Home CMS Production Clean room MedAustron HEPHY testbeams old
testbeam-MA2018 testbeam-MA2019 testbeam-MA2020 testbeam-MA2021 testbeam-MA2023 testbeam-MA2024
  MedAustron 2020, Page 9 of 15  Not logged in ELOG logo
New entries since:Thu Jan 1 01:00:00 1970
IDdown Date Author Subject Run Number DUT Beam Energy Beam Betatron Beam Chopper Beam Degrader Beam EFE Beam Orb Beam QuadScan PhantomType PhantomAngle Events Data Text Attachments
  133   Sat Jul 4 04:46:54 2020 Felix Ulrich-PurEFE SID0 spot-size measurement23Lynx+RaMon+Octavius100.4   yes SID0     

 
  
  132   Sat Jul 4 04:46:14 2020 Felix Ulrich-PurEFE SID2 spot-size measurement22Lynx+RaMon+Octavius252.7   yes SID2    tested 180s exposure time of Lynx, no signal
in Lynx or Octavius

 
  
  131   Sat Jul 4 04:44:19 2020 Felix Ulrich-PurEFE SID2 spot-size measurement21Lynx+RaMon+Octavius+Film62.4   yes SID2    tested 180s exposure time of Lynx, no signal
in Lynx or Octavius

35min irradiation of film (no signal)
  
  130   Sat Jul 4 04:43:54 2020 Felix Ulrich-PurBeta+EFE spot-size measurement20Lynx+RaMon+Octavius83yes  yes      tested 180s exposure time of Lynx   
  129   Sat Jul 4 04:43:36 2020 Felix Ulrich-PurBeta+EFE spot-size measurement19Lynx+RaMon+Octavius100.4yes  yes      tested 180s exposure time of Lynx   
  128   Sat Jul 4 04:16:41 2020 Felix Ulrich-PurBeta+EFE spot-size measurement18Lynx+RaMon+Octavius145.4yes  yes      tested 180s exposure time of Lynx   
  127   Sat Jul 4 04:16:25 2020 Felix Ulrich-PurBeta+EFE spot-size measurement17Lynx+RaMon+Octavius175.3yes  yes      tested 180s exposure time of Lynx   
  126   Sat Jul 4 04:16:09 2020 Felix Ulrich-PurBeta+EFE spot-size measurement16Lynx+RaMon+Octavius194.3yes  yes      tested 180s exposure time of Lynx   
  125   Sat Jul 4 04:15:29 2020 Felix Ulrich-PurBeta+EFE spot-size measurement15Lynx+RaMon+Octavius252.7yes  yes      tested 180s exposure time of Lynx   
  124   Sat Jul 4 04:06:58 2020 Felix Ulrich-PurBeta+EFE spot-size measurement14Lynx+RaMon+Octavius+Film62.4yes  yes      tested 120+180s exposure time of Lynx   
  123   Sat Jul 4 04:06:32 2020 Felix Ulrich-PurEFE SID0 spot-size measurement13Lynx+RaMon+Octavius83   yes SID0    tested 30s exposure time of Lynx   
  122   Sat Jul 4 04:06:04 2020 Felix Ulrich-PurEFE SID0 spot-size measurement12Lynx+RaMon+Octavius100.4   yes SID0    tested 30s exposure time of Lynx   
  121   Sat Jul 4 04:05:32 2020 Felix Ulrich-PurEFE SID0 spot-size measurement11Lynx+RaMon+Octavius175.3   yes SID0    tested 30 and 20s exposure time of Lynx   
  120   Sat Jul 4 04:05:13 2020 Felix Ulrich-PurEFE SID0 spot-size measurement10Lynx+RaMon+Octavius145.4   yes SID0    tested 60 and 30s exposure time of Lynx   
  119   Sat Jul 4 04:04:53 2020 Felix Ulrich-PurEFE SID0 spot-size measurement9Lynx+RaMon+Octavius194.3   yes SID0    tested 90,60 and 30s exposure time of Lynx   
  118   Sat Jul 4 04:04:09 2020 Felix Ulrich-PurEFE SID0 spot-size measurement8Lynx+RaMon+Octavius252.7   yes SID0    tested 90,60 and 30s exposure time of Lynx   
  117   Sat Jul 4 04:02:17 2020 Felix Ulrich-PurEFE SID0 spot-size measurement7Lynx+RaMon+Octavius+Film62.4   yes SID0    11 min irradiation of film (spot visible)

tested 180,90 and 60s exposure
time of Lynx (saturated at >=90)
  
  116   Sat Jul 4 04:01:29 2020 Felix Ulrich-PurEFE SID2 Energy measurement6Peakfinder62.4   yes SID2    6 spills per point -> no signal measurable
(always 0)
  
  115   Sat Jul 4 04:00:58 2020 Felix Ulrich-PurEFE SID2 Energy measurement5Peakfinder252.7   yes SID2    6 spills per point -> no signal measurable
(always 0)
  
  114   Sat Jul 4 04:00:31 2020 Felix Ulrich-PurEFE SID0 Energy measurement4Peakfinder252.7   yes SID0    2 spills per measurement point   
ELOG V3.1.5-fc6679b