  271   Thu Oct 7 11:13:21 2021 Philipp GagglPad Sensoren37SiC+Si62.4 Proton Unknown  10     

beamtime on 18.09.2021
CH1: SiC
CH3: Si
Bias_Si: -100 V
Bias_SiC: 1100 V

  272   Thu Oct 7 11:13:46 2021 Philipp GagglPad Sensoren38SiC+Si    Unknown        

beamtime on 18.09.2021
CH1: SiC
CH3: Si
Bias_Si: -100 V
Bias_SiC: 1100 V
Noise measurement

  222   Wed May 12 08:53:15 2021 Felix Ulrich-PurPMMA 90deg10Drehtisch145.4II(BETA+EFE)Proton Good      PMMA90

TB 24.4.21

90deg (0.2-1cm PMMA)

  223   Wed May 12 08:53:46 2021 Felix Ulrich-PurPMMA 90deg11Drehtisch100.4II(BETA+EFE)Proton Good      PMMA90

TB 24.4.21

90deg (0.2-1cm PMMA)

  221   Wed May 12 08:52:05 2021 Felix Ulrich-PurPMMA 0deg9Drehtisch145.4II(BETA+EFE)Proton Good      PMMA0

TB 24.4.21

0deg (1cm PMMA)

timing was adjusted and FADC had to be restarted

  224   Wed May 12 08:54:54 2021 Felix Ulrich-PurPMMA 0deg12Drehtisch100.4II(BETA+EFE)Proton Good      PMMA0

TB 24.4.21

0deg (1cm PMMA)

  117   Tue Mar 2 14:42:56 2021 Patrick SiebererMPW2 Tracking75MPW2+tracker252.7I(EFE+SID0)Proton Crap        

Start xx:xx:

Tracker speed test: 2MHz internal Trigger from TLU -> works.

  118   Tue Mar 2 14:45:04 2021 Patrick SiebererMPW2 Tracking76MPW2+tracker252.7I(EFE+SID0)Proton Crap        

Start xx:xx:

Tracker speed test: 4MHz internal Trigger from TLU -> works.
Tracker Setting: max_trigger=1 and trigger_busy=25. This gives ~20kHz eventrate.

  119   Tue Mar 2 14:46:49 2021 Patrick SiebererMPW2 Tracking77-82MPW2+tracker252.7I(EFE+SID0)Proton Crap        

Start 22:23 (Run79), 22:30, 22:37, 22:43 

Synchronization: Playing around with trgger_busy (range 25-500) This slows down Eventrate, but does not solve our issue of not being able to count all overflows. (This depends on the ratio of events hitting the MPW2 only, NOT on the overall busy rate)

  120   Tue Mar 2 15:30:11 2021 Patrick SiebererMPW2 Tracking83MPW2+tracker252.7I(EFE+SID0)Proton100kCrap        

Start 22:55
Threshold MPW2: 960mV, Pixel R1 C1
MaxTrigger: 1, Trigger-busy: 25
3 Koinzidenz TLU und 3er Koinzidenz Messung am MPW

  121   Tue Mar 2 15:32:52 2021 Patrick SiebererMPW2 Tracking84MPW2+tracker252.7I(EFE+SID0)Proton30kCrap        

Start 23:21
Threshold MPW2: 960mV, Pixel R1 C1
MaxTrigger: 1, Trigger-busy: 500
2 Koinzidenz TLU und 2er Koinzidenz Messmodus am MPW
Conclusion: Slowing down Tracker does not help (seen before)

  122   Tue Mar 2 15:34:49 2021 Patrick SiebererMPW2 Tracking85MPW2+tracker252.7I(EFE+SID0)Proton640kUnknown        

Start 23:24
Threshold MPW2: 960mV, Pixel R1 C1
MaxTrigger: 1, Trigger-busy: 500
2 Koinzidenz TLU und 3er Koinzidenz Messmodus am MPW

  123   Tue Mar 2 15:39:24 2021 Patrick SiebererMPW2 Tracking86MPW2+tracker252.7I(EFE+SID0)Proton46kUnknown        

Start 23:43
Threshold MPW2: 960mV, Pixel R1 C1
MaxTrigger: 1, Trigger-busy: 500
2 Koinzidenz TLU und 3er Koinzidenz Messmodus am MPW

  124   Tue Mar 2 15:40:01 2021 Patrick SiebererMPW2 Tracking87MPW2+tracker252.7I(EFE+SID0)Proton Good        

Start 23:44
Threshold MPW2: 960mV, Pixel R1 C1
MaxTrigger: 1, Trigger-busy: 500
3 Koinzidenz TLU und 3er Koinzidenz Messmodus am MPW

testing for picoscope

  125   Tue Mar 2 15:41:09 2021 Patrick SiebererMPW2 Tracking88MPW2+tracker252.7I(EFE+SID0)Proton Good        

Start 23:54
Threshold MPW2: 1200mV, Pixel R1 C1
MaxTrigger: 1, Trigger-busy: 500
3 Koinzidenz TLU und 3er Koinzidenz Messmodus am MPW

testing for picoscope

  126   Tue Mar 2 15:42:11 2021 Patrick SiebererMPW2 Tracking89MPW2+tracker252.7I(EFE+SID0)Proton1.8MUnknown        

Start 23:56
Threshold MPW2: 1200mV, Pixel R1 C1
MaxTrigger: 1, Trigger-busy: 500
2 Koinzidenz TLU und 3er Koinzidenz Messmodus am MPW

  127   Tue Mar 2 15:43:01 2021 Patrick SiebererMPW2 Tracking90MPW2+tracker252.7I(EFE+SID0)Proton300kCrap        

Start 00:34
Threshold MPW2: 960mV, Pixel R1 C1
MaxTrigger: 1, Trigger-busy: 500
2 Koinzidenz TLU und 3er Koinzidenz Messmodus am MPW
Testen neue Software: Triggernummer von 2er Koinzidenz modus, Daten von 3er Koinzidenz modus
Forgot to restart Producer

  128   Tue Mar 2 15:44:51 2021 Patrick SiebererMPW2 Tracking91MPW2+tracker252.7I(EFE+SID0)Proton235kGood        

Start 00:41
Threshold MPW2: 960mV, Pixel R1 C1
MaxTrigger: 1, Trigger-busy: 500
2 Koinzidenz TLU und Hybrid Modus (2er und 3er Koinzidenz) am MPW
Producer restartet -> works

  129   Tue Mar 2 15:46:21 2021 Patrick SiebererMPW2 Tracking92MPW2+tracker252.7I(EFE+SID0)Proton235kGood        

Start 00:xx
Threshold MPW2: 960mV, Pixel R1 C1
MaxTrigger: 1, Trigger-busy: 500
2 Koinzidenz TLU und Hybrid Modus (2er und 3er Koinzidenz) am MPW
Fill Level der MPW Fifos von 5 auf 1 geaendert

  130   Tue Mar 2 15:47:32 2021 Patrick SiebererMPW2 Tracking93MPW2+tracker252.7I(EFE+SID0)Proton Crap        

Start 00:xx
Threshold MPW2: 960mV, Pixel R1 C1
MaxTrigger: 1, Trigger-busy: 500
2 Koinzidenz TLU und Hybrid Modus (2er und 3er Koinzidenz) am MPW
Davor: Neue local runs am Tracker (...TLU_92.xml) -> Ab hier verwendet
Run for testing beam only