ID |
Date |
Author |
Subject |
Run Number |
DUT |
Beam Energy |
Rate Setting |
Particle Type |
Data |
t-Start |
t-Stop |
Duration_min |
Frontend |
Phantom |
Phantom thickness |
191
|
Sat Dec 14 15:33:35 2024 |
Felix Ulrich-Pur | plastic mouse radiography | 53 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 97.2MeV/u | Low Flux High | Helium | Good | 21:33 | 21:57 | 24 | | plastic mouse | |
run 52 was crap
pos x=175, posy=0
TRB Setup: 260V bias and 250mA pre-amps, threshold 5000
Diamond: 900mA, threshold 1000 offset |
81
|
Sat Jun 8 21:42:13 2024 |
Felix Ulrich-Pur | TOF CAL HADES LGADs WET Messung Wasser | 10 | TOF-Cal GSI | 194.3 MeV | Low Flux High | Proton | Good | 21:27 | 21:39 | | | | |
1cm plastic water (in isocentre)
6 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 1.5 MHz on LGAD
WET measurement |
190
|
Sat Dec 14 15:31:32 2024 |
Felix Ulrich-Pur | plastic mouse radiography | 51 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 97.2MeV/u | Low Flux High | Helium | Good | 21:14 | 21:30 | 16 | | plastic mouse | |
before run, mouse was aligned with lasers and shift of beam w.r.t. sensor center was taken into account
pos x=250, posy=0
TRB Setup: 260V bias and 250mA pre-amps, threshold 5000
Diamond: 900mA, threshold 1000 offset |
80
|
Sat Jun 8 21:41:04 2024 |
Felix Ulrich-Pur | TOF CAL HADES LGADs Calibrationcurve | 9 | TOF-Cal GSI | 62.4 MeV | Low Flux High | Proton | Good | 20:55 | 21:21 | | | | |
changed back to high due to beamsize vs LGAD size factor
650 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 230kHz on LGAD
TOF Calibration Curve |
189
|
Sat Dec 14 15:30:23 2024 |
Felix Ulrich-Pur | threshold and preamp tests | 50 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 97.2MeV/u | Low Flux High | Helium | Good | 20:19 | 20:35 | 16 | | | |
TRB Setup: 260V bias and 250mA pre-amps, threshold 5000 (looks good, clustersize max at ~3)
Diamond: 900mA, threshold 1000 offset
4MHz Ramon 370kHz TRB |
188
|
Sat Dec 14 15:28:46 2024 |
Felix Ulrich-Pur | threshold and preamp tests | 49 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 97.2MeV/u | Low Flux High | Helium | Good | 20:04 | 20:09 | 5 | | | |
TRB Setup: 260V bias and 250mA pre-amps, threshold 10000 (was too high)
Diamond: 900mA, threshold 1000 offset
3.5MHz Ramon |
79
|
Sat Jun 8 20:11:00 2024 |
Felix Ulrich-Pur | TOF CAL HADES LGADs Calibrationcurve | 8 | TOF-Cal GSI | 83 MeV | Low Flux Medium | Proton | Good | 19:42 | 20:51 | | | | |
300 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 35kHz on LGAD
TOF Calibration Curve |
187
|
Sat Dec 14 15:27:08 2024 |
Felix Ulrich-Pur | WET measurement slabs | 48 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 147.2 MeV/u | Low Flux High | Helium | Good | 19:30 | 19:50 | 20 | | Dense Bone Slabs | 5cm |
5cm dense bone
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000
Diamond: 900mA, threshold 1000 offset
1.5MHz Ramon
Wave crashes if 1 TDC disappears in dog d, DABC is still running |
186
|
Sat Dec 14 15:26:18 2024 |
Felix Ulrich-Pur | WET measurement slabs | 47 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 205.2 MeV/u | Low Flux High | Helium | Good | 19:16 | 19:29 | 13 | | Dense Bone Slabs | 5cm |
5cm dense bone
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000
Diamond: 900mA, threshold 1000 offset
3MHz Ramon 300kHz TRB |
78
|
Sat Jun 8 20:09:25 2024 |
Felix Ulrich-Pur | TOF CAL HADES LGADs Calibrationcurve | 7 | TOF-Cal GSI | 100.4 MeV | Low Flux Medium | Proton | Good | 19:00 | 19:42 | | | | |
300 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 65kHz on LGAD
TOF Calibration Curve |
185
|
Sat Dec 14 15:25:12 2024 |
Felix Ulrich-Pur | WET measurement slabs | 46 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 147.2 MeV/u | Low Flux High | Helium | Good | 18:55 | 19:11 | 16 | | Dense Bone Slabs | 4cm |
4cm dense bone
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000
Diamond: 900mA, threshold 1000 offset
2MHz Ramon 120kHz TRB |
184
|
Sat Dec 14 15:24:22 2024 |
Felix Ulrich-Pur | WET measurement slabs | 45 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 205.2 MeV/u | Low Flux High | Helium | Good | 18:37 | 18:51 | 14 | | Dense Bone Slabs | 4cm |
4cm dense bone
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000
Diamond: 900mA, threshold 1000 offset
3MHz Ramon 800kHz TRB |
77
|
Sat Jun 8 20:08:55 2024 |
Felix Ulrich-Pur | TOF CAL HADES LGADs Calibrationcurve | 6 | TOF-Cal GSI | 115.7MeV | Low Flux Medium | Proton | Good | 18:24 | 19:00 | | | | |
400 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 65kHz on LGAD
TOF Calibration Curve |
183
|
Sat Dec 14 15:23:21 2024 |
Felix Ulrich-Pur | WET measurement slabs | 44 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 147.2 MeV/u | Low Flux High | Helium | Good | 18:19 | 18:33 | 14 | | Dense Bone Slabs | 2cm |
3cm dense bone
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000
Diamond: 900mA, threshold 1000 offset
4MHz Ramon 200kHz TRB |
182
|
Sat Dec 14 15:21:44 2024 |
Felix Ulrich-Pur | WET measurement slabs | 43 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 205.2 MeV/u | Low Flux High | Helium | Good | 18:04 | 18:18 | 14 | | Dense Bone Slabs | 3cm |
3cm dense bone
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000
Diamond: 900mA, threshold 1000 offset
4MHz Ramon 1MHz TRB |
76
|
Sat Jun 8 20:07:44 2024 |
Felix Ulrich-Pur | TOF CAL HADES LGADs Calibrationcurve | 5 | TOF-Cal GSI | 145.4 MeV | Low Flux Medium | Proton | Good | 17:55 | 18:24 | | | | |
400 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 120kHz on LGAD
TOF Calibration Curve |
181
|
Sat Dec 14 15:20:52 2024 |
Felix Ulrich-Pur | WET measurement slabs | 42 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 147.2 MeV/u | Low Flux High | Helium | Good | 17:48 | 18:00 | 12 | | Dense Bone Slabs | 2cm |
2cm dense bone
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000
Diamond: 900mA, threshold 1000 offset
2.5MHz Ramon 250 kHz TRB |
180
|
Sat Dec 14 15:19:57 2024 |
Felix Ulrich-Pur | WET measurement slabs | 41 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 205.2 MeV/u | Low Flux High | Helium | Good | 17:32 | 17:44 | 12 | | Dense Bone Slabs | 2cm |
2cm dense bone
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000
Diamond: 900mA, threshold 1000 offset
4MHz Ramon 600 kHz TRB |
75
|
Sat Jun 8 20:05:27 2024 |
Felix Ulrich-Pur | TOF CAL HADES LGADs Calibrationcurve | 4 | TOF-Cal GSI | 175.3 MeV | Low Flux Medium | Proton | Good | 17:31 | 17:54 | | | | |
500 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 170kHz on LGAD
TOF Calibration Curve |
179
|
Sat Dec 14 15:18:48 2024 |
Felix Ulrich-Pur | WET measurement slabs | 40 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 147.2 MeV/u | Low Flux High | Helium | Good | 17:14 | 17:28 | 14 | | Dense Bone Slabs | 1cm |
1cm dense bone
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000
Diamond: 900mA, threshold 1000 offset
3.5MHz Ramon 400 kHz TRB |