ID |
Date |
Author |
Subject |
Run Number |
DUT |
Beam Energy |
Rate Setting |
Particle Type |
Data |
t-Start |
t-Stop |
Duration_min |
Frontend |
Phantom |
Phantom thickness |
204
|
Sat Dec 14 15:42:32 2024 |
Felix Ulrich-Pur | plastic mouse radiography | 66 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 97.2MeV/u | Low Flux High | Helium | Good | 01:21 | 01:34 | 13 | | plastic mouse | |
pos x=250, posy=75
TRB Setup: 260V bias and 250mA pre-amps, threshold 5000
Diamond: 900mA, threshold 1000 offset |
96
|
Sun Jun 9 01:27:38 2024 |
Felix Ulrich-Pur | TOF CAL HADES LGADs WET Messung Knochen | 25 | TOF-Cal GSI | 194.3 MeV | Low Flux High | Proton | Good | 01:16 | 01:30 | | | | |
3 cm dense bone (in isocentre)
2.8 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 600 kHz on LGAD
WET measurement |
203
|
Sat Dec 14 15:42:01 2024 |
Felix Ulrich-Pur | plastic mouse radiography | 65 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 97.2MeV/u | Low Flux High | Helium | Good | 01:08 | 01:20 | 12 | | plastic mouse | |
pos x=250, posy=-75
TRB Setup: 260V bias and 250mA pre-amps, threshold 5000
Diamond: 900mA, threshold 1000 offset |
95
|
Sun Jun 9 01:26:56 2024 |
Felix Ulrich-Pur | TOF CAL HADES LGADs WET Messung Knochen | 24 | TOF-Cal GSI | 145.4 MeV | Low Flux High | Proton | Good | 01:02 | 01:14 | | | | |
2 cm dense bone (in isocentre)
2.1 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 315 kHz on LGAD
WET measurement |
202
|
Sat Dec 14 15:41:43 2024 |
Felix Ulrich-Pur | plastic mouse radiography | 64 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 97.2MeV/u | Low Flux High | Helium | Good | 00:54 | 01:08 | 17 | | plastic mouse | |
pos x=175, posy=-75
TRB Setup: 260V bias and 250mA pre-amps, threshold 5000
Diamond: 900mA, threshold 1000 offset |
201
|
Sat Dec 14 15:41:19 2024 |
Felix Ulrich-Pur | plastic mouse radiography | 63 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 97.2MeV/u | Low Flux High | Helium | Good | 00:37 | 00:54 | 17 | | plastic mouse | |
pos x=100, posy=-75
TRB Setup: 260V bias and 250mA pre-amps, threshold 5000
Diamond: 900mA, threshold 1000 offset |
94
|
Sun Jun 9 01:25:26 2024 |
Felix Ulrich-Pur | TOF CAL HADES LGADs WET Messung Knochen | 23 | TOF-Cal GSI | 194.3 MeV | Low Flux High | Proton | Good | 00:30 | 01:02 | | | | |
2 cm dense bone (in isocentre)
- Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 1 MHz on LGAD
interlock at 00:31 till 00:50 --> 19min break
WET measurement |
200
|
Sat Dec 14 15:40:45 2024 |
Felix Ulrich-Pur | plastic mouse radiography | 62 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 97.2MeV/u | Low Flux High | Helium | Good | 00:22 | 00:37 | 15 | | plastic mouse | |
pos x=25, posy=-75
TRB Setup: 260V bias and 250mA pre-amps, threshold 5000
Diamond: 900mA, threshold 1000 offset |
93
|
Sun Jun 9 01:22:48 2024 |
Felix Ulrich-Pur | TOF CAL HADES LGADs WET Messung Knochen | 22 | TOF-Cal GSI | 145.4 MeV | Low Flux High | Proton | Good | 00:14 | 00:23 | | | | |
1 cm dense bone (in isocentre)
2.8 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 600 kHz on LGAD
WET measurement |
199
|
Sat Dec 14 15:40:09 2024 |
Felix Ulrich-Pur | plastic mouse radiography | 61 | TOF-Cal GSI (TRB)+DOGMA setup(diamond+LGAD)+RaMon | 97.2MeV/u | Low Flux High | Helium | Good | 00:09 | 00:22 | 13 | | plastic mouse | |
pos x=-25, posy=-75
TRB Setup: 260V bias and 250mA pre-amps, threshold 5000
Diamond: 900mA, threshold 1000 offset |
92
|
Sun Jun 9 00:12:41 2024 |
Felix Ulrich-Pur | TOF CAL HADES LGADs WET Messung Knochen | 21 | TOF-Cal GSI | 194.3 MeV | Low Flux High | Proton | Good | 00:00 | 00:14 | | | | |
1 cm dense bone (in isocentre)
4.5 Mhz on Ramon (after 5 Layers of Silicon and close to second iso centre) and 1.2 MHz on LGAD
WET measurement |
1
|
Sat Mar 30 06:34:39 2024 |
Bernhard Pilsl | And a new testbeam starts :D | | | | | | | | | | | | |
|
2
|
Sat Mar 30 06:53:31 2024 |
bp, ci, hh | RD50-MPW4 | | RD50-MPW4 | 800 MeV | Low Flux Very Low | Proton | Unknown | | | | Normal | | |
about to start
geo and eudaq ini attached |
10
|
Sat Mar 30 07:59:55 2024 |
bp, ci, hh | RD50-MPW4 | | RD50-MPW4 | 800 MeV | Low Flux Very Low | Proton | Crap | | | | Normal | | |
currently no beam |
11
|
Sat Mar 30 08:05:59 2024 |
Christian Irmler | RD50-MPW4 | 999 | | | | | | | | | | | |
Tracker calibration run: mpw4-tj-monopix2_00-000999.xml
Generated global xml file for MPW4 shift: mpw4-tj-monopix2_TLU_00-000999.xml
Previous local runs and xml files (mpw4-tj-monopix2_00-000212.xml) are unusable, because sensor L5.1.1.0 had a loose cotact and thus low gain.
Fixed by disconnecting and reconnecting. |
13
|
Sat Mar 30 08:34:12 2024 |
bp, ci, hh | RD50-MPW4 | | RD50-MPW4 | 800 MeV | Low Flux Very Low | Proton | Good | | | | Normal | | |
beam should be back, there was something in the nozzle too |
14
|
Sat Mar 30 08:38:21 2024 |
bp, ci, hh | RD50-MPW4 | | RD50-MPW4 | 800 MeV | Low Flux Very Low | Proton | Good | | | | Normal | | |
beam not back, fake news |
16
|
Sat Mar 30 08:56:25 2024 |
bp, ci, hh | RD50-MPW4 | | RD50-MPW4 | 800 MeV | Low Flux Very Low | Proton | Crap | | | | Normal | | |
waiting for beam
Performing MTS restart |
18
|
Sat Mar 30 09:15:13 2024 |
bp, ci, hh | RD50-MPW4 | | RD50-MPW4 | 800 MeV | Low Flux Very Low | Proton | Crap | | | | Normal | | |
moved entire setup a bit in order too better focus beam on MPW4 (hopefully) |
19
|
Sat Mar 30 09:25:05 2024 |
bp, ci, hh | RD50-MPW4 | | RD50-MPW4 | 800 MeV | Low Flux Very Low | Proton | Crap | | | | Normal | | |
bea is back  |