ID Date Author Subject Run Number DUT Beam Energy Rate Setting Particle Type Events Data Beam Betatrondown Beam Chopper Beam Degrader Beam EFE Beam Orb Beam QuadScan PhantomType PhantomAngle
  281   Sun Nov 7 18:59:28 2021 Felix Ulrich-PurLow flux test 7.11.219 62.4II(BETA+EFE) 30 spillsGood      none 

Start: 07:26 O(100kHz)

  282   Sun Nov 7 18:59:43 2021 Felix Ulrich-PurLow flux test 7.11.2110 252.7III(EFE+SID2) 30 spillsGood      none 

Start: 07:33 O(kHz)

  283   Sun Nov 7 19:00:20 2021 Felix Ulrich-PurLow flux test 7.11.2111 145.4III(EFE+SID2) 30 spillsGood      none 

Start: 07:39 O(kHz)

  284   Sun Nov 7 19:00:36 2021 Felix Ulrich-PurLow flux test 7.11.2112 62.4III(EFE+SID2) 30 spillsGood      none 

Start: 07:44 O(kHz)

  285   Sun Nov 7 19:00:54 2021 Felix Ulrich-PurPeakfinder WET meas 7.11.2113 175.3nominal  Good  10   none 

Start: 08:30
WET measurement of PMMA slabs also used for TOF-CT (GSI)

coarse run (to find Bragg range

  286   Sun Nov 7 19:02:25 2021 Felix Ulrich-PurPeakfinder WET meas 7.11.2114 175.3nominal  Good  10   none 

Start: 08:55
WET measurement of PMMA slabs also used for TOF-CT (GSI)

reference run (with Arduino Spill Start/Stop logic)
fringes in Bragg peak (trigger not optimal (maybe too much latency with Arduino))--> next measurement with "Accumalte Charge" instead of "Spill Start/Stop"

  287   Sun Nov 7 19:04:10 2021 Felix Ulrich-PurPeakfinder WET meas 7.11.2115 175.3nominal  Good  10   none 

Start: 09:14
WET measurement of PMMA slabs also used for TOF-CT (GSI)

reference run
"Accumalte Charge"

  288   Sun Nov 7 19:05:15 2021 Felix Ulrich-PurPeakfinder WET meas 7.11.2116 175.3nominal  Good  10   none 

Start: 09:38
WET measurement of PMMA slabs also used for TOF-CT (GSI)

1cm PMMA
"Accumalte Charge"

PMMA was activated. Had to wait ~10-15min after measurement

  289   Sun Nov 7 19:06:06 2021 Felix Ulrich-PurPeakfinder WET meas 7.11.2117 175.3nominal  Good  10   none 

Start: 10:06
WET measurement of PMMA slabs also used for TOF-CT (GSI)

2cm PMMA
"Accumalte Charge"

PMMA was activated. Had to wait ~10-15min after measurement

  290   Sun Nov 7 19:06:26 2021 Felix Ulrich-PurPeakfinder WET meas 7.11.2118 175.3nominal  Good  10   none 

Start: 10:30
WET measurement of PMMA slabs also used for TOF-CT (GSI)

3cm PMMA
"Accumalte Charge"

PMMA was activated. Had to wait ~10-15min after measurement

  291   Sun Nov 7 19:06:40 2021 Felix Ulrich-PurPeakfinder WET meas 7.11.2119 175.3nominal  Good  10   none 

Start: 11:00
WET measurement of PMMA slabs also used for TOF-CT (GSI)

4cm PMMA
"Accumalte Charge"

PMMA was activated. Had to wait ~10-15min after measurement

  292   Sun Nov 7 19:06:58 2021 Felix Ulrich-PurPeakfinder WET meas 7.11.2120 175.3nominal  Good  10   none 

Start: 11:24
WET measurement of PMMA slabs also used for TOF-CT (GSI)

5cm PMMA
"Accumalte Charge"

PMMA was activated. Had to wait ~10-15min after measurement

  293   Sun Nov 7 19:07:31 2021 Felix Ulrich-PurPeakfinder WET meas 7.11.2121 194.3nominal  Good  10   none 

Start: 11:44
WET measurement of PMMA slabs also used for TOF-CT (GSI)

5cm PMMA. higher energy for Felix Feichtinger
"Accumalte Charge"

coarse run

  294   Sun Nov 7 19:08:47 2021 Felix Ulrich-PurPeakfinder WET meas 7.11.2122 194.3nominal  Good  10   none 

Start: 12:00
WET measurement of PMMA slabs also used for TOF-CT (GSI)

5cm PMMA. higher energy for Felix Feichtinger
"Accumalte Charge"

more detailed measurement

  295   Sun Nov 7 19:09:50 2021 Felix Ulrich-PurFluence loss meas 7.11.2124 252.7II(BETA+EFE)Proton Good      none 

Start: 13:58

Alignment run without phantom

  296   Sun Nov 7 19:10:31 2021 Felix Ulrich-PurFluence loss meas 7.11.2125 145.4II(BETA+EFE)Proton40e6Good      none 

Start: 14:04

Reference run without phantom

  297   Sun Nov 7 19:11:24 2021 Felix Ulrich-PurFluence loss meas 7.11.2126Drehtisch145.4II(BETA+EFE)Proton60e6Good      Alu0

Start: 15:19

1 cm Alu stair phantom on rotary table (0deg--> always 1cm Al in beam)

  298   Sun Nov 7 19:12:57 2021 Felix Ulrich-PurFluence loss meas 7.11.2127Drehtisch145.4II(BETA+EFE)Proton60e6Good      Alu90

Start: 16:40

1 cm Alu stair phantom on rotary table (90deg--> 0.2-1cm Alu)

  299   Sun Nov 7 19:13:36 2021 Felix Ulrich-PurFluence loss meas 7.11.2128Drehtisch100.4II(BETA+EFE)Proton60e6Good      Alu90

Start: 18:00

1 cm Alu stair phantom on rotary table (90deg)

  300   Sun Nov 7 19:28:49 2021 Felix Ulrich-PurFluence loss meas 7.11.2129Drehtisch100.4II(BETA+EFE)Proton60e6Good      Alu0

Start: 19:28

1 cm Alu stair phantom on rotary table (0deg)