Home CMS Production Clean room MedAustron HEPHY testbeams old
testbeam-RD50-DESY-Jul2023 testbeam-RD50-CERN-Oct2022 testbeam-DESY-Oct2017 testbeam-DESY-April2017 testbeam-SPS2015 testbeam-SPS2014 testbeam-DESY14 testbeam-SPS12 testbeam-SPS11 testbeam-SPS10 testbeam-SPS09 testbeam-SPS08 LP-TPC
  Belle/Infineon/AIDA testbeam at SPS in Nov 2014, Page 6 of 6  Not logged in ELOG logo
ID Date Author Project Subject Run Number Eventsdown StartTime EndTime Data
  64   Thu Nov 20 17:16:56 2014 Axel Königcommonhow to run the EUDAQ offline monitor    Unknown

YOU need

1.) Folder EUDAQ OFFLINE Monitor

2.) ROOT

 

Starting: 

1.) open command line and chose the path where OfflineMonitor is

2.)Type -i "patch of raw file" -o test.root 100

3.) helpfile typing -h

  66   Thu Nov 20 22:11:06 2014 Hao YinBelle SVDHardwareRun012   Good

cooled see appendix (FirRun003)

PedestalRun_009

multi 6 Raw Mode

FIFO size increased

  67   Thu Nov 20 22:12:21 2014 Axel KönigcommonSPS Logbook / SPS Online Monitor    Unknown

Logbook: https://ab-dep-op-elogbook.web.cern.ch/ab-dep-op-elogbook/elogbook/secure/eLogbook.php?shiftId=1059325

 

Monitor:  http://op-webtools.web.cern.ch/op-webtools/vistar/vistars.php?usr=SPS1

  68   Fri Nov 21 01:43:54 2014 Axel KönigAIDAcalibration runAIDA_run001_calibration   Unknown

Calibration run with AIDA DUT.

Single hit

  78   Fri Nov 21 18:23:05 2014 Axel KönigcommonNew Table Calibration    Unknown

 Nach powercut musste der Tisch neu kalibriert werden. Diese Neukalibration betrifft alle Messungen ab den ersten Messungen vom AIDA Stack.

  79   Fri Nov 21 18:35:17 2014 Axel KönigAIDAImportant: Dockbox Inverter!    Unknown

Alle AIDA runs vor run_007 hatten kein Signal auf den Sensoren 5 und 6.

Grund: Inverter auf N Seite war im config file noch an (von Markus behoben). 

Runs vor 007 können noch verwendet werden obwohl sensoren 5 und 6 negative Events lieferten (kommt aber auf das negative Limit an). 

  80   Fri Nov 21 19:05:27 2014 Hao YinBelle SVDFIRRun_004  Fri Nov 21 19:05:23 CET 2014 Unknown

New ADC delay scan. cooled, -25 temperature

 

  81   Fri Nov 21 19:10:09 2014 Hao YinBelle SVDCalibrationRun_006    Unknown

cooled, latest ADC delay scan and FIRRun. 

  83   Fri Nov 21 19:42:36 2014 Hao Yin HardwareRun_033    Good

L6 are swapped, needs to be corrected later in the offline analysis.

cooled at -25

latest ADC delay and FIRRun

PedestalRun_012

CalibrationRun_006

100k trigger dead time

 

Need to change L6n apv definition to:

mod = 9,0,1,1,1,m,29,0,0,0,0,0,0,L6_n_side

apv = 9,1,34,3,30,x,0,0,5,1,7,0,0,x
apv = 9,0,36,3,50,x,0,0,5,1,6,0,0,x
apv = 9,3,38,3,30,x,0,0,5,1,9,0,0,x
apv = 9,2,40,3,30,x,0,0,5,1,8,0,0,x

to  get the right mapping of the sensor.

Attachment 1: Temp_21-11-14.PNG
Temp_21-11-14.PNG
  91   Sat Nov 22 10:20:07 2014 Axel KönigAIDAcalibration runAIDA_calibration   Unknown

Calibration run AIDA_calibration 

L:\cern14\AIDA

  92   Sat Nov 22 11:16:26 2014 Markus FriedlBelle SVDcooling    Unknown

Cooling box plant hung up (I2C error on Alix, persistent after reboot)

Power-cycled cooling plant, now working

Slowly cooling down again

  95   Sat Nov 22 16:04:05 2014 Axel KönigInfineonInfineon old calibrationInfineon_old_calibration   Unknown

Calibration run 

L:\cern14\infineon_old

  96   Sat Nov 22 16:43:18 2014 Axel Königcommontable, telescope and dut geometry     Unknown

geometry files

Attachment 1: geometry_table.pdf
Attachment 2: geometry_DUT.pdf
Attachment 3: geometry_infineon_batch4.pdf
Attachment 4: geometry_infineon_old.pdf
  98   Sat Nov 22 17:00:43 2014 Axel KönigInfineonInfineon olddamaged strixel sensors, strange IV behaviour   Unknown

1.) No events recorded for strixel sensors (run001). They seem to be damaged like they were at DESY.

2.) Very strange IV behaviour for every old Infineon module (especially TB12_baby1). 

  100   Sat Nov 22 19:29:40 2014 Axel KönigAIDADUT Position    Unknown
Attachment 1: AIDA_DUT_Position.JPG
AIDA_DUT_Position.JPG
  109   Sun Nov 23 09:49:00 2014 Markus FriedlBelle SVDcooling    Unknown

Saturday evening, the cooling became unstable which indicates the end of the CO2 supply, possibly having a mixture of liquid and gas phases.

Strange oscillations observed for the DUT temperatures until the MF valve was finally turned off close to midnight.

 

Attachment 1: temp_last_night.png
temp_last_night.png
  113   Sun Nov 23 12:39:45 2014 Axel KönigcommonPMT's of EUDET (Igor Rubinsky)    Unknown

Der zweite PM (entsprechend der Beamrichtung) war von Anfang an nicht angeschlossen an die TLU! Stattdessen war ein Atlas pixel sensor, welcher als 4ter PM fungiert angeschlossen. Genau dieser "PM" hatte 0 counts bei der TLU --> Igor sagt, dass dieser Sensor ("PM") anscheinend keine Versorgungsspannung hatte (~400V). Jetzt baut Igor diesen Atlas Sensor wieder aus und hat bereits den nicht angeschlossenen PM wieder angeschlossen an die TLU. DIeser PM zeigt aber nun zu viele counts. Deshalb behalten wir die AND mask 11 (auch jetzt bei den Belle Modulen).

  115   Sun Nov 23 15:07:54 2014 Markus FriedlBelle SVDcommon    Unknown

Geometry setup of FW/BW:

(all dimensions are in mm and refer to the sensor planes)

Attachment 1: fwbw_setup.png
fwbw_setup.png