Home CMS Production Clean room MedAustron HEPHY testbeams old
testbeam-MA2018 testbeam-MA2019 testbeam-MA2020 testbeam-MA2021 testbeam-MA2023 testbeam-MA2024
  MedAustron 2021, Page 6 of 18  Not logged in ELOG logo
New entries since:Thu Jan 1 01:00:00 1970
ID Date Author Subjectdown Run Number DUT Beam Energy Rate Setting Particle Type Events Data Beam Betatron Beam Chopper Beam Degrader Beam EFE Beam Orb Beam QuadScan PhantomType PhantomAngle
  159   Tue Mar 16 15:04:21 2021 Patrick SiebererPad Sensoren14 62.4II(BETA+EFE)Proton Good      none 

Start: 11:14
Range Shifter Out, Filter 1 IN, Filter 2 IN

  160   Tue Mar 16 15:04:52 2021 Patrick SiebererPad Sensoren15 145.4II(BETA+EFE)Proton Good      none 

Start: 11:26
Range Shifter Out, Filter 1 IN, Filter 2 IN

  161   Tue Mar 16 15:05:08 2021 Patrick SiebererPad Sensoren16 100.4II(BETA+EFE)Proton Good      none 

Start: 11:38
Range Shifter Out, Filter 1 IN, Filter 2 IN

  162   Tue Mar 16 15:05:24 2021 Patrick SiebererPad Sensoren17 194.3II(BETA+EFE)Proton Good      none 

Start: 11:46
Range Shifter Out, Filter 1 IN, Filter 2 IN

  163   Tue Mar 16 15:05:40 2021 Patrick SiebererPad Sensoren18 252.7II(BETA+EFE)Proton Good      none 

Start: 11:57
Range Shifter Out, Filter 1 IN, Filter 2 IN
Lost beam at 12:15

  164   Tue Mar 16 15:06:13 2021 Patrick SiebererPad Sensoren19 252.7II(BETA+EFE)Proton Good      none 

Start: 12:18
Range Shifter Out, Filter 1 IN, Filter 2 IN
Restart run with same settings

  165   Tue Mar 16 15:07:06 2021 Patrick SiebererPad Sensoren20 100.4I(EFE+SID0)Proton Good      none 

Start: 12:25
Range Shifter Out, Filter 1 IN, Filter 2 IN
Changed compliance of both sensors to 100 uA

  166   Tue Mar 16 15:07:56 2021 Patrick SiebererPad Sensoren21 100.4III(EFE+SID2)Proton Good      none 

Start: 12:41
Range Shifter Out, Filter 1 IN, Filter 2 IN
12:53 went inside to test if sensors are activated

  167   Tue Mar 16 15:09:04 2021 Patrick SiebererPad Sensoren22 800VProton Good      none 

Start: 13:09
Range Shifter Out, Filter 1 IN, Filter 2 IN
Changed compliance values back to 30 uA

  168   Tue Mar 16 15:09:33 2021 Patrick SiebererPad Sensoren23 800VIIProton Good      none 

Start: 13:36
Range Shifter Out, Filter 1 IN, Filter 2 IN

  169   Tue Mar 16 15:10:11 2021 Patrick SiebererPad Sensoren24 252.7nominalProton Good  10   none 

Start: 13:57
Range Shifter Out, Filter 1 IN, Filter 2 IN

  170   Tue Mar 16 15:10:41 2021 Patrick SiebererPad Sensoren25 252.7nominalProton Good  50   none 

Start: 14:06
Range Shifter Out, Filter 1 IN, Filter 2 IN

  171   Tue Mar 16 15:11:00 2021 Patrick SiebererPad Sensoren26 252.7nominalProton Good  100   none 

Start: 14:20
Range Shifter Out, Filter 1 IN, Filter 2 IN

  172   Tue Mar 16 15:11:34 2021 Patrick SiebererPad Sensoren27 252.7nominalProton Good  100   none 

Start: 14:28
Range Shifter Out, Filter 1 IN, Filter 2 IN
Power supply of si-sensor went into compliance, therefore the compliance value was set to 100 uA again
Ende 14:35

  245   Sat Sep 18 06:20:51 2021 Manuel ChristanellPad Sensoren10 62.4III(EFE+SID2)Proton          

01. SiC; 2. Si

Bias 1: 400V; Bias 2: -100 V; SiC with Cividec

  246   Sat Sep 18 07:22:07 2021 Manuel ChristanellPad Sensoren11SiC+Si62.4III(EFE+SID2)Proton          

01. SiC; 2. Si

Small SiC Signal;
Cividec for SiC

  247   Thu Oct 7 10:51:38 2021 Philipp GagglPad Sensoren12SiC+Si62.4III(EFE+SID2)Proton Good        

beamtime on 18.09.2021
CH1: SiC
CH3: Si
Bias_Si: -100 V
Bias_SiC: 400 V
Cividec for SiC

  248   Thu Oct 7 10:54:33 2021 Philipp GagglPad Sensoren13SiC+Si62.4II(BETA+EFE)Proton Unknown        

beamtime on 18.09.2021
CH1: SiC
CH3: Si
Bias_Si: -100 V
Bias_SiC: 400 V

  249   Thu Oct 7 10:56:59 2021 Philipp GagglPad Sensoren14SiC+Si62.4II(BETA+EFE)Proton Unknown        

beamtime on 18.09.2021
CH1: SiC
CH3: Si
Bias_Si: -100 V
Bias_SiC: 400 V
Repeat of run 13

  250   Thu Oct 7 10:57:31 2021 Philipp GagglPad Sensoren15SiC+Si62.4II(BETA+EFE)Proton Unknown        

beamtime on 18.09.2021
CH1: SiC
CH3: Si
Bias_Si: -100 V
Bias_SiC: 800 V
Trigger SiC

ELOG V3.1.5-fc6679b