.
Plots made with TuxOA for all different regions can be found in /home/users/bwuerkner/plots/.
|
36
|
Thu Jun 5 10:34:45 2014 |
Benedikt Würkner | Belle II | source | Silc Module | Silc Angle Measurement 10° |
Measured the Silc 03/10 Module using the Sr90 Source to have a comparison for the Eta-Distribution at different angles.
Data can be found on heros in: /home/medialib/LAB_Silc_Angle.
Plots made with TuxOA for all different regions can be found in /home/users/bwuerkner/plots/.
|
3
|
Wed Apr 23 13:05:05 2008 |
Christian Irmler | SPS Testbeam June08 | source | hybrid 01 | time correlation between TDC and sensor measurement -> 3.1 ns RMS |
|
6
|
Wed May 7 15:24:49 2008 |
Christian Irmler | SPS Testbeam June08 | module | module 10/02 | properties (noise, intcal), APVs bonded to the sensor |
Module tested with 1 and 2 rows bonded to the sensor, respectively.
HV = 100 V
Ibias (100V) = 19.1 nA
Ibias (200V) = 23.7 nA |
9
|
Wed May 7 15:26:24 2008 |
Christian Irmler | SPS Testbeam June08 | module | module 20/09 | properties (noise, intcal), APVs bonded to the sensor |
Module tested with 1 and 2 rows bonded to the sensor, respectively.
HV = 100 V
Ibias (100 V) = 25.1 nA
Ibias (200 V) = 31.4 nA |