ID Date Author Subject Run Number DUTdown Beam Energy Rate Setting Particle Type Events Data Beam Betatron Beam Chopper Beam Degrader Beam EFE Beam Orb Beam QuadScan PhantomType PhantomAngle
  289   Sun Nov 7 19:06:06 2021 Felix Ulrich-PurPeakfinder WET meas 7.11.2117 175.3nominal  Good  10   none 

Start: 10:06
WET measurement of PMMA slabs also used for TOF-CT (GSI)

2cm PMMA
"Accumalte Charge"

PMMA was activated. Had to wait ~10-15min after measurement

  290   Sun Nov 7 19:06:26 2021 Felix Ulrich-PurPeakfinder WET meas 7.11.2118 175.3nominal  Good  10   none 

Start: 10:30
WET measurement of PMMA slabs also used for TOF-CT (GSI)

3cm PMMA
"Accumalte Charge"

PMMA was activated. Had to wait ~10-15min after measurement

  291   Sun Nov 7 19:06:40 2021 Felix Ulrich-PurPeakfinder WET meas 7.11.2119 175.3nominal  Good  10   none 

Start: 11:00
WET measurement of PMMA slabs also used for TOF-CT (GSI)

4cm PMMA
"Accumalte Charge"

PMMA was activated. Had to wait ~10-15min after measurement

  292   Sun Nov 7 19:06:58 2021 Felix Ulrich-PurPeakfinder WET meas 7.11.2120 175.3nominal  Good  10   none 

Start: 11:24
WET measurement of PMMA slabs also used for TOF-CT (GSI)

5cm PMMA
"Accumalte Charge"

PMMA was activated. Had to wait ~10-15min after measurement

  293   Sun Nov 7 19:07:31 2021 Felix Ulrich-PurPeakfinder WET meas 7.11.2121 194.3nominal  Good  10   none 

Start: 11:44
WET measurement of PMMA slabs also used for TOF-CT (GSI)

5cm PMMA. higher energy for Felix Feichtinger
"Accumalte Charge"

coarse run

  294   Sun Nov 7 19:08:47 2021 Felix Ulrich-PurPeakfinder WET meas 7.11.2122 194.3nominal  Good  10   none 

Start: 12:00
WET measurement of PMMA slabs also used for TOF-CT (GSI)

5cm PMMA. higher energy for Felix Feichtinger
"Accumalte Charge"

more detailed measurement

  295   Sun Nov 7 19:09:50 2021 Felix Ulrich-PurFluence loss meas 7.11.2124 252.7II(BETA+EFE)Proton Good      none 

Start: 13:58

Alignment run without phantom

  296   Sun Nov 7 19:10:31 2021 Felix Ulrich-PurFluence loss meas 7.11.2125 145.4II(BETA+EFE)Proton40e6Good      none 

Start: 14:04

Reference run without phantom

  301   Sun Nov 7 21:04:43 2021 Felix Ulrich-PurFluence loss meas 7.11.2130 100.4II(BETA+EFE)Proton40e6Good      none 

Start: 21.04

reference run 100.4MeV

  304   Sun Nov 21 14:05:18 2021 Simon WaidComparison SiC, Si, Si LGAD and Diamond3 62.4I(EFE+SID0)Proton10000         

Bias voltage SiC: 800

Bias voltage Si LGAD: -100

From a beam perspective the SiC detector was the first target, on oscilloscope channel 4, the Si LGAD was the second target connected to oscilloscope channel 1.

Triggering source was the Si LGAD detector. The SiC detector was most probably defective.

Data was stored on the HiBPM-Osci-2 NUC under c:\measurements

  311   Sat Oct 8 07:10:31 2022 Christian IrmlerRD50-MPW3 - First Testbeam               

Tracker xml file: hvcmos_mpw3_20221008_TLU_00-000053.xml

 