ID |
Date |
Author |
Subject |
Run Number |
DUT |
Beam Energy |
Rate Setting |
Particle Type |
Data |
t-Start |
t-Stop |
Duration_min |
Frontend |
Phantom |
Phantom thickness |
198
|
Sat Dec 14 15:39:42 2024 |
Felix Ulrich-Pur | plastic mouse radiography | 60 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 97.2MeV/u | Low Flux High | Helium | Good | 23:50 | 00:08 | 18 | | plastic mouse | |
pos x=-100, posy=-75
TRB Setup: 260V bias and 250mA pre-amps, threshold 5000
Diamond: 900mA, threshold 1000 offset |
197
|
Sat Dec 14 15:39:11 2024 |
Felix Ulrich-Pur | plastic mouse radiography | 59 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 97.2MeV/u | Low Flux High | Helium | Good | 23:33 | 23:46 | 13 | | plastic mouse | |
pos x=-175, posy=-75
TRB Setup: 260V bias and 250mA pre-amps, threshold 5000
Diamond: 900mA, threshold 1000 offset |
196
|
Sat Dec 14 15:37:48 2024 |
Felix Ulrich-Pur | plastic mouse radiography | 58 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 97.2MeV/u | Low Flux High | Helium | Good | 23:17 | 23:30 | 13 | | plastic mouse | |
pos x=-175, posy=0
TRB Setup: 260V bias and 250mA pre-amps, threshold 5000
Diamond: 900mA, threshold 1000 offset |
195
|
Sat Dec 14 15:37:26 2024 |
Felix Ulrich-Pur | plastic mouse radiography | 57 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 97.2MeV/u | Low Flux High | Helium | Good | 23:02 | 23:16 | 14 | | plastic mouse | |
pos x=-100, posy=0
TRB Setup: 260V bias and 250mA pre-amps, threshold 5000
Diamond: 900mA, threshold 1000 offset |
194
|
Sat Dec 14 15:36:37 2024 |
Felix Ulrich-Pur | plastic mouse radiography | 56 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 97.2MeV/u | Low Flux High | Helium | Good | 22:46 | 23:00 | 14 | | plastic mouse | |
pos x=-25, posy=0
TRB Setup: 260V bias and 250mA pre-amps, threshold 5000
Diamond: 900mA, threshold 1000 offset |
193
|
Sat Dec 14 15:35:56 2024 |
Felix Ulrich-Pur | plastic mouse radiography | 55 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 97.2MeV/u | Low Flux High | Helium | Good | 22:12 | 22:45 | 33 | | plastic mouse | |
pos x=25, posy=0
TRB Setup: 260V bias and 250mA pre-amps, threshold 5000
Diamond: 900mA, threshold 1000 offset |
192
|
Sat Dec 14 15:34:50 2024 |
Felix Ulrich-Pur | plastic mouse radiography | 54 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 97.2MeV/u | Low Flux High | Helium | Good | 21:58 | 22:10 | 22 | | plastic mouse | |
pos x=100, posy=0
TRB Setup: 260V bias and 250mA pre-amps, threshold 5000
Diamond: 900mA, threshold 1000 offset |
191
|
Sat Dec 14 15:33:35 2024 |
Felix Ulrich-Pur | plastic mouse radiography | 53 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 97.2MeV/u | Low Flux High | Helium | Good | 21:33 | 21:57 | 24 | | plastic mouse | |
run 52 was crap
pos x=175, posy=0
TRB Setup: 260V bias and 250mA pre-amps, threshold 5000
Diamond: 900mA, threshold 1000 offset |
190
|
Sat Dec 14 15:31:32 2024 |
Felix Ulrich-Pur | plastic mouse radiography | 51 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 97.2MeV/u | Low Flux High | Helium | Good | 21:14 | 21:30 | 16 | | plastic mouse | |
before run, mouse was aligned with lasers and shift of beam w.r.t. sensor center was taken into account
pos x=250, posy=0
TRB Setup: 260V bias and 250mA pre-amps, threshold 5000
Diamond: 900mA, threshold 1000 offset |
189
|
Sat Dec 14 15:30:23 2024 |
Felix Ulrich-Pur | threshold and preamp tests | 50 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 97.2MeV/u | Low Flux High | Helium | Good | 20:19 | 20:35 | 16 | | | |
TRB Setup: 260V bias and 250mA pre-amps, threshold 5000 (looks good, clustersize max at ~3)
Diamond: 900mA, threshold 1000 offset
4MHz Ramon 370kHz TRB |
188
|
Sat Dec 14 15:28:46 2024 |
Felix Ulrich-Pur | threshold and preamp tests | 49 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 97.2MeV/u | Low Flux High | Helium | Good | 20:04 | 20:09 | 5 | | | |
TRB Setup: 260V bias and 250mA pre-amps, threshold 10000 (was too high)
Diamond: 900mA, threshold 1000 offset
3.5MHz Ramon |
187
|
Sat Dec 14 15:27:08 2024 |
Felix Ulrich-Pur | WET measurement slabs | 48 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 147.2 MeV/u | Low Flux High | Helium | Good | 19:30 | 19:50 | 20 | | Dense Bone Slabs | 5cm |
5cm dense bone
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000
Diamond: 900mA, threshold 1000 offset
1.5MHz Ramon
Wave crashes if 1 TDC disappears in dog d, DABC is still running |
186
|
Sat Dec 14 15:26:18 2024 |
Felix Ulrich-Pur | WET measurement slabs | 47 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 205.2 MeV/u | Low Flux High | Helium | Good | 19:16 | 19:29 | 13 | | Dense Bone Slabs | 5cm |
5cm dense bone
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000
Diamond: 900mA, threshold 1000 offset
3MHz Ramon 300kHz TRB |
185
|
Sat Dec 14 15:25:12 2024 |
Felix Ulrich-Pur | WET measurement slabs | 46 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 147.2 MeV/u | Low Flux High | Helium | Good | 18:55 | 19:11 | 16 | | Dense Bone Slabs | 4cm |
4cm dense bone
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000
Diamond: 900mA, threshold 1000 offset
2MHz Ramon 120kHz TRB |
184
|
Sat Dec 14 15:24:22 2024 |
Felix Ulrich-Pur | WET measurement slabs | 45 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 205.2 MeV/u | Low Flux High | Helium | Good | 18:37 | 18:51 | 14 | | Dense Bone Slabs | 4cm |
4cm dense bone
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000
Diamond: 900mA, threshold 1000 offset
3MHz Ramon 800kHz TRB |
183
|
Sat Dec 14 15:23:21 2024 |
Felix Ulrich-Pur | WET measurement slabs | 44 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 147.2 MeV/u | Low Flux High | Helium | Good | 18:19 | 18:33 | 14 | | Dense Bone Slabs | 2cm |
3cm dense bone
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000
Diamond: 900mA, threshold 1000 offset
4MHz Ramon 200kHz TRB |
182
|
Sat Dec 14 15:21:44 2024 |
Felix Ulrich-Pur | WET measurement slabs | 43 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 205.2 MeV/u | Low Flux High | Helium | Good | 18:04 | 18:18 | 14 | | Dense Bone Slabs | 3cm |
3cm dense bone
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000
Diamond: 900mA, threshold 1000 offset
4MHz Ramon 1MHz TRB |
181
|
Sat Dec 14 15:20:52 2024 |
Felix Ulrich-Pur | WET measurement slabs | 42 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 147.2 MeV/u | Low Flux High | Helium | Good | 17:48 | 18:00 | 12 | | Dense Bone Slabs | 2cm |
2cm dense bone
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000
Diamond: 900mA, threshold 1000 offset
2.5MHz Ramon 250 kHz TRB |
180
|
Sat Dec 14 15:19:57 2024 |
Felix Ulrich-Pur | WET measurement slabs | 41 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 205.2 MeV/u | Low Flux High | Helium | Good | 17:32 | 17:44 | 12 | | Dense Bone Slabs | 2cm |
2cm dense bone
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000
Diamond: 900mA, threshold 1000 offset
4MHz Ramon 600 kHz TRB |
179
|
Sat Dec 14 15:18:48 2024 |
Felix Ulrich-Pur | WET measurement slabs | 40 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 147.2 MeV/u | Low Flux High | Helium | Good | 17:14 | 17:28 | 14 | | Dense Bone Slabs | 1cm |
1cm dense bone
TRB Setup: 260V bias and 250mA pre-amps, threshold 1000
Diamond: 900mA, threshold 1000 offset
3.5MHz Ramon 400 kHz TRB |