Entry  Sat Dec 14 15:33:35 2024, Felix Ulrich-Pur, plastic mouse radiography, 53, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 21:33, 21:57, 24, , plastic mouse,  

run 52 was crap

pos x=175, posy=0

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000

Diamond: 900mA, threshold 1000 offset

Entry  Sat Jun 8 21:42:13 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 10, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Good, 21:27, 21:39, , , ,  

1cm plastic water (in isocentre)
6 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 1.5 MHz on LGAD

WET measurement

Entry  Sat Dec 14 15:31:32 2024, Felix Ulrich-Pur, plastic mouse radiography, 51, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 21:14, 21:30, 16, , plastic mouse,  

before run, mouse was aligned with lasers and shift of beam w.r.t. sensor center was taken into account

pos x=250, posy=0

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000

Diamond: 900mA, threshold 1000 offset

Entry  Sat Jun 8 21:41:04 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs Calibrationcurve, 9, TOF-Cal GSI, 62.4 MeV, Low Flux High, Proton, Good, 20:55, 21:21, , , ,  

changed back to high due to beamsize vs LGAD size factor
650 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 230kHz on LGAD

TOF Calibration Curve

Entry  Sat Dec 14 15:30:23 2024, Felix Ulrich-Pur, threshold and preamp tests, 50, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 20:19, 20:35, 16, , ,  

TRB Setup: 260V bias and 250mA pre-amps, threshold 5000 (looks good, clustersize max at ~3)

Diamond: 900mA, threshold 1000 offset

4MHz Ramon 370kHz TRB

Entry  Sat Dec 14 15:28:46 2024, Felix Ulrich-Pur, threshold and preamp tests, 49, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 97.2MeV/u, Low Flux High, Helium, Good, 20:04, 20:09, 5, , ,  

TRB Setup: 260V bias and 250mA pre-amps, threshold 10000 (was too high)

Diamond: 900mA, threshold 1000 offset

3.5MHz Ramon

Entry  Sat Jun 8 20:11:00 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs Calibrationcurve, 8, TOF-Cal GSI, 83 MeV, Low Flux Medium, Proton, Good, 19:42, 20:51, , , ,  

300 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 35kHz on LGAD

TOF Calibration Curve

Entry  Sat Dec 14 15:27:08 2024, Felix Ulrich-Pur, WET measurement slabs, 48, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 147.2 MeV/u, Low Flux High, Helium, Good, 19:30, 19:50, 20, , Dense Bone Slabs, 5cm 

5cm dense bone

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

1.5MHz Ramon

Wave crashes if 1 TDC disappears in dog d, DABC is still running

Entry  Sat Dec 14 15:26:18 2024, Felix Ulrich-Pur, WET measurement slabs, 47, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 205.2 MeV/u, Low Flux High, Helium, Good, 19:16, 19:29, 13, , Dense Bone Slabs, 5cm 

5cm dense bone

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

3MHz Ramon 300kHz TRB

Entry  Sat Jun 8 20:09:25 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs Calibrationcurve, 7, TOF-Cal GSI, 100.4 MeV, Low Flux Medium, Proton, Good, 19:00, 19:42, , , ,  

300 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 65kHz on LGAD

TOF Calibration Curve

Entry  Sat Dec 14 15:25:12 2024, Felix Ulrich-Pur, WET measurement slabs, 46, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 147.2 MeV/u, Low Flux High, Helium, Good, 18:55, 19:11, 16, , Dense Bone Slabs, 4cm 

4cm dense bone

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

2MHz Ramon 120kHz TRB

Entry  Sat Dec 14 15:24:22 2024, Felix Ulrich-Pur, WET measurement slabs, 45, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 205.2 MeV/u, Low Flux High, Helium, Good, 18:37, 18:51, 14, , Dense Bone Slabs, 4cm 

4cm dense bone

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

3MHz Ramon 800kHz TRB

Entry  Sat Jun 8 20:08:55 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs Calibrationcurve, 6, TOF-Cal GSI, 115.7MeV, Low Flux Medium, Proton, Good, 18:24, 19:00, , , ,  

400 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 65kHz on LGAD

TOF Calibration Curve

Entry  Sat Dec 14 15:23:21 2024, Felix Ulrich-Pur, WET measurement slabs, 44, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 147.2 MeV/u, Low Flux High, Helium, Good, 18:19, 18:33, 14, , Dense Bone Slabs, 2cm 

3cm dense bone

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

4MHz Ramon 200kHz TRB

Entry  Sat Dec 14 15:21:44 2024, Felix Ulrich-Pur, WET measurement slabs, 43, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 205.2 MeV/u, Low Flux High, Helium, Good, 18:04, 18:18, 14, , Dense Bone Slabs, 3cm 

3cm dense bone

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

4MHz Ramon 1MHz TRB

Entry  Sat Jun 8 20:07:44 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs Calibrationcurve, 5, TOF-Cal GSI, 145.4 MeV, Low Flux Medium, Proton, Good, 17:55, 18:24, , , ,  

400 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 120kHz on LGAD

TOF Calibration Curve

Entry  Sat Dec 14 15:20:52 2024, Felix Ulrich-Pur, WET measurement slabs, 42, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 147.2 MeV/u, Low Flux High, Helium, Good, 17:48, 18:00, 12, , Dense Bone Slabs, 2cm 

2cm dense bone

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

2.5MHz Ramon 250 kHz TRB

Entry  Sat Dec 14 15:19:57 2024, Felix Ulrich-Pur, WET measurement slabs, 41, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 205.2 MeV/u, Low Flux High, Helium, Good, 17:32, 17:44, 12, , Dense Bone Slabs, 2cm 

2cm dense bone

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

4MHz Ramon 600 kHz TRB

Entry  Sat Jun 8 20:05:27 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs Calibrationcurve, 4, TOF-Cal GSI, 175.3 MeV, Low Flux Medium, Proton, Good, 17:31, 17:54, , , ,  

500 khz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 170kHz on LGAD

TOF Calibration Curve

Entry  Sat Dec 14 15:18:48 2024, Felix Ulrich-Pur, WET measurement slabs, 40, TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon, 147.2 MeV/u, Low Flux High, Helium, Good, 17:14, 17:28, 14, , Dense Bone Slabs, 1cm 

1cm dense bone

TRB Setup: 260V bias and 250mA pre-amps, threshold 1000

Diamond: 900mA, threshold 1000 offset

3.5MHz Ramon 400 kHz TRB