| ID | 
Date | 
Author | 
Subject | 
Run Number | 
DUT | 
Wafer | 
HV | 
Fluence | 
Resistivity | 
Beam Energy | 
Rate Setting | 
Particle Type | 
Data | 
t-Start | 
t-Stop | 
Duration_min  | 
| 
  385  
 | 
Thu Jul 13 16:19:11 2023 | 
tb-crew | fw testing |   | RD50-MPW3 | unknown | 91 | unirradiated | 1.9K | 5.6 | Low Flux Low | electron | Unknown |   |   |   | 
| ran into noise again, about to increase thr | 
| 
  387  
 | 
Thu Jul 13 17:35:31 2023 | 
tb-crew | proper data taking |   | RD50-MPW3 | unknown | 91 | unirradiated | 1.9K | 4.2 | Low Flux Low | electron | Good |   |   |   | 
| moving telescopes back closer together as "gbl" tracking seems to have a problem with the new geometry (or we are underestimating material budget of our DUT) | 
| 
  389  
 | 
Thu Jul 13 18:26:34 2023 | 
tb-crew | proper data taking |   | RD50-MPW3 | unknown | 91 | unirradiated | 1.9K | 4.2 | Low Flux Low | electron | Good |   |   |   | 
| disconnected piggy | 
| 
  394  
 | 
Fri Jul 14 09:47:03 2023 | 
tb-crew | proper data taking |   | RD50-MPW3 | unknown | 91 | unirradiated | 1.9K | 5.6 | Low Flux Low | electron | Good |   |   |   | 
| Petra is most likely causing our efficiency drops in time, testing now if timeshift is rate dependent (see energy) | 
| 
  398  
 | 
Fri Jul 14 10:29:10 2023 | 
tb-crew | proper data taking |   | RD50-MPW3 | unknown | 91 | unirradiated | 1.9K | 4.8 | Low Flux Low | electron | Good |   |   |   | 
| reduced thr to 1.15 and deactivated rows, trying to go as low with thr as possible without noise for efficiency | 
| 
  403  
 | 
Fri Jul 14 10:48:46 2023 | 
tb-crew | proper data taking |   | RD50-MPW3 | unknown | 91 | unirradiated | 1.9K | 4.2 | Low Flux Low | electron | Good |   |   |   | 
| thr 1.08V is noise | 
| 
  413  
 | 
Fri Jul 14 13:56:09 2023 | 
tb-crew | threshold optimization |   | RD50-MPW3 | unknown | 91 | 1e14 | 1.9K | 4.2 | Low Flux Low | electron | Good |   |   |   | 
| now really going for Sams irradiated chip | 
| 
  415  
 | 
Fri Jul 14 13:57:18 2023 | 
tb-crew | irradiated |   | RD50-MPW3 | unknown | 91 | 1e14 | 1.9K | 4.2 | Low Flux Low | electron | Good |   |   |   | 
| that was noise | 
| 
  417  
 | 
Fri Jul 14 14:05:27 2023 | 
tb-crew | irradiated |   | RD50-MPW3 | unknown | 91 | 1e14 | 1.9K | 4.2 | Low Flux Low | electron | Good |   |   |   | 
| that was overkill, thr 1.3V it is | 
| 
  419  
 | 
Fri Jul 14 14:51:11 2023 | 
tb-crew | irradiated |   | RD50-MPW3 | unknown | 91 | 1e14 | 1.9K | 4.2 | Low Flux Low | electron | Good |   |   |   | 
| change geometry btw due to montage off cooling setup | 
| 
  420  
 | 
Fri Jul 14 15:50:55 2023 | 
tb-crew | irradiated |   | RD50-MPW3 | unknown | 91 | 1e14 | 1.9K | 4.2 | Low Flux Low | electron | Good |   |   |   | 
| installed cooling setup | 
| 
  423  
 | 
Fri Jul 14 15:54:43 2023 | 
tb-crew | irradiated |   | RD50-MPW3 | unknown | 91 | 1e14 | 1.9K | 4.2 | Low Flux Low | electron | Good |   |   |   | 
| thr 1.20V is noise | 
| 
  427  
 | 
Fri Jul 14 16:26:16 2023 | 
tb-crew | irradiated |   | RD50-MPW3 | unknown | 91 | 1e14 | 1.9K | 4.2 | Low Flux Low | electron | Good |   |   |   | 
| thr 1.225V,  -8.6 deg C, was running in noise, deactivated lower 30 rows | 
| 
  437  
 | 
Sat Jul 15 10:03:28 2023 | 
tb-crew | irradiated |   | RD50-MPW3 | unknown | 91 | 1e14 | 1.9K | 4.2 | Low Flux Low | electron | Good |   |   |   | 
| that was noise | 
| 
  438  
 | 
Sat Jul 15 10:04:38 2023 | 
tb-crew | irradiated |   | RD50-MPW3 | unknown | 91 | 1e14 | 1.9K | 4.2 | Low Flux Low | electron | Good |   |   |   | 
| masking all rows below row 50 | 
| 
  440  
 | 
Sat Jul 15 10:06:28 2023 | 
tb-crew | irradiated |   | RD50-MPW3 | unknown | 91 | 1e14 | 1.9K | 4.2 | Low Flux Low | electron | Good |   |   |   | 
| there is  a noisy pixel | 
| 
  442  
 | 
Sat Jul 15 10:22:32 2023 | 
tb-crew | irradiated |   | RD50-MPW3 | unknown | 105 | 1e14 | 1.9K | 4.2 | Low Flux Low | electron | Good |   |   |   | 
| changed bias, is short before breakdown now | 
| 
  445  
 | 
Sat Jul 15 10:40:09 2023 | 
tb-crew | irradiated |   | RD50-MPW3 | unknown | 105 | 1e14 | 1.9K | 4.2 | Low Flux Low | electron | Good |   |   |   | 
| pixel 57:30 is noisy, masking it | 
| 
  451  
 | 
Sat Jul 15 11:18:48 2023 | 
tb-crew | irradiated |   | RD50-MPW3 | unknown | 105 | 1e14 | 1.9K | 4.2 | Low Flux Low | electron | Good |   |   |   | 
| masked 2 more pixel | 
| 
  453  
 | 
Sat Jul 15 11:21:28 2023 | 
tb-crew | irradiated |   | RD50-MPW3 | unknown | 105 | 1e14 | 1.9K | 4.2 | Low Flux Low | electron | Good |   |   |   | 
| problems with noisy pixels now, masked 56:30  |