Home CMS Production Clean room MedAustron HEPHY testbeams old
testbeam-MA2018 testbeam-MA2019 testbeam-MA2020 testbeam-MA2021 testbeam-MA2023 testbeam-MA2024
  MedAustron 2021, Page 3 of 18  Not logged in ELOG logo
Entry  Sun Jan 10 07:19:17 2021, Maximilian Tomaschek, Lowflux recommissioning, 28, Ramon+Tracker, 252.7, , , 10 spills 1e6 events, Good, 10%, , 10, 1, , , ,  

Spotsize 

Entry  Sun Jan 10 07:20:36 2021, Maximilian Tomaschek, Lowflux recommissioning, 30, Ramon+Tracker, 194.3, , , 10 spills 1e6 events, Good, 10%, , 10, 1, , , ,  

Spotsize ,

Entry  Sun Jan 10 07:22:59 2021, Maximilian Tomaschek, Lowflux recommissioning, 31, Ramon+Tracker, 175.3, , , 10 spills 1e6 events, Good, 10%, , 10, 1, , , ,  

Spotsize

Entry  Sun Jan 10 07:23:25 2021, Maximilian Tomaschek, Lowflux recommissioning, 32, Ramon+Tracker, 145.4, , , 10 spills 1e6 events, Good, 10%, , , 10, , , ,  

Spotsize

Entry  Sun Jan 10 07:23:39 2021, Maximilian Tomaschek, Lowflux recommissioning, 33, Ramon+Tracker, 100.4, , , 10 spills 1e6 events, Unknown, 10%, , 10, 1, , , ,  

Spotsize, trigger stopped

Entry  Sun Jan 10 07:24:04 2021, Maximilian Tomaschek, Lowflux recommissioning, 34, Ramon+Tracker, 83, , , 10 spills 1e6 events, Good, 10%, , 10, 1, , , ,  

Spotsize

Entry  Sun Jan 10 07:24:28 2021, Maximilian Tomaschek, Lowflux recommissioning, 35, Ramon+Tracker, 62.4, , , 10 spills 1e6 events, Good, 10%, , 10, 1, , , ,  

Spotsize

Entry  Wed Jan 20 17:12:15 2021, Felix Ulrich-Pur, Lowflux recommissioning, 4, Ramon+Tracker, 145.4, , , 30 spills, , 10%, , 10, 1, , , ,  

Rate

Entry  Wed Jan 20 17:25:34 2021, Felix Ulrich-Pur, Lowflux p800 commissioning, 56, Ramon+Tracker, 800, , , 30 spills, , 10%, 60%, 20, 10, 6dB, , ,  

MT=1, BSY=50,
1cm off in +x and +y direction
Setting IV

Entry  Wed Jan 20 17:27:58 2021, Felix Ulrich-Pur, Lowflux p800 commissioning, 57, Ramon+Tracker, 800, , , 30 spills, , 10%, 60%, 20, 10, 6dB, , ,  

MT=1, BSY=50,
2cm off in +x and +y direction (wrong direction)
setting IV
 

Entry  Wed Jan 20 17:28:56 2021, Felix Ulrich-Pur, Lowflux p800 commissioning, 58, Ramon+Tracker, 800, , , 30 spills, , 10%, 60%, 20, 10, 6dB, , ,  

MT=1, BSY=50,
looks +0.5vcm x, -0.5cm y
rate quite low
setting IV
 

Entry  Wed Jan 20 17:29:42 2021, Felix Ulrich-Pur, Lowflux p800 commissioning, 59, Ramon+Tracker, 800, , , 30 spills, , 10%, 60%, 20, 10, 6dB, , ,  

MT=1, BSY=50,
x is centered now, y looks 0.25cm off
setting IV
 

Entry  Wed Jan 20 17:30:17 2021, Felix Ulrich-Pur, Lowflux p800 commissioning, 60, Ramon+Tracker, 800, , , 30 spills, , 10%, 60%, 20, 10, 6dB, , ,  

MT=1, BSY=50,
rate few 100Hz
x looks off, y centered now (after -0.25cm in y correction)
setting IV
 

Entry  Wed Jan 20 17:31:04 2021, Felix Ulrich-Pur, Lowflux p800 commissioning, 61, Ramon+Tracker, 800, , , 30 spills, Crap, 10%, 60%, 10, 10, 6dB, , ,  

wrong degrader setting
setting IV
 

Entry  Wed Jan 20 17:31:30 2021, Felix Ulrich-Pur, Lowflux p800 commissioning, 63, Ramon+Tracker, 800, , , 30 spills, , 10%, 60%, 20, 10, 6dB, , ,  

MT=1, BSY=50,
beam almost centered (total correction +15mm x (to the left in beam direction) and +7.5 mm y (down in beam direction)

rate between 100Hz and 70 kHz
setting IV
 

Entry  Wed Jan 20 17:33:58 2021, Felix Ulrich-Pur, Lowflux p800 commissioning, 64, Ramon+Tracker, 800, , , 30 spills, , 10%, 60%, 20, 10, 6dB, , ,  

MT=1, BSY=50,
beam  centered (total correction +12.5mm x (to the left in beam direction) and +7.5 mm y (down in beam direction)

rate between 100Hz and 70 kHz, few 100Hz spills
setting IV
 

Entry  Wed Jan 20 17:35:25 2021, Felix Ulrich-Pur, Lowflux p800 commissioning, 62, Ramon+Tracker, 800, , , 30 spills, , 10%, 60%, 20, 10, 6dB, , ,  

MT=1, BSY=50,
now correct degrader, beam still off
setting IV
 

Entry  Wed Jan 20 17:35:51 2021, Felix Ulrich-Pur, Lowflux p800 commissioning, 66, Ramon+Tracker, 800, , , 30 spills, , 10%, 62%, 20, 10, 6dB, , ,  

MT=1, BSY=50,
beam is centered
rate few 100kHz and two peaks in beam profile
setting IV
 

Entry  Wed Jan 20 17:36:21 2021, Felix Ulrich-Pur, Lowflux p800 commissioning, 65, Ramon+Tracker, 800, , , 30 spills, , 10%, 80%, 20, 10, 6dB, , ,  

MT=1, BSY=50,
beam shifted again
jumps in rate too high
setting IV
 

Entry  Wed Jan 20 17:41:27 2021, Felix Ulrich-Pur, Lowflux p800 commissioning, 67, Ramon+Tracker, 800, , , 30 spills, , 10%, 60.5%, 20, 10, 6dB, , ,  

MT=1, BSY=50,
beam shifted again
max 30kHz, low rate
setting IV
 

ELOG V3.1.5-fc6679b