Home CMS Production Clean room MedAustron HEPHY testbeams old
testbeam-MA2018 testbeam-MA2019 testbeam-MA2020 testbeam-MA2021 testbeam-MA2023 testbeam-MA2024 testbeam-MA2025
  Testbeams at MedAustron in 2025, Page 4 of 4  Not logged in ELOG logo
ID Date Author Subject Run Number DUT Beam Energy Rate Settingdown Particle Type Data t-Start t-Stop Duration_min Frontend Phantom Phantom thickness
  69   Sat Mar 15 16:50:39 2025 RD50 Generic UserSlab Test Water 1cm9Sandwich-iCT setup(GSI)+RaMon205.2 MeV/uLow Flux MediumHeliumGood16:3517:2348 Plastic Water Slabs1cm

RaMon Rate: 1.3MHz

  70   Sat Mar 15 17:30:18 2025 RD50 Generic UserSlab Test Water 2cm10Sandwich-iCT setup(GSI)+RaMon205.2 MeV/uLow Flux MediumHelium 17:3018:2454 Plastic Water Slabs2cm

RaMon Rate 1.5 MHz

  71   Sat Mar 15 18:37:36 2025 RD50 Generic UserSlab Test Water 3cm11Sandwich-iCT setup(GSI)+RaMon205.2 MeV/uLow Flux MediumHelium 18:4119:2847 Plastic Water Slabs3cm

RaMon Rate 1.5MHz

  72   Sat Mar 15 19:37:42 2025 RD50 Generic UserSlab Test Water 4cm12Sandwich-iCT setup(GSI)+RaMon205.2 MeV/uLow Flux MediumHelium 19:3920:4162 Plastic Water Slabs4cm

RaMon Rate 1.4MHz

  32   Sat Mar 15 01:29:25 2025 elog-producerFast RO Test215TJ-Monopix2800 MeVLow Flux LowProtonUnknown15.03.2025 01:26:1315.03.2025 01:29:513HVNone 
automatic log for run 215
Comment:
Lower rates then Very low settings, so just a quick run here
Fast RO for Monopix:

Delay = 2
rd_frz_delay = 2

LOAD_CONF=10
FREEZE_STOP_CONF=11
STOP_CONF=10
  36   Sat Mar 15 02:15:34 2025 elog-producerFast RO Test221TJ-Monopix2800 MeVLow Flux LowProtonUnknown15.03.2025 02:02:1515.03.2025 02:16:0013HVNone 
automatic log for run 221
Comment:
  40   Sat Mar 15 02:56:07 2025 elog-producerenergy scans225TJ-Monopix2252.7 MeVLow Flux LowProtonUnknown   HVNone 
8b10b W02R01, 1 MHz rate (low flux low)
  62   Sat Mar 15 08:58:26 2025 RD50 Generic UserAlignment Run 800MeV1Sandwich-iCT setup(GSI)+RaMon800 MeVLow Flux LowProtonGood07:4008:2040   

RaMon Rate not measured

front sensors 200mu: ~40-60Hz pro Streifen

Comments:

stopped run due to too low rate

 

  51   Sat Mar 15 04:39:13 2025 elog-producertest rate limits236TJ-Monopix2800 MeVLow Flux HighProtonUnknown15.03.2025 04:38:0415.03.2025 04:39:391HVNone 
automatic log for run 236
Comment:
Increase csinti thr to -0.2
  52   Sat Mar 15 04:42:48 2025 elog-producertest rate limits237TJ-Monopix2800 MeVLow Flux HighProtonUnknown15.03.2025 04:40:2815.03.2025 04:43:142HVNone 
automatic log for run 237
Comment:
Increase csinti thr to -0.3
  63   Sat Mar 15 09:51:23 2025 RD50 Generic UserAlignment Run 800MeV3Sandwich-iCT setup(GSI)+RaMon800 MeVLow Flux HighProtonUnknown9:4410:0319   

RaMon Rate 2MHz

front sensors 200mu: ~50-100kHz pro Streifen

Comments:

Factor 10 lost more data than measured

 

  67   Sat Mar 15 15:02:39 2025 RD50 Generic UserRate testing7Sandwich-iCT setup(GSI)+RaMon205.2 MeV/uLow Flux HighHeliumGood15:0415:2521   

RaMon Rate 4.5MHz

Comments:

From File 7 until 21 optional false, and from 22 true (everything deleted)

  73   Sat Mar 15 21:25:18 2025 RD50 Generic UserSlab Test Water 5cm14Sandwich-iCT setup(GSI)+RaMon205.2 MeV/uLow Flux HighHeliumGood21:2621:5014 Plastic Water Slabs5cm

RaMon rate: 4.5MHz

  74   Sat Mar 15 21:29:28 2025 RD50 Generic UserSlab and rate testing (with Michael)13Sandwich-iCT setup(GSI)+RaMon205.2 MeV/uLow Flux HighHeliumGood20:4421:2238 Plastic Water Slabs4cm

totzeit optimieren und deadtime finden

hex wert mal 10ns deadtime

13 ab file 10 besat effort
ab 52 wieder ohne best effort

600

  75   Wed Mar 19 11:23:53 2025 bpjust testing            
This is message text
  76   Wed Mar 19 11:27:46 2025 bpjust testing            
This is message text
ELOG V3.1.5-fc6679b