Home CMS Production Clean room MedAustron HEPHY testbeams old
testbeam-MA2018 testbeam-MA2019 testbeam-MA2020 testbeam-MA2021 testbeam-MA2023 testbeam-MA2024
  MedAustron 2021, Page 2 of 18  Not logged in ELOG logo
ID Date Author Subject Run Number DUT Beam Energy Rate Setting Particle Type Events Data Beam Betatron Beam Chopper Beam Degraderdown Beam EFE Beam Orb Beam QuadScan PhantomType PhantomAngle
  68   Wed Jan 20 17:51:28 2021 Felix Ulrich-PurLowflux p800 commissioning77Ramon+Tracker800  30 spills 100%60.5%20306dB   

MT=1, BSY=50,

beam off
potential setting VII
steering setting IV + 1cm in +y direction

  69   Wed Jan 20 17:51:50 2021 Felix Ulrich-PurLowflux p800 commissioning79Ramon+Tracker800  30 spills 100%60.5%20306dB   

MT=1, BSY=50

beam off
potential setting VII
steering setting IV + 1cm in +y direction

  70   Wed Jan 20 17:52:12 2021 Felix Ulrich-PurLowflux p800 commissioning80Ramon+Tracker800  30 spills 100%60.5%20306dB   

MT=1, BSY=50

beam off
potential setting VII
steering setting IV + shift in -y direction

  71   Wed Jan 20 17:52:28 2021 Felix Ulrich-PurLowflux p800 commissioning81Ramon+Tracker800  30 spills 100%60.5%20306dB   

MT=1, BSY=50,

beam off
potential setting VII
steering setting IV + bigger shift in -y direction

  72   Wed Jan 20 17:52:49 2021 Felix Ulrich-PurLowflux p800 commissioning82Ramon+Tracker800  30 spills 100%60.5%20306dB   

MT=1, BSY=50,
potential setting VII
new steering +12.5mm x,+10mm y ( -beam direction), still off

  73   Wed Jan 20 17:55:13 2021 Felix Ulrich-PurLowflux p800 commissioning83Ramon+Tracker800  30 spills 100%60.5%20306dB   

MT=1, BSY=50,
final setting VII
new steering VII +12.5mm x,+13mm y ( -beam direction)

  74   Wed Jan 20 17:56:06 2021 Felix Ulrich-PurLowflux p800 commissioning84Ramon+Tracker800  30 spills 12%60.5%20306dB   

MT=1, BSY=50,
setting VI steering VII,beam is off again

  75   Wed Jan 20 17:57:16 2021 Felix Ulrich-PurLowflux p800 commissioning85Ramon+Tracker800  30 spillsUnknown100%60.5%20106dB   

MT=1, BSY=50,
setting V steering VII,beam looks centered, but rate and spot looks similar to setting VII, maybe wrong acc setting? --> don't use

  235   Sun Jun 20 19:38:28 2021 Felix Ulrich-PurHADES increased rate run (wrong setting9HADES-LGADs800VIIProton Good  20   none0

TB 19.6.21

Start 19:43
rate setting was not set correctly deg20 in combination with setting VII instead of nominal flux deg10
 

  93   Sat Feb 27 18:50:16 2021 Felix Ulrich-PurWaveforms SiC+Si19SiC+Si+RaMon252.7nominal  Good  100     

Start 13:40

Degrader 100 
Messung aufgenommen, keine Screenshots

channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.100 V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x = 8
y = 68

  96   Sat Feb 27 18:56:58 2021 Felix Ulrich-PurWaveforms SiC+Si22SiC+Si+RaMon252.7nominalProton Good  100     

Start 14:08

Degrader 100


channel 1 : Si upstream
channel 2: SiC downstream

400 V Bias

Referenz: 2.100 V channel 2, 0-Level: 2.310V channel 2
Referenz: 2.477V channel 1, 0-Level: 2.315V channel 1

position:

x = 8
y = 68

  171   Tue Mar 16 15:11:00 2021 Patrick SiebererPad Sensoren26 252.7nominalProton Good  100   none 

Start: 14:20
Range Shifter Out, Filter 1 IN, Filter 2 IN

  172   Tue Mar 16 15:11:34 2021 Patrick SiebererPad Sensoren27 252.7nominalProton Good  100   none 

Start: 14:28
Range Shifter Out, Filter 1 IN, Filter 2 IN
Power supply of si-sensor went into compliance, therefore the compliance value was set to 100 uA again
Ende 14:35

  1   Sun Jan 10 06:27:26 2021 Maximilian TomaschekLowflux recommissioning1Ramon+Tracker252.7  30 spills 10% 101    

Rate,15s spill pause, delay can be set by operator

  2   Sun Jan 10 06:32:13 2021 Maximilian TomaschekLowflux recommissioning2Ramon+Tracker194.3  30 spills 10% 101    

Rate

  3   Sun Jan 10 06:33:22 2021 Maximilian TomaschekLowflux recommissioning3Ramon+Tracker175.3  30 spills 10% 101    

Rate

  4   Sun Jan 10 06:33:44 2021 Maximilian TomaschekLowflux recommissioning5Ramon+Tracker100.4  30 spills 10% 101    

Rate

  5   Sun Jan 10 06:38:16 2021 Maximilian TomaschekLowflux recommissioning6Ramon+Tracker83  30 spills 10% 101    

Rate

  6   Sun Jan 10 06:40:06 2021 Maximilian TomaschekLowflux recommissioning7Ramon+Tracker62.4  30 spills 10% 101    

Rate, up till now looks like the same

  7   Sun Jan 10 06:42:03 2021 Maximilian TomaschekLowflux recommissioning8Ramon+Tracker252.7  30 spills   101 2  

Rate, factor 10 too high, role ~30kHz instead of 3kHz, Beta nominal

ELOG V3.1.5-fc6679b