ID Date Author Subject Run Number DUT Beam Energy Rate Settingup Particle Type Events Data Beam Betatron Beam Chopper Beam Degrader Beam EFE Beam Orb Beam QuadScan PhantomType PhantomAngle
  244   Sun Jun 20 20:05:38 2021 Felix Ulrich-PurHADES testrun with increased rate 800MeV17HADES-LGADs800nominalProton Good  10   none0

TB 20.6.21

Start 20:10
Mean rate O(200MHz)
 

  285   Sun Nov 7 19:00:54 2021 Felix Ulrich-PurPeakfinder WET meas 7.11.2113 175.3nominal  Good  10   none 

Start: 08:30
WET measurement of PMMA slabs also used for TOF-CT (GSI)

coarse run (to find Bragg range

  286   Sun Nov 7 19:02:25 2021 Felix Ulrich-PurPeakfinder WET meas 7.11.2114 175.3nominal  Good  10   none 

Start: 08:55
WET measurement of PMMA slabs also used for TOF-CT (GSI)

reference run (with Arduino Spill Start/Stop logic)
fringes in Bragg peak (trigger not optimal (maybe too much latency with Arduino))--> next measurement with "Accumalte Charge" instead of "Spill Start/Stop"

  287   Sun Nov 7 19:04:10 2021 Felix Ulrich-PurPeakfinder WET meas 7.11.2115 175.3nominal  Good  10   none 

Start: 09:14
WET measurement of PMMA slabs also used for TOF-CT (GSI)

reference run
"Accumalte Charge"

  288   Sun Nov 7 19:05:15 2021 Felix Ulrich-PurPeakfinder WET meas 7.11.2116 175.3nominal  Good  10   none 

Start: 09:38
WET measurement of PMMA slabs also used for TOF-CT (GSI)

1cm PMMA
"Accumalte Charge"

PMMA was activated. Had to wait ~10-15min after measurement

  289   Sun Nov 7 19:06:06 2021 Felix Ulrich-PurPeakfinder WET meas 7.11.2117 175.3nominal  Good  10   none 

Start: 10:06
WET measurement of PMMA slabs also used for TOF-CT (GSI)

2cm PMMA
"Accumalte Charge"

PMMA was activated. Had to wait ~10-15min after measurement

  290   Sun Nov 7 19:06:26 2021 Felix Ulrich-PurPeakfinder WET meas 7.11.2118 175.3nominal  Good  10   none 

Start: 10:30
WET measurement of PMMA slabs also used for TOF-CT (GSI)

3cm PMMA
"Accumalte Charge"

PMMA was activated. Had to wait ~10-15min after measurement

  291   Sun Nov 7 19:06:40 2021 Felix Ulrich-PurPeakfinder WET meas 7.11.2119 175.3nominal  Good  10   none 

Start: 11:00
WET measurement of PMMA slabs also used for TOF-CT (GSI)

4cm PMMA
"Accumalte Charge"

PMMA was activated. Had to wait ~10-15min after measurement

  292   Sun Nov 7 19:06:58 2021 Felix Ulrich-PurPeakfinder WET meas 7.11.2120 175.3nominal  Good  10   none 

Start: 11:24
WET measurement of PMMA slabs also used for TOF-CT (GSI)

5cm PMMA
"Accumalte Charge"

PMMA was activated. Had to wait ~10-15min after measurement

  293   Sun Nov 7 19:07:31 2021 Felix Ulrich-PurPeakfinder WET meas 7.11.2121 194.3nominal  Good  10   none 

Start: 11:44
WET measurement of PMMA slabs also used for TOF-CT (GSI)

5cm PMMA. higher energy for Felix Feichtinger
"Accumalte Charge"

coarse run

  294   Sun Nov 7 19:08:47 2021 Felix Ulrich-PurPeakfinder WET meas 7.11.2122 194.3nominal  Good  10   none 

Start: 12:00
WET measurement of PMMA slabs also used for TOF-CT (GSI)

5cm PMMA. higher energy for Felix Feichtinger
"Accumalte Charge"

more detailed measurement