  29   Sat Jun 13 00:14:44 2020 Felix Ulrich-PurCalibration TERA MPV 4004TERA-Cal800106020106dB none   

Setting 4

stopped because one slice could not be calibrated due to SW error

  30   Sat Jun 13 00:29:22 2020 Felix Ulrich-PurCalibration TERA MPV 4005TERA-Cal800106020106dB none   

Setting 4

S9 did not converge run stopped

 

  31   Sat Jun 13 00:34:32 2020 Felix Ulrich-PurCalibration TERA MPV 4006TERA-Cal800106020106dB none   

Setting 4

 

  28   Sat Jun 13 00:01:43 2020 Felix Ulrich-PurCalibration TERA MPV 3003TERA-Cal800106020106dB none   

Setting 4

 

  32   Sat Jun 13 00:54:47 2020 Felix Ulrich-PurCalibration TERA MPV 3007TERA-Cal800106020106dB none   

Setting 4

repeated 300, because of SW error in previous run

 

  26   Fri Jun 12 23:10:40 2020 Felix Ulrich-PurCalibration TERA MPV 2001TERA-Cal800106020106dB none   

Setting 4

was stopped, cause timout setting was not adjusted to spill structure

  27   Fri Jun 12 23:47:07 2020 Felix Ulrich-PurCalibration TERA MPV 2002TERA-Cal800106020106dB none   

Setting 4

slice R1 did not converge therofor 9 steps

  33   Sat Jun 13 01:07:52 2020 Felix Ulrich-PurCalibration TERA MPV 2008TERA-Cal800106020106dB none   

Setting 4

repeated 200, because of SW error in previous run

 

  3   Sun May 10 14:51:12 2020 Christian IrmlerCalibration Runs1Diamond + DSSD trackerNoise      none   

ADC delay, FIR, ped, noise,cal with DSSD tracker + diamond
DSSD HV= +/- 50V
Dia HV = +/- 250V

diamond not yet pumped

calibration files: pct_dia_01_silimond_tb_01-000001.xml

  92   Fri Jun 19 19:22:10 2020 Felix Ulrich-PurCalibration Diamond,Silimond,pCT6+Diamand30Diamond           

300V-800V in 100V steps

xml files:

/mnt/data/xmls/local/pct_dia_01_diamond_800V_02-000030.xml
/mnt/data/xmls/global/pct_dia_01_diamond_800V_beam_02-000030.xml

all calibration files

800 is only example replace with according voltage

  14   Sun May 10 17:44:50 2020 Felix Ulrich-PurCalibration Diamond+pCT6 500V11Diamond + DSSD-pCT6252.7nonenone    none   

Diamond+pCT6
Dia HV = +/- 250V

ADC, FIR, ped, noise, cal

  13   Sun May 10 17:38:55 2020 Felix Ulrich-PurCalibration Diamond+pCT6 400V10Diamond + DSSD-pCT6252.7nonenone    none   

Diamond+pCT6
Dia HV = +/- 200V

ADC, FIR, ped, noise, cal

  7   Sun May 10 16:47:10 2020 Christian IrmlerCalibration Diamond 600V4Diamond252.7nonenone    none   

diamond only
Dia HV = +/- 300V

ADC, FIR, ped, noise, cal

  6   Sun May 10 16:44:46 2020 Christian IrmlerCalibration Diamond 500V3Diamond252.7nonenone    none   

diamond only
Dia HV = +/- 250V

ADC, FIR, ped, noise, cal

  12   Sun May 10 17:21:28 2020 Felix Ulrich-PurCalibration Diamond 400V9Diamond252.7nonenone    none   

Diamond only
Dia HV = +/- 200V

ADC, FIR, ped, noise, cal

  11   Sun May 10 17:16:58 2020 Christian IrmlerCalibration Diamond 300V8Diamond252.7nonenone    none   

Diamond only
Dia HV = +/- 150V

ADC, FIR, ped, noise, cal

  9   Sun May 10 17:01:42 2020 Christian IrmlerCalibration Diamond + pCT-6 600V6Diamond + DSSD-pCT6252.7nonenone    none   

pCT-6 + diamond

DSSD HV = +/-50V
Dia HV = +/- 300V

ADC, FIR, ped, noise, cal

  10   Sun May 10 17:14:00 2020 Christian IrmlerCalibration Diamond + pCT-6 300V7Diamond + DSSD-pCT6252.7nonenone    none   

pCT-6 + diamond

DSSD HV = +/-50V
Dia HV = +/- 150V

ADC, FIR, ped, noise, cal

  124   Sat Jul 4 04:06:58 2020 Felix Ulrich-PurBeta+EFE spot-size measurement14Lynx+RaMon+Octavius+Film62.4yes  yes      

tested 120+180s exposure time of Lynx

  125   Sat Jul 4 04:15:29 2020 Felix Ulrich-PurBeta+EFE spot-size measurement15Lynx+RaMon+Octavius252.7yes  yes      

tested 180s exposure time of Lynx