Entry  Sun Aug 25 13:04:26 2024, tb-crew, only base, , RD50-MPW4, 62.4 MeV, Low Flux Very Low, Proton, Unknown, , , , Normal 
End of (promising) shift
Entry is currently edited by RD50 Generic User on 193.170.243.131  Entry  Sun Aug 25 10:02:35 2024, RD50 Generic User, init_geo_file, , , , , , , , , ,  detectors.geo
position of duts needs to be switched
Entry  Mon Aug 26 09:18:25 2024, RD50 Generic User, geo, , , , , , , , , ,  detectors_new.geo
new geo file with adjusted dut position
Entry  Sun Aug 25 08:51:02 2024, tb-crew, VNFB scan, 53, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, 25.08.2024 08:36:24, 25.08.2024 08:51:28, 15, Normal peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 53
Comment:
vnfb = 10
Entry  Sun Aug 25 08:59:46 2024, tb-crew, VNFB scan, 54, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, 25.08.2024 08:53:16, 25.08.2024 09:00:13, 6, Normal peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 54
Comment:
vnfb = 20
Entry  Sun Aug 25 09:12:07 2024, tb-crew, VNFB scan, 55, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, 25.08.2024 09:01:26, 25.08.2024 09:12:34, 11, Normal peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 55
Comment:
vnfb = 30
Entry  Sun Aug 25 09:24:15 2024, tb-crew, VNFB scan, 56, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, 25.08.2024 09:13:35, 25.08.2024 09:24:42, 11, Normal peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 56
Comment:
vnfb = 40
Entry  Sun Aug 25 09:35:54 2024, tb-crew, VNFB scan, 57, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, 25.08.2024 09:25:43, 25.08.2024 09:36:22, 10, Normal peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 57
Comment:
vnfb = 50
Entry  Sun Aug 25 09:47:51 2024, tb-crew, VNFB scan, 58, RD50-MPW4, 800 MeV, Low Flux Very Low, Proton, Unknown, 25.08.2024 09:37:13, 25.08.2024 09:48:18, 11, Normal peary_config.cfgmatrix_mpw4_base.txtmatrix_mpw4_piggy.txtmpw4.conf
automatic log for run 58
Comment:
vnfb = 60
Entry  Sun Jun 9 00:05:11 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser (wrong energy), 16, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Crap, 22:50, 22:52, ,  
wrong energy don't use
Entry  Sat Jun 8 21:42:13 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 10, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Good, 21:27, 21:39, ,  
1cm plastic water (in isocentre)
6 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 1.5 MHz on LGAD

WET measurement
Entry  Sat Jun 8 21:43:26 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 11, TOF-Cal GSI, 145.4 MeV, Low Flux High, Proton, Good, 21:39, 21:48, ,  
1cm plastic water (in isocentre)
4 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 700 kHz on LGAD

WET measurement
Entry  Sat Jun 8 21:54:59 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 12, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Good, 21:53, 22:05, ,  
2cm plastic water (in isocentre)
5 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 1.2 MHz on LGAD

WET measurement
Entry  Sat Jun 8 22:12:06 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 13, TOF-Cal GSI, 145.4 MeV, Low Flux High, Proton, Good, 22:05, 22:34, ,  
2cm plastic water (in isocentre)
3.2 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 600 kHz on LGAD

WET measurement
Entry  Sat Jun 8 22:38:41 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 14, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Good, 22:19, 22.34, ,  
3 cm plastic water (in isocentre)
4.3 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 1MHz on LGAD

WET measurement
Entry  Sun Jun 9 00:04:23 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 15, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Good, 22:34, 22:50, ,  
3 cm plastic water (in isocentre)
2.6 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 400 kHz on LGAD

WET measurement
Entry  Sun Jun 9 00:06:01 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 17, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Good, 22:53, 23:04, ,  
4 cm plastic water (in isocentre)
3.8 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 900 kHz on LGAD

WET measurement
Entry  Sun Jun 9 00:07:17 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 18, TOF-Cal GSI, 145.4 MeV, Low Flux High, Proton, Good, 23:04, 23:20, ,  
4 cm plastic water (in isocentre)
2 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 500 kHz on LGAD

WET measurement
Entry  Sun Jun 9 00:08:07 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 19, TOF-Cal GSI, 194.3 MeV, Low Flux High, Proton, Good, 23:20, 23:30, ,  
5 cm plastic water (in isocentre)
3.7 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 795 kHz on LGAD

WET measurement
Entry  Sun Jun 9 00:10:38 2024, Felix Ulrich-Pur, TOF CAL HADES LGADs WET Messung Wasser, 20, TOF-Cal GSI, 145.4 MeV, Low Flux High, Proton, Good, 23:31, 00:00, ,  
5 cm plastic water (in isocentre)
2 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 330 kHz on LGAD

14 min break with beam off (DAQ failed, had to power cycle TRB3s and all DIRICH boards-->out of synch DIRICH)