Home
CMS Production
Clean room
MedAustron
HEPHY
testbeams
old
testbeam-rd50-desy-Jul2025
testbeam-rd50-desy-Oct2024
testbeam-RD50-DESY-Apr2024
testbeam-RD50-DESY-Jul2023
testbeam-RD50-CERN-Oct2022
testbeam-DESY-Oct2017
testbeam-DESY-April2017
testbeam-SPS2015
testbeam-SPS2014
testbeam-DESY14
testbeam-SPS12
testbeam-SPS11
testbeam-SPS10
testbeam-SPS09
testbeam-SPS08
LP-TPC
Testbeam at DESY in July 2025, Page 4 of 4
Not logged in
Find
|
Login
|
Help
Full |
Summary
|
Threaded
|
Hide attachments
-- All entries --
Last day
Last 3 Days
Last week
Last month
Last 3 Months
Last 6 Months
Last Year
-- DUT --
W3-2 5E14
W3-x1 1E15
W8-2 1E15
W3-5 2E15
W8-3 2E15
W3-7 3E15
W8-x 3E15
W3-x 5E15
W3-1 5E14
W3-x2 1E15
W3-6 2E15
W3-1x3 3E15
W8-x 3E15
W3-x 5E15
W3 0E00
W8 0E00
-- Beam Energy --
1.0
1.2
1.4
1.6
1.8
2.0
2.2
2.4
2.6
2.8
3.0
3.2
3.4
3.6
3.8
4.0
4.2
4.4
4.6
4.8
5.0
5.2
5.4
5.6
5.8
6.0
74 Entries
Goto page
Previous
1
,
2
,
3
, 4
All
ID
Date
Author
Subject
Run Number
DUT
Biasing
Beam Energy
thr
HV
scan-val
Data
t-Start
t-Stop
Duration_min
5
Mon Jun 30 17:32:39 2025
tb-crew
test
W3 0E00
Backside
4.2
Crap
1st log from hut TB22
28
Tue Jul 1 09:18:15 2025
tb-crew
system test
W3 0E00
Backside
4.2
Crap
changed alignment of DUT
31
Tue Jul 1 11:11:54 2025
tb-crew
system test
W3 0E00
Backside
4.2
Unknown
changed CarBoard due to sync problems
34
Tue Jul 1 11:20:36 2025
tb-crew
system test
W3 0E00
Backside
4.2
0.95
-190
Unknown
seeing Correlations for first time
36
Tue Jul 1 11:22:25 2025
tb-crew
system test
W3 0E00
Backside
4.2
0.95
-190
Unknown
change of CaR-board led to correlations
Still unclear why we have to ignore DUT trigger veto
38
Tue Jul 1 13:55:45 2025
tb-crew
system test
W3 0E00
Backside
4.2
0.95
-190
Unknown
got 1st "nice" efficiency results, 87% with poor sw-alignment
51
Tue Jul 1 16:39:57 2025
tb-crew
system test
W3 0E00
Backside
4.2
0.95
-190
Unknown
moved stage to get telescope centered properly
60
Tue Jul 1 17:09:51 2025
tb-crew
system test
W3 0E00
Backside
4.2
0.95
-190
Unknown
might have found good ROI
62
Tue Jul 1 17:22:02 2025
tb-crew
system test
W3 0E00
Backside
4.2
0.95
-190
Unknown
ROI is nice now, efficiency still a bit crappy with 95%
65
Tue Jul 1 18:48:36 2025
tb-crew
system test
W8 0E00
Backside
4.2
Unknown
changed DUT - hoping for full efficiency with sample not carrying thermocouple
67
Tue Jul 1 19:17:46 2025
tb-crew
system test
W8 0E00
Backside
4.2
0.95
-190
Unknown
that was noise
68
Tue Jul 1 19:20:18 2025
tb-crew
system test
W8 0E00
Backside
4.2
0.95
-190
Unknown
using Christophers calibration, no more noise
71
Tue Jul 1 23:34:29 2025
tb-crew
system test
W8 0E00
Backside
4.2
0.95
-190
Unknown
Run 63 is 99,9% efficiency with ts_mode=TLU
bug in MPW4 config (conf_reg_ts_ctrl)
74
Wed Jul 2 00:37:34 2025
tb-crew
system test
W3 0E00
Backside
4.2
0.95
-190
Unknown
going to bed