  6   Sat Jan 17 10:40:24 2026 elog_producerRate Test8ToF-iCT-Setup + DS LGAD + Ramon124.4 MeV/uLow Flux MediumHeliumGood17.01.2026 10:31:1217.01.2026 10:40:289NormalNone 
automatic log for run 8
Comment:
trigger window: 85ns post, 1.915 us pre
  7   Sat Jan 17 10:48:31 2026 elog_producerRate Test9ToF-iCT-Setup + DS LGAD + Ramon124.4 MeV/uLow Flux HighHeliumGood17.01.2026 10:42:2317.01.2026 10:48:356NormalNone 
automatic log for run 9
Comment:
trigger window 85 ns post, 1.915 us pre, 70mb/s events, 15mb/s drop
  8   Sat Jan 17 10:55:49 2026 elog_producerRate Test10ToF-iCT-Setup + DS LGAD + Ramon124.4 MeV/uLow Flux HighHeliumGood17.01.2026 10:51:1517.01.2026 10:55:534NormalNone 
automatic log for run 10
Comment:
trigger window 85 ns post, 640 ns pre, 52mb/s events, 4mb/s drop, 9E+6 particle rate ramon
  9   Sat Jan 17 11:00:50 2026 elog_producerRate Test11ToF-iCT-Setup + DS LGAD + Ramon124.4 MeV/uLow Flux HighHeliumGood17.01.2026 10:57:2217.01.2026 11:00:543NormalNone 
automatic log for run 11
Comment:
trigger window 85ns post, 2.915 us pre, 80mb/s events, 25mb/s drop, 9E+6 particle rate ramon
  10   Sat Jan 17 11:06:15 2026 elog_producerRate Test12ToF-iCT-Setup + DS LGAD + Ramon124.4 MeV/uLow Flux HighHeliumGood17.01.2026 11:02:0117.01.2026 11:06:194NormalNone 
automatic log for run 12
Comment:
trigger window 85ns post, 3.915 us pre, 100mb/s events, 40mb/s drop, 9E+6 particle rate ramon
  11   Sat Jan 17 11:13:44 2026 elog_producerRate Test13ToF-iCT-Setup + DS LGAD + Ramon124.4 MeV/uLow Flux MediumHeliumGood17.01.2026 11:06:5117.01.2026 11:13:486NormalNone 
automatic log for run 13
Comment:
trigger window 85ns post, 3.915 us pre, 33mb/s events, 1mb/s drop, 4E+6 particle rate ramon
  12   Sat Jan 17 11:30:05 2026 elog_producerRate Test14ToF-iCT-Setup + DS LGAD + Ramon124.4 MeV/uLow Flux MediumHeliumGood17.01.2026 11:14:4617.01.2026 11:30:0915NormalNone 
automatic log for run 14
Comment:
trigger window 85ns post, 2.915 us pre, 30mb/s events, 0.6mb/s drop, 4E+6 particle rate ramon
  13   Sat Jan 17 11:33:33 2026 elog_producerRate Test15ToF-iCT-Setup + DS LGAD + Ramon124.4 MeV/uLow Flux LowHeliumGood17.01.2026 11:31:1917.01.2026 11:33:372NormalNone 
automatic log for run 15
Comment:
trigger window 85ns post, 2.915 us pre, 25kb/s events, 0mb/s drop, 6E+3 particle rate ramon
  55   Sun Jun 7 03:44:28 2026 elog_producerRate Test5ToF-iCT-Setup + DS LGAD + Ramon402MeV/uLow Flux HighHeliumUnknown07.06.2026 03:23:4907.06.2026 03:44:4020NormalNone 
automatic log for run 5
Comment:
3 us trigger window
110 mb/s /eb
rejected: 70 mb/s
  56   Sun Jun 7 04:04:12 2026 elog_producerRate Test6ToF-iCT-Setup + DS LGAD + Ramon402MeV/uLow Flux HighHeliumUnknown07.06.2026 03:52:4807.06.2026 04:04:2411NormalNone 
automatic log for run 6
Comment:
new threshold scan

trigger window: 200 ns
threshold at 1500
  85   Sat Jul 11 19:34:51 2026 elog_producerRate Test5TOF-Cal_GSI_8LGAD+RaMon+SIPM-Setup200 MeV/uNominalMixed Beam (He+C)Unknown11.07.2026 19:01:2611.07.2026 19:34:5133NormalRW3 
automatic log for run 5
Comment:
In beam: 3 cm RW3 + PTW34070 Ionization Chamber + 11 cm RW3 + PTW34070 Ionization Chamber + 1 cm RW3

trigger on plane 4
use max rate settings (nominal with 10% He) of mixed beam
rate per strip: 60-90khz for front detector
about MHz per detector
data written: 12MB/s per eb
absolut: 50 MB/s leading
  83   Sat Jul 11 18:48:23 2026 elog_producerQA+alignment3TOF-Cal_GSI_8LGAD+RaMon+SIPM-Setup200 MeV/uNominalMixed Beam (He+C)Unknown11.07.2026 18:19:2511.07.2026 18:48:2428NormalRW3 
automatic log for run 3
Comment:
In beam: 3 cm RW3 + PTW34070 Ionization Chamber + 11 cm RW3 + PTW34070 Ionization Chamber + 1 cm RW3

Alignment mixed beam:
y-axis: aligned
x-axis: maximum at strip 18 (setup is 4-5 strips off to the right (beam is left))

SIPM
18:45: thr rescanned
  67   Sun Jun 7 11:09:39 2026 Harald HanderkasPhantom iCT runs log entries16-295ToF-iCT-Setup + DS LGAD + Ramon124.4 MeV/uLow Flux HighHeliumUnknown07.06.2026 06:49:1507.06.2026 11:06:55257NormalCylinder with inserts 
Run settings and log file for runs
  65   Sun Jun 7 06:48:30 2026 elog_producerPhantom Testrun without DAQ15ToF-iCT-Setup + DS LGAD + Ramon124.4 MeV/uLow Flux HighHeliumUnknown07.06.2026 06:43:5107.06.2026 06:48:424NormalNone 
automatic log for run 15
Comment:
  66   Sun Jun 7 11:06:43 2026 elog_producerPhantom Testrun without DAQ16ToF-iCT-Setup + DS LGAD + Ramon124.4 MeV/uLow Flux HighHeliumUnknown07.06.2026 06:49:1507.06.2026 11:06:55257NormalNone 
automatic log for run 16
Comment:
iCT tomographie runs
  101   Sun Jul 12 06:52:16 2026 elog_producerHead Phantom21TOF-Cal_GSI_8LGAD+RaMon+SIPM-Setup200 MeV/uNominalMixed Beam (He+C)Unknown12.07.2026 02:20:2512.07.2026 06:52:17271NormalHead 
automatic log for run 21
Comment:

Start automatet Spot measurements of head phantom
2:19: thr were not set correctly, no data at beginning of run 21
2:23: reset trigger window to 3us

runs 21-25 high statistics spots
ab run 81: 1mm Platte zusätzlich
  86   Sat Jul 11 20:53:37 2026 elog_producerHe Alignment + mixed beam verification6TOF-Cal_GSI_8LGAD+RaMon+SIPM-Setup402MeV/uLow Flux HighHeliumUnknown11.07.2026 19:45:1511.07.2026 20:53:3868NormalNone 
automatic log for run 6
Comment:
ohne empty bucket
now ripple filter are in
setup moved by 3mm
moved back by 6mm
He only first sensor: mean is strip 21
last sensor: mean is on strip 19

20:36: go back to mixed beam for verification
alignment looks good for mixed beam
  52   Sun Jun 7 01:18:32 2026 elog_producerGSI Sensor Test2ToF-iCT-Setup + DS LGAD + Ramon800 MeVLow Flux HighProtonUnknown07.06.2026 00:37:1507.06.2026 01:18:4441NormalNone 
automatic log for run 2
Comment:
  53   Sun Jun 7 02:41:32 2026 elog_producerGSI Sensor Test3ToF-iCT-Setup + DS LGAD + Ramon800 MeVLow Flux HighProtonUnknown07.06.2026 01:19:4707.06.2026 02:41:4381NormalNone 
automatic log for run 3
Comment:
Trigger on last detector
  80   Sat Jul 11 16:26:54 2026 elog_producerGSI Sensor Test1TOF-Cal_GSI_8LGAD+RaMon+SIPM-Setup800 MeVLow Flux HighProtonUnknown11.07.2026 16:17:2911.07.2026 16:26:549NormalRW3 
automatic log for run 1
Comment:
In beam: 3 cm RW3 + PTW34070 Ionization Chamber + 11 cm RW3 + PTW34070 Ionization Chamber + 1 cm RW3

Powersupply failed, restart in next run