Home CMS Production Clean room MedAustron HEPHY testbeams old
testbeam-MA2018 testbeam-MA2019 testbeam-MA2020 testbeam-MA2021 testbeam-MA2023 testbeam-MA2024
  MedAustron 2021, Page 9 of 18  Not logged in ELOG logo
ID Date Author Subject Run Number DUT Beam Energy Rate Setting Particle Typedown Events Data Beam Betatron Beam Chopper Beam Degrader Beam EFE Beam Orb Beam QuadScan PhantomType PhantomAngle
  256   Thu Oct 7 11:01:19 2021 Philipp GagglPad Sensoren21SiC+Si62.4I(EFE+SID0)Proton Unknown        

beamtime on 18.09.2021
CH1: SiC
CH3: Si
Bias_Si: -100 V
Bias_SiC: 800 V

  257   Thu Oct 7 11:01:53 2021 Philipp GagglPad Sensoren22SiC+Si62.4I(EFE+SID0)Proton Unknown        

beamtime on 18.09.2021
CH1: SiC
CH3: Si
Bias_Si: -100 V
Bias_SiC: 600 V

  258   Thu Oct 7 11:02:15 2021 Philipp GagglPad Sensoren23SiC+Si100.4I(EFE+SID0)Proton Unknown        

beamtime on 18.09.2021
CH1: SiC
CH3: Si
Bias_Si: -100 V
Bias_SiC: 600 V

  259   Thu Oct 7 11:02:43 2021 Philipp GagglPad Sensoren24SiC+Si100.4I(EFE+SID0)Proton Unknown        

beamtime on 18.09.2021
CH1: SiC
CH3: Si
Bias_Si: -100 V
Bias_SiC: 800 V

  260   Thu Oct 7 11:03:04 2021 Philipp GagglPad Sensoren25SiC+Si100.4I(EFE+SID0)Proton Unknown        

beamtime on 18.09.2021
CH1: SiC
CH3: Si
Bias_Si: -100 V
Bias_SiC: 1100 V

  268   Thu Oct 7 11:12:10 2021 Philipp GagglPad Sensoren34SiC+Si62.4 Proton Unknown  10     

beamtime on 18.09.2021
CH1: SiC
CH3: Si
Bias_Si: -100 V
Bias_SiC: 400 V

  269   Thu Oct 7 11:12:44 2021 Philipp GagglPad Sensoren35SiC+Si62.4 Proton Unknown  10     

beamtime on 18.09.2021
CH1: SiC
CH3: Si
Bias_Si: -100 V
Bias_SiC: 600 V

  270   Thu Oct 7 11:13:05 2021 Philipp GagglPad Sensoren36SiC+Si62.4 Proton Unknown  10     

beamtime on 18.09.2021
CH1: SiC
CH3: Si
Bias_Si: -100 V
Bias_SiC: 800 V

  271   Thu Oct 7 11:13:21 2021 Philipp GagglPad Sensoren37SiC+Si62.4 Proton Unknown  10     

beamtime on 18.09.2021
CH1: SiC
CH3: Si
Bias_Si: -100 V
Bias_SiC: 1100 V

  295   Sun Nov 7 19:09:50 2021 Felix Ulrich-PurFluence loss meas 7.11.2124 252.7II(BETA+EFE)Proton Good      none 

Start: 13:58

Alignment run without phantom

  296   Sun Nov 7 19:10:31 2021 Felix Ulrich-PurFluence loss meas 7.11.2125 145.4II(BETA+EFE)Proton40e6Good      none 

Start: 14:04

Reference run without phantom

  297   Sun Nov 7 19:11:24 2021 Felix Ulrich-PurFluence loss meas 7.11.2126Drehtisch145.4II(BETA+EFE)Proton60e6Good      Alu0

Start: 15:19

1 cm Alu stair phantom on rotary table (0deg--> always 1cm Al in beam)

  298   Sun Nov 7 19:12:57 2021 Felix Ulrich-PurFluence loss meas 7.11.2127Drehtisch145.4II(BETA+EFE)Proton60e6Good      Alu90

Start: 16:40

1 cm Alu stair phantom on rotary table (90deg--> 0.2-1cm Alu)

  299   Sun Nov 7 19:13:36 2021 Felix Ulrich-PurFluence loss meas 7.11.2128Drehtisch100.4II(BETA+EFE)Proton60e6Good      Alu90

Start: 18:00

1 cm Alu stair phantom on rotary table (90deg)

  300   Sun Nov 7 19:28:49 2021 Felix Ulrich-PurFluence loss meas 7.11.2129Drehtisch100.4II(BETA+EFE)Proton60e6Good      Alu0

Start: 19:28

1 cm Alu stair phantom on rotary table (0deg)

  301   Sun Nov 7 21:04:43 2021 Felix Ulrich-PurFluence loss meas 7.11.2130 100.4II(BETA+EFE)Proton40e6Good      none 

Start: 21.04

reference run 100.4MeV

  302   Sun Nov 21 13:51:53 2021 Simon WaidComparison SiC, Si, Si LGAD and Diamond1SiC+Si62.4I(EFE+SID0)Proton10000         

Bias voltage SiC: 800

Bias voltage Si: -150

From a beam perspective the Si detector was the first target, on oscilloscope  channel 1, SiC was the second target connected to oscilloscope channel 3.

Triggering source was the SiC detector.

Data was stored on the HiBPM-Osci-2 NUC under c:\measurements

  303   Sun Nov 21 13:55:05 2021 Simon WaidComparison SiC, Si, Si LGAD and Diamond2SiC+Si62.4I(EFE+SID0)Proton1000000         

Bias voltage SiC: 800

Bias voltage Si: -150

From a beam perspective the Si detector was the first target, on oscilloscope  channel 1, SiC was the second target connected to oscilloscope channel 3.

Triggering source was the Si detector.

The compliancy of the SiC SMU triggered after approx. 50% of the events had been acquired after this event the HV for the SiC HV was shut down.

Data was stored on the HiBPM-Osci-2 NUC under c:\measurements

  304   Sun Nov 21 14:05:18 2021 Simon WaidComparison SiC, Si, Si LGAD and Diamond3 62.4I(EFE+SID0)Proton10000         

Bias voltage SiC: 800

Bias voltage Si LGAD: -100

From a beam perspective the SiC detector was the first target, on oscilloscope channel 4, the Si LGAD was the second target connected to oscilloscope channel 1.

Triggering source was the Si LGAD detector. The SiC detector was most probably defective.

Data was stored on the HiBPM-Osci-2 NUC under c:\measurements

  305   Sun Nov 21 14:53:00 2021 Simon WaidComparison SiC, Si, Si LGAD and Diamond4Diamond+LGAD62.4I(EFE+SID0)Proton100000         

Bias voltage Diamond: 500

Bias voltage Si LGAD: -100

From a beam perspective the Diamond detector was the first target, on oscilloscope channel 4, the Si LGAD was the second target connected to oscilloscope channel 1.

Triggering source was the Si LGAD detector.

Data was stored on the HiBPM-Osci-2 NUC under c:\measurements

ELOG V3.1.5-fc6679b