ID |
Date |
Author |
Project |
Subject |
Run Number |
Events |
StartTime |
EndTime |
Data |
76
|
Fri Nov 21 18:20:47 2014 |
Axel König | AIDA | hardware run | AIDA_run005 | 20045 | | | Unknown |
AIDA
single hit
Voltage=100V, I=5.4357uA
(new table calibration): V:71mm, H:80mm
config: L:\cern14\AIDA
data: L:\cern14\AIDA
EUDAQ File: Run000979
--Stopped 21.11 / 12:12--
DaqEvents: 20045 (-600 ped)
EUDAQ EVENTS: ~18k |
77
|
Fri Nov 21 18:21:45 2014 |
Axel König | AIDA | hardware run | AIDA_run006 | ~1k | | | Unknown |
verwerfen! |
78
|
Fri Nov 21 18:23:05 2014 |
Axel König | common | New Table Calibration | | | | | Unknown |
Nach powercut musste der Tisch neu kalibriert werden. Diese Neukalibration betrifft alle Messungen ab den ersten Messungen vom AIDA Stack. |
79
|
Fri Nov 21 18:35:17 2014 |
Axel König | AIDA | Important: Dockbox Inverter! | | | | | Unknown |
Alle AIDA runs vor run_007 hatten kein Signal auf den Sensoren 5 und 6.
Grund: Inverter auf N Seite war im config file noch an (von Markus behoben).
Runs vor 007 können noch verwendet werden obwohl sensoren 5 und 6 negative Events lieferten (kommt aber auf das negative Limit an). |
80
|
Fri Nov 21 19:05:27 2014 |
Hao Yin | Belle SVD | FIRRun_004 | | | Fri Nov 21 19:05:23 CET 2014 | | Unknown |
New ADC delay scan. cooled, -25 temperature
|
81
|
Fri Nov 21 19:10:09 2014 |
Hao Yin | Belle SVD | CalibrationRun_006 | | | | | Unknown |
cooled, latest ADC delay scan and FIRRun. |
82
|
Fri Nov 21 19:25:57 2014 |
Axel König | AIDA | hardware run | AIDA_run007 | 26770 | | | Unknown |
AIDA
multi hit
Voltage=100V, I=5.4357uA
(new table calibration): V:71mm, H:65mm
config: L:\cern14\AIDA
data: L:\cern14\AIDA
EUDAQ File: Run000983
--Stopped 21.11 / 19:20--
DaqEvents: 26770 (-600 ped)
EUDAQ EVENTS: 26171 (+1 trigger offset) |
84
|
Fri Nov 21 22:08:32 2014 |
Hao Yin | | PedestalRun | 013 | 10000 | | | Unknown |
correct mapping
cooled -25 °C
L3n deactivated
|
85
|
Fri Nov 21 23:33:54 2014 |
Axel König | AIDA | hardware run (PC crashed) | AIDA_run008 | --- | | | Unknown |
22:00 SVD3 PC crashed!!! data will be collected. |
86
|
Fri Nov 21 23:34:36 2014 |
Axel König | AIDA | hardware run | AIDA_run009 | 32902 | | | Unknown |
AIDA
multi hit
Voltage=100V, I=5.2190uA
(new table calibration): V:71mm, H:100mm
config: L:\cern14\AIDA
data: L:\cern14\AIDA
EUDAQ File: Run000983
--Stopped 21.11 / 19:20--
DaqEvents: 32902 (-600 ped)
EUDAQ EVENTS: DAQ events -600 (+1 trigger offset)
|
|
87
|
Fri Nov 21 23:40:10 2014 |
Axel König | AIDA | hardware run | AIDA_run010 | 26329 | | | Unknown |
AIDA
multi hit
Voltage=100V, I=5.3470uA
(new table calibration): V:71mm, H:90mm
config: L:\cern14\AIDA
data: L:\cern14\AIDA
EUDAQ File: run000987
DaqEvents: 26329 (-600 ped)
EUDAQ EVENTS: DAQ events -600 (+1 trigger offset)
|
|
88
|
Sat Nov 22 00:11:36 2014 |
Axel König | AIDA | hardware run | AIDA_run011 | 24916 | | | Unknown |
AIDA
multi hit
Voltage=100V, I=5.3470uA
(new table calibration): V:71mm, H:80mm
config: L:\cern14\AIDA
data: L:\cern14\AIDA
EUDAQ File: run000988
DaqEvents: 24916 (-600 ped)
EUDAQ EVENTS: DAQ events -600 (+1 trigger offset) |
89
|
Sat Nov 22 00:58:04 2014 |
Axel König | AIDA | hardware run | AIDA_run012 | 276006 | | | Unknown |
AIDA
multi hit
Voltage=100V, I=5.3417uA
(new table calibration): V:51mm, H:70mm
config: L:\cern14\AIDA
data: L:\cern14\AIDA
EUDAQ File: run000989
DaqEvents: 276006 (-600 ped)
EUDAQ EVENTS: 275390 |
90
|
Sat Nov 22 07:56:33 2014 |
Axel König | AIDA | hardware run | AIDA_run013 | 55629 | | | Unknown |
AIDA
multi hit
Voltage=100V, I=5.3417uA
(new table calibration): V:51mm, H:45mm
config: L:\cern14\AIDA
data: L:\cern14\AIDA
EUDAQ File: run000990
DaqEvents: 55629 (-600 ped)
EUDAQ EVENTS: 55003
|
|
91
|
Sat Nov 22 10:20:07 2014 |
Axel König | AIDA | calibration run | AIDA_calibration | | | | Unknown |
Calibration run AIDA_calibration
L:\cern14\AIDA |
92
|
Sat Nov 22 11:16:26 2014 |
Markus Friedl | Belle SVD | cooling | | | | | Unknown |
Cooling box plant hung up (I2C error on Alix, persistent after reboot)
Power-cycled cooling plant, now working
Slowly cooling down again |
93
|
Sat Nov 22 11:37:16 2014 |
Benedikt Würkner | Belle SVD | HardwareRun | 036 | 450378 | 2014-11-22 1:00:00 | 2014-11-22 10:27:54 | Unknown |
Not completely cooled since cooling failed at 1:42 in the night
Bias voltage set to +-60V
Some error with event no17 or 18 leading to the data not being easily analysable requiring a hack in TuxOA
Pedestal_013 (cooled)
Noisefile: HardwareRun_036.noi |
Attachment 1: SPS_Testbeam-Temp.xlsx
|
Attachment 2: Screen_Shot_2014-11-22_at_12.41.44.png
|
|
94
|
Sat Nov 22 15:31:49 2014 |
Hao Yin | Belle SVD | HardwareRun | 038 | 43788 | 2014-11-22 15:18:40 | | Unknown |
Cooled down to -20
Bias voltage set to +-60V
Pedestal_013 (cooled)
Noisefile: HardwareRun_036.noi
Zero suppressed |
95
|
Sat Nov 22 16:04:05 2014 |
Axel König | Infineon | Infineon old calibration | Infineon_old_calibration | | | | Unknown |
Calibration run
L:\cern14\infineon_old |
96
|
Sat Nov 22 16:43:18 2014 |
Axel König | common | table, telescope and dut geometry | | | | | Unknown |
geometry files |
Attachment 1: geometry_table.pdf
|
Attachment 2: geometry_DUT.pdf
|
Attachment 3: geometry_infineon_batch4.pdf
|
Attachment 4: geometry_infineon_old.pdf
|