Message ID: 340     Entry time: Sat Dec 13 12:26:16 2025
Author: elog_producer 
Subject: mixed beam 
Run Number:
DUT: Sandwich-iCT setup(GSI)+RaMon 
Beam Energy: 200 MeV/u 
Rate Setting: reduced_mixed_beam 
Particle Type: Mixed Beam (He+C) 
Data: Good 
t-Start: 13.12.2025 11:47:19 
t-Stop: 13.12.2025 12:26:41 
Duration_min: 39 
Frontend: Normal 
Phantom: Mixed Beam QA phantom (ICs) 
Phantom thickness: 17cm 
automatic log for run 9
Comment:
xx% He
mixed beam QA test run
Attachment 1: ramon.conf  3 kB  | Hide | Hide all
# example config file: Ex0.conf
[RunControl]
#pCT_STOP_RUN_AFTER_N_SECONDS = 3000000
# from the base RunControl.cc
#EUDAQ_CTRL_PRODUCER_LAST_START = my_pd0
#EUDAQ_CTRL_PRODUCER_FIRST_STOP = my_pd0
# Steer which values to display in the GUI: producerName and displayed value are seperated by a ",". 
ADDITIONAL_DISPLAY_NUMBERS = "log,_SERVER"


[Producer.aida_tlu]
## GENERAL PARAMETERS
verbose= 1
skipconf= 0
confid= 20201311
delayStart= 200  

## HDMI CONFIGURATION
#  4-bits to determine direction of HDMI pins
HDMI1_set= 0x7 #1= output (driven by TLU), 0 = input. bit 0=CONT, 1=SPARE, 2=TRIG, 3=BUSY for eudet, Busy needs to be 0. (0b0111 = 0x7)
HDMI2_set= 0x7
HDMI3_set= 0x7
HDMI4_set= 0x7
#  Clock source (0= no clock, 1= Si5345, 2= FPGA)
HDMI1_clk = 1
HDMI2_clk = 1
HDMI3_clk = 1
HDMI4_clk = 1
#  Enable/Disable clock on differential LEMO
LEMOclk = 0

## PMT POWER CONFIGURATION
PMT1_V= 0.8 #0.6363 #Szinti 4x4cm
PMT2_V= 0.8 #0.6363 #Szinti 4x4cm
PMT3_V= 0.8
PMT4_V= 0.8

## TRIGGER CONFIGURATION
trigMaskHi = 0x00000000
trigMaskLo = 0x00000008 # 0x00000008 Koinzidenz 1+2 ; 0x00000040 Koinzidenz 2+3
in0_STR = 0	# 0 
in0_DEL = 0	# 18
in1_STR = 0	# 0
in1_DEL = 0	# 18
in2_STR = 0	# 0
in2_DEL = 0 	# 0
in3_STR = 0
in3_DEL = 0
in4_STR = 0
in4_DEL = 0
in5_STR = 0 #streches pulse from 6.25ns (at 0) to 12.5ns
in5_DEL = 0
# Generate internal triggers (in Hz, 0= no triggers)
InternalTriggerFreq= 0 #4000 #00

## DISCRIMINATOR THRESHOLDS
DACThreshold0 = -0.025 #-0.025 #-0.012
DACThreshold1 = -0.025 #-0.025 #-0.012
DACThreshold2 = -0.012
DACThreshold3 = -0.12
DACThreshold4 = -0.12 #-0.0066
DACThreshold5 = -0.12

## DUT CONFIGURATION
 # DUTMask Which DUTs are on
#DUTMask= 0x3 #6 bit 0= DUT1, bit 1= DUT2, bit 2= DUT3, bit 3= DUT4
#DUTMask= 0x6

 # DUTMaskMode Define AIDA (11) or EUDET (00) mode (2 bits per DUT)
DUTMaskMode= 0x30 #0x00

 # In EUDET mode: 0 = standard trigger/busy mode, 1 = raising BUSY outside handshake vetoes triggers (2 bits per DUT, 0x3= DUT 4! -backwards!)
DUTMaskModeModifier= 0x30 #0x30 #0xC0 PS: Check what this does! 0x0 würde ich hier eingeben

 # Ignore the BUSY signal for a DUT (0xF)
DUTIgnoreBusy= 0xF #ignores triggers
DUTIgnoreShutterVeto= 0x0
EnableRecordData = 1


# EnableShutterMode: 0x0. If 1, shutter mode is enabled. If 0, shutter mode is disabled.
EnableShutterMode= 0x0
# Define which input is used for shutter source [0 - 5]
ShutterSource = 5
# 32-bit counter of clocks. Set to 0 to not use internal shutter generator.
InternalShutterInterval = 0
# 32-bit counter of clocks
ShutterOnTime = 0 #start of shutter after trigger
# 32-bit counter of clocks
ShutterVetoOffTime = 0 #time after input-trigger, until first trigger is sent to output
# 32-bit counter of clocks
ShutterOffTime = 200000000 #end of shutter after trigger #200000000 = 5s (multiples of 25ns)

## DATA COLLECTOR
EUDAQ_DC= tlu_dc

## TLU MONITOR PARAMETER
FileComment="LU_MedAustron"
PlotWindowSizeinns=40e9
MinSpillTimeOnins=4.5
MinSpilldownTimeins=0.5
ThreshEvBegin=1
ThreshEvEnd=1
ThreshPartBegin=1
ThreshPartEnd=1



[LogCollector.log]
# Currently, all LogCollectors have a hardcoded runtime name: log
# nothing

[Producer.elog1]


ELOG V3.1.5-fc6679b